510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating, Dispersion Aortic Perfusion Cannula

K123370510(k) TraditionalCleared Mar 13, 2013132-day reviewClass II21 CFR 870.4210Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Edwards Lifesciences, LLC · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Cleared on Mar 13, 2013 as substantially equivalent to K831769 (Research Industries Corp., 1983) and 1 other predicate after a 132-day review. Its family tree reaches back 1 generation to 1983. 1 later device descends from it. It has been recalled 4 times.

Decided
Mar 13, 2013
Review
132 days
Class
Class II
Product code
DWF
Recalls
4
Predicates
2

Its longest line of descent runs back 1 generation to K831769 (1983). 2 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2024: 1 descendant clearedK831769 · Various Cardiovascular Surgical Device · Research Industries Corp. · 1983 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant K831769 · Various Cardiovascular Surgical Device · Research Industries Corp. · 1983 · recalled0
  1. Root · 1 generation backK831769 Various Cardiovascular Surgical DeviceResearch Industries Corp. · 1983 · Recalled
  2. This deviceK123370 EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo…Edwards Lifesciences, LLC · 2013
  3. Descendant · cites it directly · 0 of its ownK233895 EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A…Edwards Lifesciences · 2024

Most influential descendants

Showing 1 of 1 · See all
  • K233895EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A, EZF21TA…Edwards Lifesciences · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K831769Various Cardiovascular Surgical DeviceResearch Industries Corp.19838
0 (this device)K123370EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating…Edwards Lifesciences, LLC2013—
+1K233895EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A…Edwards Lifesciences2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
FIG. 1 — Lineage of K123370, 1983–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWF device

  1. #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K831769Various Cardiovascular Surgical Device
Summary page 1
…ter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass 21 CFR Part 870.4210, Product Code DWF, Class 11 Predicate Device: K831769 - Aortic Perfusion Cannulae K033463 - Aortic Perfusion Pediatric Cannulae Device Description:…
Research Industries Corp.19838
*K033463Edwards Lifesciences Research Medical Pediatric Arterial Cannulae
Summary page 1
…21 CFR Part 870.4210, Product Code DWF, Class 11 Predicate Device: K831769 - Aortic Perfusion Cannulae K033463 - Aortic Perfusion Pediatric Cannulae Device Description: Edwards Aortic Perfusion Cannulae are sterile, non-pyrogenic, singl…
Edwards Lifesciences, LLC20041

Also cleared against K831769: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233895EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A, EZF21TA…Edwards Lifesciences20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 22, 2019Class IIZ-0420-2020Under Investigation by firmTheir is a possibility that the cannula may separate from its connector, potentially causing a breach of circuit during cardiopulmonary bypass which could result in significant blood loss.
Mar 7, 2018Class IIZ-3028-2018Labeling False and MisleadingThe expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Mar 7, 2018Class IIZ-3027-2018Labeling False and MisleadingThe expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Aug 26, 2013Class IIZ-0134-2014Process controlEdwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K233895 (Edwards Lifesciences, 2024) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K831769.

How does its category compare?

All DWF devices

DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · DWF

4in 2025

2007: 22008: 62009: 82010: 32011: 52012: 102013: 92014: 142015: 92016: 32017: 52018: 52019: 32020: 32021: 52022: 22023: 22024: 32025: 42026 (partial year): 8
2007high 14 (2014) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWF

89in 2025

2008: 95 days median2009: 148 days median2010: 62 days median2011: 85 days median2012: 110 days median2013: 110 days median2014: 110 days median2015: 46 days median2016: 32 days median2017: 127 days median2018: 30 days median2019: 120 days median2020: 30 days median2021: 358 days median2024: 29 days median2025: 89 days median2026 (partial year): 81 days median
2007high 358 (2021) · low 29 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWF

4in 2024

2007: 252008: 252009: 212010: 262011: 132012: 372013: 14, 1 Class I2014: 62015: 3, 1 Class I2016: 32017: 2, 1 Class I2018: 102019: 22021: 32024: 42026 (partial year): 20
2007high 37 (2012) · low 0 (2020)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DWF

1,020in 2025

2007: 1212008: 1262009: 992010: 1972011: 2122012: 2332013: 2382014: 3252015: 2182016: 1442017: 1012018: 1182019: 1792020: 1422021: 1622022: 2762023: 4702024: 4012025: 1,020
2007high 1,020 (2025) · low 99 (2009)2025

MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—25121
200869525126
200981482199
201036226197
201158513212
20121011037233
2013911014 (1)238
2014141106325
20159463 (1)218
20163323144
201751272 (1)101
201853010118
201931202179
20203300142
202153583162
20222—0276
20232—0470
20243294401
202548901,020
202688120—

What is it indicated for?

Aortic perfusion cannulae are Intended for perfusion of the aorta during short-term (5 6 hours) cardiopulmonary bypass procedures.

Aortic Perfusion Cannulae may be used in pediatric or adult populations based on the flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates.

Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123370: EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating, Dispersion Aortic Perfusion Cannula.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123370

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026