510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Pentaray Nav Eco High-Density Mapping Catheter

K123837510(k) SpecialCleared Mar 5, 201382-day reviewClass II21 CFR 870.1220Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biosense Webster, Inc. · MTD Catheter, Intracardiac Mapping, High-Density Array

Cleared on Mar 5, 2013 as substantially equivalent to K120425 (Biosense Webster, Inc., 2012) after an 82-day review. Its family tree reaches back 5 generations to 1999. 1 of the 23 devices in its family, counting itself, has a Class I recall. 11 later devices descend from it. It has been recalled 5 times.

Decided
Mar 5, 2013
Review
82 days
Class
Class II
Product code
MTD
Recalls
5
Predicates
1

Its longest line of descent runs back 5 generations to K983171 (1999). 11 ancestors in all, the oldest K970264 (1997). 11 later devices descend from it, 4 citing it directly. 1 of the 23 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors2000: 2 other ancestors1998: 1 other ancestor1997: 1 other ancestor2019: 1 descendant cleared2020: 2 descendants cleared2021: 1 descendant cleared2022: 1 descendant cleared2024: 3 descendants cleared, 1 with a Class I recall12023: 3 descendants clearedK983171 · Constellation Multiple Electrode Pacing and Recording System · Boston Scientific Corp · 1999 · recalledK992777 · Constellation Catheters, Model 8039 · Boston Scientific Corp · 1999 · recalledK021232 · Constellation Catheters, Model 8031M · Boston Scientific Corp · 2002 · recalledK050217 · Flower High-Density Mapping Catheter, 4-4-4 MM Electrode Spacing), (2-6-2 MM Electrode Spacing) · Biosense Webster, Inc. · 2005K120425 · Pentaray Nav High-Density Mapping Catheter · Biosense Webster, Inc. · 2012 · recalled1995200020052010201520202025
  1. Root · 5 generations backK983171 Constellation Multiple Electrode Pacing and Recording SystemBoston Scientific Corp · 1999 · Recalled
  2. 4 generations backK992777 Constellation Catheters, Model 8039Boston Scientific Corp · 1999 · Recalled
  3. 3 generations backK021232 Constellation Catheters, Model 8031MBoston Scientific Corp · 2002 · Recalled
  4. 2 generations backK050217 Flower High-Density Mapping Catheter, 4-4-4 MM Electrode Spacing)…Biosense Webster, Inc. · 2005
  5. 1 generation backK120425 Pentaray Nav High-Density Mapping CatheterBiosense Webster, Inc. · 2012 · Recalled
  6. This deviceK123837 Pentaray Nav Eco High-Density Mapping CatheterBiosense Webster, Inc. · 2013
  7. Descendant · cites it directly · 5 of its ownK201750 Pentaray Nav eco High-Density Mapping CatheterBiosense Webster, Inc. · 2020
  8. Descendant · 2 generations on · 3 of its ownK231312 PENTARAY® NAV ECO High Density Mapping Catheter, DECANAV® Mapping…Biosense Webster, Inc. · 2023
  9. Descendant · cites it directly · 2 of its ownK193237 Carto Octaray Mapping Catheter with Trueref TechnologyBiosense Webster, Inc. · 2020
  10. Descendant · 2 generations on · 1 of its ownK211438 OPTRELL Mapping Catheter with TRUEref TechnologyBiosense Webster, Inc. · 2022
  11. Descendant · 3 generations on · 0 of its ownK241224 Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter…Surgical Instrument Service and Savings Inc.(Dba Medline Ren · 2024 · Recalled (Class I)

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2019: 12020: 22021: 12022: 12023: 32024: 3
2019high 3 (2023) · low 0 (2025)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 11 · See all
  • K201750Pentaray Nav eco High-Density Mapping CatheterBiosense Webster, Inc. · 2020 · cites it directly5descendants
  • K231312PENTARAY® NAV ECO High Density Mapping Catheter, DECANAV® Mapping Catheter, Webster® CS…Biosense Webster, Inc. · 2023 · 2 generations on3descendants
  • K193237Carto Octaray Mapping Catheter with Trueref TechnologyBiosense Webster, Inc. · 2020 · cites it directly2descendants
  • K211438OPTRELL Mapping Catheter with TRUEref TechnologyBiosense Webster, Inc. · 2022 · 2 generations on1descendant
  • K241224Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter with EZ Steer…Surgical Instrument Service and Savings Inc.(Dba Medline Ren · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K983171Constellation Multiple Electrode Pacing and Recording SystemBoston Scientific Corp19991
−4K992777Constellation Catheters, Model 8039Boston Scientific Corp19991
−3K021232Constellation Catheters, Model 8031MBoston Scientific Corp20021
−2K050217Flower High-Density Mapping Catheter, 4-4-4 MM Electrode Spacing), (2-6-2 MM…Biosense Webster, Inc.2005—
−1K120425Pentaray Nav High-Density Mapping CatheterBiosense Webster, Inc.20125
0 (this device)K123837Pentaray Nav Eco High-Density Mapping CatheterBiosense Webster, Inc.2013—
+1K201750Pentaray Nav eco High-Density Mapping CatheterBiosense Webster, Inc.2020—
+2K231312PENTARAY® NAV ECO High Density Mapping Catheter, DECANAV® Mapping Catheter…Biosense Webster, Inc.2023—
+1K193237Carto Octaray Mapping Catheter with Trueref TechnologyBiosense Webster, Inc.2020—
+2K211438OPTRELL Mapping Catheter with TRUEref TechnologyBiosense Webster, Inc.2022—
+3K241224Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter with EZ Steer…Surgical Instrument Service and Savings Inc.(Dba Medline Ren2024Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
200020
200120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202020
202110
202210
202330
202431
FIG. 1 — Lineage of K123837, 1999–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MTD device

