# K123195: Carescape Respiratory Modules, E-Sco, E-Scov, E-Scaiov and Accessories

- Pathway: 510(k) (Traditional)
- Applicant: GE Healthcare Finland Oy
- Decided: Feb 28, 2013 after 140 days
- Product code: CCL (Analyzer, Gas, Oxygen, Gaseous-Phase)
- Device class: 2
- Regulation: 21 CFR 868.1720
- Recalls of this device: 4 (3 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K001814 (2000). 2 ancestors in all.
13 later devices descend from it; 2 cite it directly.
5 of 16 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K051092: Datex-Ohmeda S/5 Compact Airway Module (Model Family E-Caiovx) E-Caiovx, E-Caiov, E-Caio, E-Covx, E-Cov, E-CO & Accesacc, GE Healthcare (2005) Summary page 2.

## Most influential descendants
- K130584: Monitor B40 (2013), 8 descendants of its own
- K133576: Monitor B40 (2014), 7 descendants of its own
- K151063: Monitor B40 (2015), 5 descendants of its own
- K171580: Monitor B125, Monitor B105 (2017), 4 descendants of its own
- K143676: B40i Patient Monitor (2015), 4 descendants of its own

## Recalls of this device
- Sep 27, 2019 · Class I · Z-0757-2020: Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Sep 27, 2019 · Class I · Z-0754-2020: Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Sep 27, 2019 · Class I · Z-0756-2020: Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Aug 8, 2014 · Class II · Z-2617-2014: GE Healthcare has recently become aware of a potential safety issue due to a failure of an O2 sensor component associated with the CARESCAPE respiratory modules, the Airway Gas Option and their respec

## Indications for use
The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCAiO, E-sCAiOV) are indicated for use with a host device for monitoring respiratory parameters (C02, 02, N20, anesthetic agents, anesthetic agent identification, and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric, and neonatal patients. When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy. These modules are intended for use by qualified medical personnel only.

Source page: https://510kreport.com/device/K123195
Methodology: https://510kreport.com/methodology
