510(k) Report

510(k) · Cleared

Cleared

Synthes Synflate Vertevral Balloon System

K130146510(k) TraditionalCleared May 20, 2013118-day reviewClass II21 CFR 888.3027

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Synthes (Usa), LLC · NDN Cement, Bone, Vertebroplasty

Cleared on May 20, 2013 as substantially equivalent to K110604 (Synthes, 2011) and 1 other predicate after a 118-day review. Its family tree reaches back 6 generations to 2000.

Decided
May 20, 2013
Review
118 days
Class
Class II
Product code
NDN
Recalls
None
Predicates
2

Its longest line of descent runs back 6 generations to K001134 (2000). 18 ancestors in all, the oldest K981251 (1998). No later device cites it as a predicate yet.

  1. Root · 6 generations backK001134 Gazelle Balloon Dilatation CatheterBoston Scientific Corp · 2000
  2. 5 generations backK021735 Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc. · 2002 · Recalled
  3. 4 generations backK050389 Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp · 2005
  4. 3 generations backK053116 Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp · 2005
  5. 2 generations backK080982 Sterling ES Monorail and Over-the-Wire Pta Balloon Dilatation…Boston Scientific Corp · 2008
  6. 1 generation backK110604 VBB SystemSynthes · 2011 · Recalled
  7. This deviceK130146 Synthes Synflate Vertevral Balloon SystemSynthes (Usa), LLC · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K001134Gazelle Balloon Dilatation CatheterBoston Scientific Corp2000—
−5K021735Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc.200223
−4K050389Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp2005—
−3K053116Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp2005—
−2K080982Sterling ES Monorail and Over-the-Wire Pta Balloon Dilatation CathetersBoston Scientific Corp2008—
−1K110604VBB SystemSynthes20111
0 (this device)K130146Synthes Synflate Vertevral Balloon SystemSynthes (Usa), LLC2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
199910
200010
200110
200310
200420
200520
200910
201020
FIG. 1 — Lineage of K130146, 2000–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NDN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NDN device

  1. #1 K241775 XeliteMed VertehighFix High Viscosity Spinal…score 14.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K110604VBB System
Summary page 1
…oplasty Class 11; Product Code NDN Predicate Devices: Synthes Vertebral Body Balloon (VBB) System (KI 110604) Kyphon KyphX Xpander Inflatable Bone Tamp (K04 1454) Device Description:…
Synthes20111
*K041454Kyphx Xpander Inflatable Bone Tamps Model#KO8A,KO9A,K13A
Summary page 1
…Devices: Synthes Vertebral Body Balloon (VBB) System (KI 110604) Kyphon KyphX Xpander Inflatable Bone Tamp (K04 1454) Device Description: The Synflate Vertebral Balloon System is a comprehensive array of…
Kyphon, Inc.20045

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 7 of the 18 ancestors have a recall of any class.

How does its category compare?

All NDN devices

NDN Cement, Bone, Vertebroplasty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · NDN

2in 2025

2007: 92008: 12009: 62010: 52011: 112012: 72013: 72014: 72015: 62016: 42017: 52018: 32019: 22020: 42021: 22022: 22023: 22024: 32025: 22026 (partial year): 1
2007high 11 (2011) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NDN

90in 2024

2007: 128 days median2009: 211 days median2010: 215 days median2011: 164 days median2012: 90 days median2013: 118 days median2014: 106 days median2015: 76 days median2016: 154 days median2017: 125 days median2018: 111 days median2020: 157 days median2024: 90 days median
2007high 215 (2010) · low 76 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NDN

4in 2013

2009: 12013: 4
2007high 4 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NDN

626in 2025

2007: 212008: 392009: 702010: 3962011: 982012: 892013: 2422014: 2312015: 2042016: 1822017: 1912018: 1442019: 2272020: 2122021: 3442022: 2702023: 3112024: 4542025: 626
2007high 626 (2025) · low 21 (2007)2025

MAUDE reports naming product code NDN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20079128021
20081—039
20096211170
201052150396
201111164098
2012790089
201371184242
201471060231
20156760204
201641540182
201751250191
201831110144
20192—0227
202041570212
20212—0344
20222—0270
20232—0311
20243900454
20252—0626
20261—0—

What is it indicated for?

The Synflate Vertebral Balloon System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethylmethacrylate (PMM4A) bone cements indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130146: Synthes Synflate Vertevral Balloon System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K130146

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026