510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

AU Bicarbonate Reagent

K131546510(k) TraditionalCleared Oct 9, 2013133-day reviewClass II21 CFR 862.1160Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter Ireland, Inc. · KHS Enzymatic, Carbon-Dioxide

Cleared on Oct 9, 2013 after a 133-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). It has been recalled 4 times.

Decided
Oct 9, 2013
Review
133 days
Class
Class II
Product code
KHS
Recalls
4
Predicates
1

The summary cites one number, which could not be confirmed as its predicate (see the table below).

FIG. 1 — Lineage of K131546Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KHS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK9003Carbon Dioxide - DST
Number could not be read
Summary page 1
…cate Device: Cndidate(s) Predicate Ma1nufacturer Docket Number AU Biabonate Reagent .1 Carbon Dioxide - DST Trac K9003 5.0 Description: The AU Bicarbonate reagent it is a liquid, ready to use and consists of four R1 reagent vials in…

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 14, 2025Class IIZ-0997-2026Nonconforming Material/ComponentBicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.
Apr 5, 2023Class IIZ-1608-2023Under Investigation by firmThere is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
Apr 5, 2023Class IIZ-1607-2023Under Investigation by firmThere is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
Apr 5, 2023Class IIZ-1606-2023Under Investigation by firmThere is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.

No ancestor or descendant has a Class I recall.

How does its category compare?

All KHS devices

KHS Enzymatic, Carbon-Dioxide: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · KHS

1in 2017

2007: 22008: 12009: 12010: 32011: 12012: 22013: 22014: 12016: 12017: 1
2007high 3 (2010) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KHS

143in 2010

2010: 143 days median
2007high 143 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KHS

1in 2025

2013: 12019: 12023: 12025: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KHS

40in 2025

2008: 22009: 142011: 52012: 32013: 22014: 62015: 12016: 42017: 72018: 152019: 352020: 312021: 142022: 102023: 252024: 302025: 40
2007high 40 (2025) · low 0 (2007)2025

MAUDE reports naming product code KHS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20081—02
20091—014
2010314300
20111—05
20122—03
20132—12
20141—06
20150—01
20161—04
20171—07
20180—015
20190—135
20200—031
20210—014
20220—010
20230—125
20240—030
20250—140
20260—0—

What is it indicated for?

AU Bicarbonate reagent is intended for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers.

Bicarbonate measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131546: AU Bicarbonate Reagent.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131546

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026