Mckesson Israel, Ltd. · MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Cleared on Dec 5, 2013 as substantially equivalent to K050561 (Medcon, Ltd., 2005) and 1 other predicate after a 195-day review. Its family tree reaches back 6 generations to 1991. 3 later devices descend from it. It has been recalled 10 times.
- Decided
- Dec 5, 2013
- Review
- 195 days
- Class
- Class II
- Product code
- MWI
- Recalls
- 10
- Predicates
- 2
Its longest line of descent runs back 6 generations to K910772 (1991). 38 ancestors in all, the oldest K862740 (1986). 3 later devices descend from it, 3 citing it directly.
- Root · 6 generations backK910772 Propaq 102, 104 and 106Protocol Systems, Inc. · 1991
- 5 generations backK972020 Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc. · 1998
- 4 generations backK980879 Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc. · 1998
- 3 generations backK982776 9303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc. · 1998
- 2 generations backK020440 NewcorSiemens Medical Systems, Inc. · 2002 · Recalled
- 1 generation backK101571 Xper Information Management and Flex Cardio Physiomonitoring SystemsWitt Biomedical Corp · 2010 · Recalled
- This deviceK131497 Mckesson Cardiology HemoMckesson Israel, Ltd. · 2013
- Descendant · cites it directly · 1 of its ownK170032 QMAPP Amplifier; QMAPP Hemodynamic Monitoring System; QMAPP PCM…Fysicon BV · 2017
- Descendant · cites it directly · 0 of its ownK241766 QMAPP® (Hemo, Hemo Lite, PCM, GO, Hybrid)Fysicon BV · 2025
- Descendant · cites it directly · 0 of its ownK230881 Change Healthcare Cardiology Hemodynamics™Change Healthcare Israel, Ltd. · 2023 · Recalled
Most influential descendants
Showing 3 of 3 · See all- K170032QMAPP Amplifier; QMAPP Hemodynamic Monitoring System; QMAPP PCM; QMAPP GO; QMAPP EPFysicon BV · 2017 · cites it directly1descendant
- K241766QMAPP® (Hemo, Hemo Lite, PCM, GO, Hybrid)Fysicon BV · 2025 · cites it directly0descendants
- K230881Change Healthcare Cardiology Hemodynamics™Change Healthcare Israel, Ltd. · 2023 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K910772 | Propaq 102, 104 and 106 | Protocol Systems, Inc. | 1991 | — |
| −5 | K972020 | Oscillomate Non-Invasive Blood Pressure Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −4 | K980879 | Oscillomate Nibp / SP02 Ems Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −3 | K982776 | 9303 Neonatal/Adult Vital Signs Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −2 | K020440 | Newcor | Siemens Medical Systems, Inc. | 2002 | 10 |
| −1 | K101571 | Xper Information Management and Flex Cardio Physiomonitoring Systems | Witt Biomedical Corp | 2010 | 8 |
| 0 (this device) | K131497 | Mckesson Cardiology Hemo | Mckesson Israel, Ltd. | 2013 | — |
| +1 | K170032 | QMAPP Amplifier; QMAPP Hemodynamic Monitoring System; QMAPP PCM; QMAPP GO… | Fysicon BV | 2017 | — |
| +1 | K241766 | QMAPP® (Hemo, Hemo Lite, PCM, GO, Hybrid) | Fysicon BV | 2025 | — |
| +1 | K230881 | Change Healthcare Cardiology Hemodynamics™ | Change Healthcare Israel, Ltd. | 2023 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1986 | 1 | 0 |
| 1990 | 2 | 0 |
| 1991 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 1 | 0 |
| 1995 | 5 | 0 |
| 1997 | 2 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 5 | 0 |
| 2001 | 1 | 0 |
| 2002 | 3 | 0 |
| 2003 | 2 | 0 |
| 2004 | 2 | 0 |
| 2005 | 2 | 0 |
| 2007 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2017 | 1 | 0 |
| 2023 | 1 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MWI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MWI device
- #1 K230457 Empatica Health Monitoring Platform…score 28.0 · 39 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K050561 | WindsurferSummary page 1…ication Product code: MWI Subsequent Product code: DQK Device Class: Class If Predicate Device: Medeon Ltd., Windsurfer (K050561I), WITT BIOMEDICAL CORP Philips Xper Flex Cardio Physiomonitoring System/… | 2005 |
| * | K101571 | Xper Information Management and Flex Cardio Physiomonitoring SystemsSummary page 1…T BIOMEDICAL CORP Philips Xper Flex Cardio Physiomonitoring System/ Xper Information Management System (K 10 157 1) Intended Use / Indications for Use McKesson Cardiologyim Hemo is intended for complete physiological/hemodynamic monitoring, cl… | 2010 |
| † | K012226 | Monitoring System,Model Argus PB-1000 | 2002 |
What descends from it? (3)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 22, 2025 | Class II | Z-2071-2025 | Software design | Due to complaints, software update may cause software to unexpectedly shutdown. |
