510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Dimension Ammonia Flex Reagent Cartridge

K123320510(k) TraditionalCleared Feb 15, 2013112-day reviewClass I21 CFR 862.1065Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics · JIF Enzymatic Method, Ammonia

Cleared on Feb 15, 2013 as substantially equivalent to K863840 (E.I. Dupont DE Nemours & Co., Inc., 1986) after a 112-day review. Its family tree reaches back 1 generation to 1986. It has been recalled 3 times.

Decided
Feb 15, 2013
Review
112 days
Class
Class I
Product code
JIF
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K863840 (1986). No later device cites it as a predicate yet.

K863840 · Meth. & Cali. for the Dimension Clini/Chem. System · E.I. Dupont DE Nemours & Co., Inc. · 19861985199019952000200520102015
  1. Root · 1 generation backK863840 Meth. & Cali. for the Dimension Clini/Chem. SystemE.I. Dupont DE Nemours & Co., Inc. · 1986
  2. This deviceK123320 Dimension Ammonia Flex Reagent CartridgeSiemens Healthcare Diagnostics · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K863840Meth. & Cali. for the Dimension Clini/Chem. SystemE.I. Dupont DE Nemours & Co., Inc.1986—
0 (this device)K123320Dimension Ammonia Flex Reagent CartridgeSiemens Healthcare Diagnostics2013—
FIG. 1 — Lineage of K123320, 1986–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JIF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K863840Meth. & Cali. for the Dimension Clini/Chem. System
Summary page 1
…cate device used to demonstrate substantial equivalence to the Dimension® Ammonia (AMM) Flex® reagent cartridge is the Dimension® Ammonia (AMON) previously cleared under k863840. 7. Device Description: The Dimension® Ammonia assay (AMM) is an enzymatic method that uses glutamate dehydrogenase (GLDH) and a…
E.I. Dupont DE Nemours & Co., Inc.1986—

Also cleared against K863840: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 29, 2017Class IIZ-1649-2018Device DesignVarious assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
Dec 16, 2016Class IIZ-1140-2017Under Investigation by firmSiemens Healthcare Diagnostics has determined that Dimension AMM (DF119) flex reagent cartridge lots EB7180, FB7152, BA7194, EA7223, BA7250 do not meet the 60-day calibration interval claim due to reagent instability and results may show…
Dec 16, 2016Class IIZ-1141-2017Under Investigation by firmSiemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interval claim due to reagent instability and results may show an…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 4 other devices also cite its predicate K863840.

How does its category compare?

All JIF devices

JIF Enzymatic Method, Ammonia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · JIF

1in 2019

2013: 22019: 1
2007high 2 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JIF

Too few JIF 510(k)s in any year since 2007 to give a median.

Recalls per year · JIF

1in 2015

2007: 12009: 12015: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JIF

5in 2025

2009: 12010: 32011: 72013: 22014: 12015: 32016: 42017: 72018: 12019: 12020: 12021: 12022: 22023: 42024: 12025: 5
2007high 7 (2011) · low 0 (2007)2025

MAUDE reports naming product code JIF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20080—00
20090—11
20100—03
20110—07
20120—00
20132—02
20140—01
20150—13
20160—04
20170—07
20180—01
20191—01
20200—01
20210—01
20220—02
20230—04
20240—01
20250—05
20260—0—

What is it indicated for?

The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma (heparin or EDTA) on the Dimension® clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders such as cirrhosis, hepatitis and Reye's syndrome

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123320: Dimension Ammonia Flex Reagent Cartridge.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123320

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026