# K122818: Olympus Maj-172 Instrument Tray for LTF Videoscope

- Pathway: 510(k) (Traditional)
- Applicant: Olympus Keymed (Medical & Industrial Equipment) LT
- Decided: Jan 4, 2013 after 112 days
- Product code: KCT (Sterilization Wrap Containers, Trays, Cassettes & Other Accessories)
- Device class: 2
- Regulation: 21 CFR 880.6850
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 6 generations back to K844002 (1984). 11 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 13 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K033222: Olympus Sterilization Trays, Olympus Winter & Ibe GmbH (2004) Summary page 1.
- K092682: Gyrus Acmi Flexible Endoscope Storage-Sterilization Trays, Gyrus Acmi, Inc. (2009) Summary page 1.

## Most influential descendants
- K133116: Video Extended Length Tray (Plastic - Radel), Video Tray Single Level (Plastic - Radel), Video Arthroscope Tray (Plastic (2014), 0 descendants of its own

## Indications for use
The Olympus MAJ-172 Instrument Tray for LTF Videoscope is intended to be used to enclose Olympus medical devices including hand instruments, trocars, camera heads, adapter, and endoscopes; to be sterilized by a health care provider. It is intended to allow steam or ethylene oxide sterilization of the enclosed medical device. The Olympus MAJ-172 Instrument Tray for LTF Videoscope is a purpose designed transport and sterilization case indicated for use in medical facilities to accommodate an endoscope from the Olympus LTF range of videoscopes, plus certain identified accessories, for reprocessing by autoclaving (steam) or ETO sterilization (depending on the model). LTF Type 160 and VH videoscopes are compatible with Steam and ETO sterilization cycles. LTF Type V3 and VP are compatible only with ETO sterilization cycles. The Olympus MAJ-172 Instrument Tray for LTF Videoscope is designed to secure and store a single videoscope and Its accessoriesfor sterilization at a healthcare facility. Refer to the instrument tray instructions for use for the proper orientation of the videoscope and accessories within the instrument tray. The recommended sterilization cycles for the LTF videoscopes are as follows' ETO Sterilization... Steam Sterilization: 100% ETO concentration: 735 mgIL Vacuum: 0.016 MPa minimum Temp: 57 0C *.Pressure: 0.101 MPa minimum Relative Humidity: 70% Temp: 135 C Processing (Hold) lime: 1 hour Exposure lime: 3 minutes Aeration ime: 12-hours Drying lime: 20 minutes

Source page: https://510kreport.com/device/K122818
Methodology: https://510kreport.com/methodology