510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Fresenius Dry Acid Dissolution Units

K131611510(k) TraditionalCleared Aug 29, 201387-day reviewClass II21 CFR 876.5820Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care North America · KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Cleared on Aug 29, 2013 after an 87-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 2 later devices descend from it. It has been recalled 2 times.

Decided
Aug 29, 2013
Review
87 days
Class
Class II
Product code
KPO
Recalls
2
Predicates
0

2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2020: 1 descendant cleared2026: 1 descendant cleared2010201520202025
  1. This deviceK131611 Fresenius Dry Acid Dissolution UnitsFresenius Medical Care North America · 2013
  2. Descendant · cites it directly · 1 of its ownK191093 Isopure Dry Acid Dissolution SystemIsopure, Corp. · 2020
  3. Descendant · 2 generations on · 0 of its ownK260799 Nipro Dry Acid Concentrate MixerNipro Technical Solutions, Incoporated · 2026

Most influential descendants

Showing 2 of 2 · See all
  • K191093Isopure Dry Acid Dissolution SystemIsopure, Corp. · 2020 · cites it directly1descendant
  • K260799Nipro Dry Acid Concentrate MixerNipro Technical Solutions, Incoporated · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K131611Fresenius Dry Acid Dissolution UnitsFresenius Medical Care North America2013—
+1K191093Isopure Dry Acid Dissolution SystemIsopure, Corp.2020—
+2K260799Nipro Dry Acid Concentrate MixerNipro Technical Solutions, Incoporated2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202610
FIG. 1 — Lineage of K131611, 2013–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPO device

  1. #1 K233159 pureFLOW 402, pureFLOW 400, pureFLOW 406…score 14.5 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191093Isopure Dry Acid Dissolution SystemIsopure, Corp.20201

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 23, 2015Class IIZ-0295-2016Device DesignDissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Oct 23, 2015Class IIZ-0296-2016Device DesignDissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K191093 (Isopure, Corp., 2020) has gone furthest, with 1 descendant of its own.

How does its category compare?

All KPO devices

KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder): FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · KPO

2in 2025

2007: 22008: 22009: 12011: 12012: 22013: 12014: 22015: 12016: 12017: 32018: 22020: 52021: 12022: 22023: 32024: 32025: 22026 (partial year): 2
2007high 5 (2020) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPO

112in 2024

2017: 192 days median2020: 270 days median2023: 263 days median2024: 112 days median
2007high 270 (2020) · low 112 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPO

6in 2025

2007: 52008: 22009: 12011: 22012: 102015: 32016: 12017: 12018: 112019: 12020: 312022: 22024: 12025: 62026 (partial year): 9
2007high 31 (2020) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPO

95in 2025

2007: 52008: 82009: 32010: 1072011: 2232012: 702013: 3,8972014: 15,9162015: 8,3402016: 2402017: 802018: 972019: 792020: 2442021: 632022: 252023: 592024: 2592025: 95
2007high 15,916 (2014) · low 3 (2009)2025

MAUDE reports naming product code KPO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—55
20082—28
20091—13
20100—0107
20111—2223
20122—1070
20131—03,897
20142—015,916
20151—38,340
20161—1240
20173192180
20182—1197
20190—179
2020527031244
20211—063
20222—225
20233263059
202431121259
20252—695
20262—9—

What is it indicated for?

The Fresenius Medical Care Dry Acid Dissolution Unit mixes Fresenius Medical Care distributed dry acid concentrate products with hemodialysis quality water. The resulting liquid acid concentrates are intended for use in three-stream hemodialysis machines calibrated for acid and bicarbonate concentrates.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K131611: Fresenius Dry Acid Dissolution Units.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131611

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026