| Dec 2, 2024 | Class II | Z-0957-2025 | Software design | Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s). |
| Dec 2, 2024 | Class II | Z-0958-2025 | Software design | Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s). |
| Aug 21, 2024 | Class II | Z-0128-2025 | Use error | A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs… |
| Aug 21, 2024 | Class II | Z-0126-2025 | Use error | A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs… |
| Mar 25, 2019 | Class II | Z-1490-2019 | Software Design Change | users are not notified of procedure medication discrepancies between the Vitals and Meds, the Procedure Notes and Patient Common Data screens in Horizon/McKesson Cardiology Hemo" (Hemo). The discrepancy may also affect administered… |
| Mar 25, 2019 | Class II | Z-1489-2019 | Software Design Change | users are not notified of procedure medication discrepancies between the Vitals and Meds, the Procedure Notes and Patient Common Data screens in Horizon/McKesson Cardiology Hemo" (Hemo). The discrepancy may also affect administered… |
| Mar 12, 2018 | Class II | Z-2968-2018 | Software Design Change | Change Healthcare has identified an issue where, under certain circumstances, the Real Time Monitor (RTM) may not display physiological signals. |
| Aug 27, 2014 | Class II | Z-0042-2015 | Software design (manufacturing process) | Software Error: The McKesson Cardiology Hemo calculation section incorrectly converts the Hemoglobin value before it is utilized in the applicable formula calculations. |
| Jun 25, 2014 | Class II | Z-0093-2015 | Labeling design | McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration.… |
The 5 latest of 10 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 22 years and passes through 5 companies; a new generation was cleared every 4.1 years on median.
- 11 of the 38 ancestors have a recall of any class.
- Of its descendants, K170032 (Fysicon BV, 2017) has gone furthest, with 1 descendant of its own.
How does its category compare?
All MWI devicesMWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.
FDA decisions per year · MWI
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MWI
247in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MWI
9so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MWI
182in 2025
MAUDE reports naming product code MWI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 9 | 42 | 1 | 3 |
| 2008 | 9 | 49 | 1 | 3 |
| 2009 | 6 | 112 | 2 | 20 |
| 2010 | 17 | 58 | 2 | 17 |
| 2011 | 11 | 78 | 3 | 16 |
| 2012 | 13 | 104 | 5 | 31 |
| 2013 | 20 | 85 | 3 | 42 |
| 2014 | 16 | 148 | 6 | 30 |
| 2015 | 19 | 127 | 1 | 31 |
| 2016 | 12 | 165 | 0 | 51 |
| 2017 | 13 | 205 | 2 | 85 |
| 2018 | 11 | 217 | 0 | 158 |
| 2019 | 14 | 255 | 4 | 169 |
| 2020 | 8 | 138 | 0 | 110 |
| 2021 | 9 | 161 | 0 | 188 |
| 2022 | 5 | 231 | 2 | 168 |
| 2023 | 8 | 250 | 0 | 179 |
| 2024 | 10 | 202 | 5 | 120 |
| 2025 | 4 | 247 | 0 | 182 |
| 2026 | 6 | 239 | 9 | — |
What is it indicated for?
McKesson Cardiology Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment. McKesson Cardiology Hemo is also intended for patient/procedural data management, such as documentation, logging, reporting, trending, storing, reviewing, carrying out clinical calculations and exporting various representations of the acquired data. Data may also be acquired from…
Read the full statement (142 words)
McKesson Cardiology Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
McKesson Cardiology Hemo is also intended for patient/procedural data management, such as documentation, logging, reporting, trending, storing, reviewing, carrying out clinical calculations and exporting various representations of the acquired data. Data may also be acquired from and/or sent to other devices, such as physiological monitoring systems, information management systems, image acquisition/storage devices, and other medical devices.
McKesson Cardiology Hemo is intended for use in the areas of: cardiology, cardiac catheterization, electrophysiology, radiology, invasive radiology, and other areas where patient/procedural data management is needed.
User-adjustable alarms (both visual and audible) available in the system alert the operator to anomalous occurrences and facilitate timely responses.
Use of the system is not intended for unattended patient monitoring or in situations where arrhythmia detection is required.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K131497 · printed Nov 6, 2026
Cite this page
510(k) Report. “K131497: Mckesson Cardiology Hemo.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131497