510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Volista 400; Volista 600

K130513510(k) SpecialCleared May 8, 201370-day reviewClass II21 CFR 878.4580Recalls 19

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Maquet S.A.S. · FTD Lamp, Surgical

Cleared on May 8, 2013 as substantially equivalent to K070442 (Maquet S.A., 2007) and 1 other predicate after a 70-day review. Its family tree reaches back 4 generations to 1991. It has been recalled 19 times.

Decided
May 8, 2013
Review
70 days
Class
Class II
Product code
FTD
Recalls
19
Predicates
2

Its longest line of descent runs back 4 generations to K904965 (1991). 6 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK904965 Shampaine 1000 Series Angenieux AX-14 Duo,AX-4Smith & Nephew Equipment Group · 1991
  2. 3 generations backK040735 Alm X'ten (X10) Surgical Light SystemAlm S.A. · 2004 · Recalled
  3. 2 generations backK070442 Powerled Surgical Light SystemMaquet S.A. · 2007 · Recalled
  4. 1 generation backK113679 Lucea Led Surgical Light SystemMaquet S.A.S. · 2012 · Recalled
  5. This deviceK130513 Volista 400; Volista 600Maquet S.A.S. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K904965Shampaine 1000 Series Angenieux AX-14 Duo,AX-4Smith & Nephew Equipment Group1991—
−3K040735Alm X'ten (X10) Surgical Light SystemAlm S.A.20046
−2K070442Powerled Surgical Light SystemMaquet S.A.20073
−1K113679Lucea Led Surgical Light SystemMaquet S.A.S.20121
0 (this device)K130513Volista 400; Volista 600Maquet S.A.S.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199710
FIG. 1 — Lineage of K130513, 1991–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FTD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070442Powerled Surgical Light System
Summary page 1
…ice classification Class 11,acording to regulation number 21 CER 878.4580 Predicate Device identification " MAQUET POWERLEDTM Surgical Light System - 510(k) No. K070442 * MAQUET LUCEA LED® Surgical Light System - 510(k) No. Ki113679…
Maquet S.A.20073
*K113679Lucea Led Surgical Light System
Summary page 1
…R 878.4580 Predicate Device identification " MAQUET POWERLEDTM Surgical Light System - 510(k) No. K070442 * MAQUET LUCEA LED® Surgical Light System - 510(k) No. Ki113679 Page 1 of 4 MAQUET S.A.S. FDA 51…
Maquet S.A.S.20121

Also cleared against K070442: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 8, 2023Class IIZ-0758-2024Equipment maintenanceA potential for a light system to fall in the operating room.
May 6, 2022Class IIZ-1308-2022Process change controlPaint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical operation into the…
May 28, 2019Class IIZ-1860-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1862-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1852-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1861-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1858-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1857-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1853-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1856-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1859-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1854-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1864-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1863-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
May 28, 2019Class IIZ-1855-2019Device DesignPotential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
Jun 14, 2018Class IIZ-2502-2018Process controlCertain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket…
Apr 19, 2018Class IIZ-1937-2018Device DesignPotential keypad failure on VOLISTA StandOP Surgical Lights.
Feb 7, 2018Class IIZ-1349-2018Component design/selectionThe manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder…
May 29, 2015Class IIZ-1990-2015Device DesignMaquet Suspension Arms used with PowerLED/X-TEN/Volista Surgical Light Systems may grind at the junction of the suspension arms which can cause paint chips or particles to fall into the surgical field, when the suspension arms are mounted…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 4.8 years on median.
  • 5 of the 6 ancestors have a recall of any class.
  • 6 other devices also cite its predicate K070442.

How does its category compare?

All FTD devices

FTD Lamp, Surgical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · FTD

3in 2013

2007: 12010: 22011: 12013: 3
2007high 3 (2013) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FTD

70in 2013

2013: 70 days median
2007high 70 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FTD

6in 2023

2007: 112009: 92010: 32016: 12023: 6
2007high 11 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FTD

81in 2025

2007: 72008: 102009: 222010: 212011: 302012: 262013: 292014: 292015: 452016: 182017: 102018: 162019: 802020: 1052021: 652022: 1362023: 2512024: 1442025: 81
2007high 251 (2023) · low 7 (2007)2025

MAUDE reports naming product code FTD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—117
20080—010
20090—922
20102—321
20111—030
20120—026
2013370029
20140—029
20150—045
20160—118
20170—010
20180—016
20190—080
20200—0105
20210—065
20220—0136
20230—6251
20240—0144
20250—081
20260—0—

What is it indicated for?

MAQUET TRIOP VOLISTA surgical lights are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130513: Volista 400; Volista 600.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130513

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026