510(k) Report

510(k) · Cleared

Cleared

Icrco 3600M

K122510510(k) TraditionalCleared Apr 26, 2013252-day reviewClass II21 CFR 892.1715

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icrco, Inc. · MUE Full Field Digital, System, X-Ray, Mammographic

Cleared on Apr 26, 2013 after a 252-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Apr 26, 2013
Review
252 days
Class
Class II
Product code
MUE
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K122510Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUE device

  1. #1 K253678 MAMMOMAT B.brilliant (11372950)score 13.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All MUE devices

MUE Full Field Digital, System, X-Ray, Mammographic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · MUE

3in 2025

2007: 62008: 52009: 82010: 102011: 102012: 42013: 72014: 12015: 12016: 12017: 82018: 22019: 22020: 42021: 42022: 32023: 12024: 12025: 32026 (partial year): 4
2007high 10 (2010) · low 1 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUE

255in 2025

2011: 192 days median2012: 123 days median2013: 207 days median2017: 95 days median2020: 188 days median2021: 164 days median2022: 404 days median2025: 255 days median2026 (partial year): 185 days median
2007high 404 (2022) · low 95 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUE

4in 2025

2008: 12009: 12010: 22011: 52014: 22017: 62018: 12019: 12023: 12025: 42026 (partial year): 1
2007high 6 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUE

40in 2025

2008: 42009: 32010: 22011: 102012: 62013: 112014: 72015: 62016: 192017: 1152018: 1452019: 802020: 492021: 252022: 362023: 292024: 332025: 40
2007high 145 (2018) · low 0 (2007)2025

MAUDE reports naming product code MUE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076—00
20085—14
20098—13
201010—22
201110192510
2012412306
20137207011
20141—27
20151—06
20161—019
20178956115
20182—1145
20192—180
20204188049
20214164025
20223404036
20231—129
20241—033
20253255440
202641851—

What is it indicated for?

The iCR3600M CR Mammography Feature together with iCR3600M CR Mammography Cassette comprise a device which, when used in conjunction with an iCRco 3600 CR System and a mammographic x-ray machine, generates digital mammographic images. The device is intended to be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammographic systems. The mammographic images can be interpreted by an MQSA qualified physician using either hardcopy film or softeopy display at a workstation.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K122510: Icrco 3600M.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122510

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026