510(k) · Cleared
ClearedAmplivue C. Difficile Assay
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Quidel Corporation · OZN C. Difficile Toxin Gene Amplification Assay
Cleared on Dec 13, 2012 as substantially equivalent to DEN120013 (unknown applicant, ) after a 43-day review. 1 later device descends from it.
- Decided
- Dec 13, 2012
- Review
- 43 days
- Class
- Class II
- Product code
- OZN
- Recalls
- None
- Predicates
- 1
1 ancestor in all. 1 later device descends from it, 1 citing it directly.
- This deviceK123355 Amplivue C. Difficile AssayQuidel Corporation · 2012
- Descendant · cites it directly · 0 of its ownK132235 Imdx C.Difficile for Abbott M2000Intelligent Medical Devices, Inc. · 2013
Most influential descendants
Showing 1 of 1 · See all- K132235Imdx C.Difficile for Abbott M2000Intelligent Medical Devices, Inc. · 2013 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K123355 | Amplivue C. Difficile Assay | Quidel Corporation | 2012 | — |
| +1 | K132235 | Imdx C.Difficile for Abbott M2000 | Intelligent Medical Devices, Inc. | 2013 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2013 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OZN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OZN device
- #1 K260045 cobas® Cdiff Nucleic acid test for use on the…score 13.4 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | DEN120013 | C. Difficile Nucleic Acid Amplification Test Assay Our reading of the summary gave K113358, which is not in FDA’s 510(k) list. The device the summary names is FDA’s De Novo DEN120013, C. Difficile Nucleic Acid Amplification Test Assay (), granted —; older De Novo requests were first filed under a 510(k) number, so this links to the grant. |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K132235 | Imdx C.Difficile for Abbott M2000 | 2013 | 0 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All OZN devicesOZN C. Difficile Toxin Gene Amplification Assay: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · OZN
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OZN
89in 2017
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OZN
2in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OZN
3in 2025
MAUDE reports naming product code OZN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 2 | — | 0 | 0 |
| 2013 | 4 | 79 | 0 | 0 |
| 2014 | 1 | — | 0 | 0 |
| 2015 | 1 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 5 | 89 | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 2 | — | 0 | 5 |
| 2022 | 0 | — | 0 | 1 |
| 2023 | 1 | — | 0 | 2 |
| 2024 | 0 | — | 1 | 7 |
| 2025 | 1 | — | 2 | 3 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The AmpliVueTM C. difficile assay is an in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile Toxin A gene (tcdA) in unformed stool specimens of patients suspected of having Clostridium difficile-associated disease (CDAD). The AmpliVue T M C. difficile assay is intended for use as an aid in diagnosis of CDAD. The assay utilizes helicase-dependent amplification (HDA) for the amplification of a highly conserved fragment of the Toxin A gene sequence and a self-contained disposable amplification detection device that allows for manual evaluation of assay results.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K123355 · printed Nov 5, 2026
Cite this page
510(k) Report. “K123355: Amplivue C. Difficile Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K123355
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026