510(k) · Cleared
Recalled · Class II · 2019EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating, Dispersion Aortic Perfusion Cannula
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Edwards Lifesciences, LLC · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Cleared on Mar 13, 2013 as substantially equivalent to K831769 (Research Industries Corp., 1983) and 1 other predicate after a 132-day review. Its family tree reaches back 1 generation to 1983. 1 later device descends from it. It has been recalled 4 times.
- Decided
- Mar 13, 2013
- Review
- 132 days
- Class
- Class II
- Product code
- DWF
- Recalls
- 4
- Predicates
- 2
Its longest line of descent runs back 1 generation to K831769 (1983). 2 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 1 generation backK831769 Various Cardiovascular Surgical DeviceResearch Industries Corp. · 1983 · Recalled
- This deviceK123370 EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo…Edwards Lifesciences, LLC · 2013
- Descendant · cites it directly · 0 of its ownK233895 EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A…Edwards Lifesciences · 2024
Most influential descendants
Showing 1 of 1 · See all- K233895EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A, EZF21TA…Edwards Lifesciences · 2024 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K831769 | Various Cardiovascular Surgical Device | Research Industries Corp. | 1983 | 8 |
| 0 (this device) | K123370 | EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating… | Edwards Lifesciences, LLC | 2013 | — |
| +1 | K233895 | EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A… | Edwards Lifesciences | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DWF device
- #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K831769 | Various Cardiovascular Surgical DeviceSummary page 1…ter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass 21 CFR Part 870.4210, Product Code DWF, Class 11 Predicate Device: K831769 - Aortic Perfusion Cannulae K033463 - Aortic Perfusion Pediatric Cannulae Device Description:… | 1983 |
| * | K033463 | Edwards Lifesciences Research Medical Pediatric Arterial CannulaeSummary page 1…21 CFR Part 870.4210, Product Code DWF, Class 11 Predicate Device: K831769 - Aortic Perfusion Cannulae K033463 - Aortic Perfusion Pediatric Cannulae Device Description: Edwards Aortic Perfusion Cannulae are sterile, non-pyrogenic, singl… | 2004 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K233895 | EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A, EZF21TA… | 2024 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 22, 2019 | Class II | Z-0420-2020 | Under Investigation by firm | Their is a possibility that the cannula may separate from its connector, potentially causing a breach of circuit during cardiopulmonary bypass which could result in significant blood loss. |
| Mar 7, 2018 | Class II | Z-3028-2018 | Labeling False and Misleading | The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life. |
| Mar 7, 2018 | Class II | Z-3027-2018 | Labeling False and Misleading | The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life. |
| Aug 26, 2013 | Class II | Z-0134-2014 | Process control | Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DWF devicesDWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.
FDA decisions per year · DWF
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DWF
89in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DWF
4in 2024
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DWF
1,020in 2025
MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 25 | 121 |
| 2008 | 6 | 95 | 25 | 126 |
| 2009 | 8 | 148 | 21 | 99 |
| 2010 | 3 | 62 | 26 | 197 |
| 2011 | 5 | 85 | 13 | 212 |
| 2012 | 10 | 110 | 37 | 233 |
| 2013 | 9 | 110 | 14 (1) | 238 |
| 2014 | 14 | 110 | 6 | 325 |
| 2015 | 9 | 46 | 3 (1) | 218 |
| 2016 | 3 | 32 | 3 | 144 |
| 2017 | 5 | 127 | 2 (1) | 101 |
| 2018 | 5 | 30 | 10 | 118 |
| 2019 | 3 | 120 | 2 | 179 |
| 2020 | 3 | 30 | 0 | 142 |
| 2021 | 5 | 358 | 3 | 162 |
| 2022 | 2 | — | 0 | 276 |
| 2023 | 2 | — | 0 | 470 |
| 2024 | 3 | 29 | 4 | 401 |
| 2025 | 4 | 89 | 0 | 1,020 |
| 2026 | 8 | 81 | 20 | — |
What is it indicated for?
Aortic perfusion cannulae are Intended for perfusion of the aorta during short-term (5 6 hours) cardiopulmonary bypass procedures.
Aortic Perfusion Cannulae may be used in pediatric or adult populations based on the flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates.
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K123370 · printed Nov 6, 2026
Cite this page
510(k) Report. “K123370: EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating, Dispersion Aortic Perfusion Cannula.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123370
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026