# K130276: Dimension Exl with LM System

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Healthcare Diagnostics, Inc.
- Decided: Mar 22, 2013 after 45 days
- Product code: JJE (Analyzer, Chemistry (Photometric, Discrete), For Clinical Use)
- Device class: 1
- Regulation: 21 CFR 862.2160
- Recalls of this device: 6 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 3 generations back to K970389 (1997). 3 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K073604: Dimension Exl with LM Clinical Chemistry System, Dimension FT4L Flex Reagent Cartridge, and Loci Thyroid Calibrator, Mod, Dade Behring, Inc. (2008) Summary page 2.

## Most influential descendants
- K140842: Dimension Loci Free Thyroxine Flex Reagent Cartridge, FT4L, Dimension Loci Thyroid Stimulating Hormone Flex Reagent Cart (2014), 0 descendants of its own

## Recalls of this device
- Sep 28, 2018 · Class II · Z-0409-2019: Thermal chamber motors have electrical cables with single insulation rather than the required double insulation. There is no exposed current due to the single insulation that could result in a potenti
- Nov 10, 2015 · Class II · Z-0458-2016: Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located
- Nov 10, 2015 · Class II · Z-0459-2016: Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located
- Apr 2, 2015 · Class II · Z-1615-2015: Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as describ
- Jul 16, 2014 · Class II · Z-2553-2014: The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
- May 27, 2014 · Class III · Z-2133-2014: There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a

## Indications for use
The Dimension® EXLTM with LM system is an in vitro diagnostic device that is intended to measure a variety of analytes in human body fluids. The system utilizes photometric, turbidimetric, chemiluminescence and integrated ion selective muitisensor technology for chemical and immunochemnical applications for clinical use. The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXLTM with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.