  1. #1 K231312 PENTARAY® NAV ECO High Density Mapping…score 16.4 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120425Pentaray Nav High-Density Mapping Catheter
Summary page 1
…(21 CFR 870.1220, Product Code MTD) Predicate Devices: PENTrARAY® NAV High-Density Mapping Catheter 5 10O(k): K 120425 Manufacturer: Biosense Webster, Inc. 3333 Diamond Canyon Road…
Biosense Webster, Inc.20125

Also cleared against K120425: 1 other device →

What descends from it? (11)

See all 11

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201750Pentaray Nav eco High-Density Mapping CatheterBiosense Webster, Inc.20205
K193237Carto Octaray Mapping Catheter with Trueref TechnologyBiosense Webster, Inc.20202
K201806Reprocessed PentaRay Nav eco High-Density Mapping CatheterSterilmed Inc. (Johnson and Johnson)20210
K190785Reprocessed PentaRay Nav eco High-Density Mapping CatheterInnovative Health, LLC20190

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 22, 2020Class IIZ-2372-2020Error in labelingMapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
May 17, 2017Class IIZ-0211-2018OtherContraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
May 17, 2017Class IIZ-0212-2018OtherContraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
Apr 14, 2016Class IIZ-1591-2016Use errorBiosense Webster is recalling the Pentaray Catheters to clarify the contraindication language in the Instructions For Use (IFU) and product labeling for this catheter relative to patients with prosthetic valves.
Feb 20, 2014Class IIZ-1176-2014Process designBiosense Webster is recalling the PentaRay Nav Catheter because it has the potential to have partial tip separation from the catheter shaft. Biosense Webster is expanding the recall on 5/1/14 to include all lots of the PentaRay Nav…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K241224Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter with EZ Steer Technology and Medline…Surgical Instrument Service and Savings Inc.(Dba Medline Ren · 2024 · 3 generations on · ◆ in the 2024 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 2 companies; a new generation was cleared every 2.5 years on median.
  • 7 of the 11 ancestors have a recall of any class.
  • Of its descendants, K201750 (Biosense Webster, Inc., 2020) has gone furthest, with 5 descendants of its own.
  • 1 other device also cites its predicate K120425.

How does its category compare?

All MTD devices

MTD Catheter, Intracardiac Mapping, High-Density Array: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · MTD

2in 2024

2012: 22013: 22014: 22016: 12017: 22018: 12020: 42021: 12022: 22023: 32024: 22026 (partial year): 2
2007high 4 (2020) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MTD

66in 2023

2020: 169 days median2023: 66 days median
2007high 169 (2020) · low 66 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MTD

2in 2017

2017: 2
2007high 2 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MTD

385in 2025

2007: 12008: 32009: 62010: 192011: 142012: 172013: 302014: 282015: 362016: 352017: 382018: 492019: 772020: 872021: 992022: 1452023: 1842024: 3782025: 385
2007high 385 (2025) · low 1 (2007)2025

MAUDE reports naming product code MTD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—01
20080—03
20090—06
20100—019
20110—014
20122—017
20132—030
20142—028
20150—036
20161—035
20172—238
20181—049
20190—077
20204169087
20211—099
20222—0145
20233660184
20242—0378
20250—0385
20262—0—

What is it indicated for?

The Biosense Webster PENTARAY® NAV eco High-Density Mapping Catheter is indicated for

multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording

or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular

regions of the heart. The PENTARAY® NAV eco High-Density Mapping Catheter provides

location information when used with compatible CARTO® 3 EP Navigation Systems. (This

catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.x.)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123837: Pentaray Nav Eco High-Density Mapping Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123837

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026