# K123370: EZ Glide Aortic Cannula, Aortic Perfusion Cannula with Duraflo Coating, Dispersion Aortic Perfusion Cannula

- Pathway: 510(k) (Traditional)
- Applicant: Edwards Lifesciences, LLC
- Decided: Mar 13, 2013 after 132 days
- Product code: DWF (Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass)
- Device class: 2
- Regulation: 21 CFR 870.4210
- Recalls of this device: 4 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K831769 (1983). 2 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K831769: Various Cardiovascular Surgical Device, Research Industries Corp. (1983) Summary page 1.
- K033463: Edwards Lifesciences Research Medical Pediatric Arterial Cannulae, Edwards Lifesciences, LLC (2004) Summary page 1.

## Most influential descendants
- K233895: EZ Glide Aortic Perfusion Cannulae (EZC21A, EZC21TA, EZC24A, EZC24TA, EZF21A, EZF21TA, EZF24A, EZF24TA, EZS21A, EZS21TA, EZS24A and EZS24TA); OptiSite Arterial Perfusion Cannulae (OPTI16, OPTI18, OPTI20 and OPTI22); EndoReturn Arterial Cannulae (ER21B and ER23B); (2024), 0 descendants of its own

## Recalls of this device
- Oct 22, 2019 · Class II · Z-0420-2020: Their is a possibility that the cannula may separate from its connector, potentially causing a breach of circuit during cardiopulmonary bypass which could result in significant blood loss.
- Mar 7, 2018 · Class II · Z-3028-2018: The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
- Mar 7, 2018 · Class II · Z-3027-2018: The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
- Aug 26, 2013 · Class II · Z-0134-2014: Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.

## Indications for use
Aortic perfusion cannulae are Intended for perfusion of the aorta during short-term (5 6 hours) cardiopulmonary bypass procedures. Aortic Perfusion Cannulae may be used in pediatric or adult populations based on the flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.