510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Heartstring Iii Proximal Seal System

K130382510(k) TraditionalCleared Jun 5, 2013111-day reviewClass II21 CFR 870.4450Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corporation · DXC Clamp, Vascular

Cleared on Jun 5, 2013 as substantially equivalent to K080169 (Guidant Corporation, Cardiac Surgery, 2008) after a 111-day review. Its family tree reaches back 4 generations to 1980. 2 later devices descend from it. It has been recalled 3 times.

Decided
Jun 5, 2013
Review
111 days
Class
Class II
Product code
DXC
Recalls
3
Predicates
1

Its longest line of descent runs back 4 generations to K800122 (1980). 2 later devices descend from it, 1 citing it directly.

  1. Root · 4 generations backK800122 Hancock Aortic PunchHancock Laboratories, Inc. · 1980
  2. 3 generations backK980128 Baladi Inverter with Coaxial Aortic PunchCardio Medical Solutions, Inc. · 1998
  3. 2 generations backK022880 Guidant Proximal Seal System, Models PSS-4500 & PS-4500Guidant Corp. · 2002 · Recalled
  4. 1 generation backK080169 Heartstring Iii Proximal Seal SystemGuidant Corporation, Cardiac Surgery · 2008
  5. This deviceK130382 Heartstring Iii Proximal Seal SystemBoston Scientific Corporation · 2013
  6. Descendant · cites it directly · 1 of its ownK203307 VIOLAVascular Graft Solutions, Ltd. · 2021
  7. Descendant · 2 generations on · 0 of its ownK230248 VIOLAVascular Graft Solutions, Ltd. · 2023

Most influential descendants

Showing 2 of 2 · See all
  • K203307VIOLAVascular Graft Solutions, Ltd. · 2021 · cites it directly1descendant
  • K230248VIOLAVascular Graft Solutions, Ltd. · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K800122Hancock Aortic PunchHancock Laboratories, Inc.1980—
−3K980128Baladi Inverter with Coaxial Aortic PunchCardio Medical Solutions, Inc.1998—
−2K022880Guidant Proximal Seal System, Models PSS-4500 & PS-4500Guidant Corp.20021
−1K080169Heartstring Iii Proximal Seal SystemGuidant Corporation, Cardiac Surgery2008—
0 (this device)K130382Heartstring Iii Proximal Seal SystemBoston Scientific Corporation2013—
+1K203307VIOLAVascular Graft Solutions, Ltd.2021—
+2K230248VIOLAVascular Graft Solutions, Ltd.2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202310
FIG. 1 — Lineage of K130382, 1980–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXC device

  1. #1 K262086 PreludeSYNC™ Radial Compression Devicescore 13.5 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080169Heartstring Iii Proximal Seal System
Summary page 1
…I Proximal Seal System Vascular Clamp According to 21 CFR 870.4450 of the Federal Food, Drug and Cosmetic Act, the device classification is Class II, Product code DXC. K080169 HEARTSTRING III Proximal Seal System (SE: 8 Feb 2008) The HEARTSTRING III-Proximal Seal System is a device designed to enable th…
Guidant Corporation, Cardiac Surgery2008—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203307VIOLAVascular Graft Solutions, Ltd.20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 15, 2025Class IIZ-2588-2025Under Investigation by firmThree failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Aug 15, 2025Class IIZ-2587-2025Under Investigation by firmThree failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Aug 15, 2025Class IIZ-2586-2025Under Investigation by firmThree failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 5 companies; a new generation was cleared every 5.4 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K203307 (Vascular Graft Solutions, Ltd., 2021) has gone furthest, with 1 descendant of its own.

How does its category compare?

All DXC devices

DXC Clamp, Vascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · DXC

2in 2025

2007: 102008: 32009: 42010: 22011: 62012: 52013: 142014: 42015: 52016: 32017: 32018: 22019: 22020: 32021: 62022: 12023: 42024: 12025: 22026 (partial year): 4
2007high 14 (2013) · low 1 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXC

215in 2023

2007: 52 days median2008: 29 days median2009: 83 days median2011: 51 days median2012: 73 days median2013: 89 days median2014: 157 days median2015: 71 days median2016: 91 days median2017: 90 days median2020: 209 days median2021: 37 days median2023: 215 days median2026 (partial year): 125 days median
2007high 215 (2023) · low 29 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXC

2in 2023

2008: 22009: 22010: 12011: 32014: 12022: 12023: 2
2007high 3 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXC

442in 2025

2007: 2982008: 3022009: 5392010: 6002011: 6912012: 5482013: 9202014: 4942015: 3742016: 4942017: 4162018: 3912019: 3932020: 4042021: 3712022: 2732023: 6742024: 3752025: 442
2007high 920 (2013) · low 273 (2022)2025

MAUDE reports naming product code DXC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200710520298
20083292302
20094832539
20102—1600
20116513691
20125730548
201314890920
201441571494
20155710374
20163910494
20173900416
20182—0391
20192—0393
202032090404
20216370371
20221—1273
202342152674
20241—0375
20252—0442
202641250—

What is it indicated for?

The HEARTSTRING III Proximal Seal System is intended for use by Physicians during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis to the aorta without application of an aortic clamp.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130382: Heartstring Iii Proximal Seal System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130382

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026