510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification System, Conchasmart Column

K131912510(k) TraditionalCleared Mar 6, 2014253-day reviewClass II21 CFR 868.5450Recalls 3 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Teleflex, Inc. · BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Cleared on Mar 6, 2014 as substantially equivalent to K063758 (Teleflex Medical, 2007) and 2 other predicates after a 253-day review. Its family tree reaches back 6 generations to 1992. 2 of the 18 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 3 times, including 2 Class I recalls.

Decided
Mar 6, 2014
Review
253 days
Class
Class II
Product code
BTT
Recalls
3 (2 Class I)
Predicates
3

Its longest line of descent runs back 6 generations to K913368 (1992). 16 ancestors in all, the oldest K881625 (1988). 1 later device descends from it, 1 citing it directly. 2 of the 18 devices in this family, counting this one, have a Class I recall.

  1. Root · 6 generations backK913368 Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd. · 1992
  2. 5 generations backK953392 HC500 Servo-Controlled Heated Respiratory Humidifier (& Accessories)Fisher & Paykel Electronics, Ltd. · 1996
  3. 4 generations backK983112 Respiratory Humidifier, Model # MR850Fisher & Paykel Healthcare, Ltd. · 1998 · Recalled
  4. 3 generations backK020332 Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd. · 2003 · Recalled
  5. 2 generations backK033710 MR850 Respiratory Humidifier, Model MR850JHUFisher & Paykel Healthcare, Ltd. · 2004
  6. 1 generation backK061736 Comfort Flo Humidification System and Nasal CannulaTeleflex Medical · 2006
  7. This deviceK131912 Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification…Teleflex, Inc. · 2014
  8. Descendant · cites it directly · 0 of its ownK162242 Comfort Flo Humidification SystemTeleflexmedical, Inc. · 2017 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K162242Comfort Flo Humidification SystemTeleflexmedical, Inc. · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K913368Models MR700/MR720/MR730 Dual Servo Resp Humid AccFisher & Paykel Electronics, Ltd.1992—
−5K953392HC500 Servo-Controlled Heated Respiratory Humidifier (& Accessories)Fisher & Paykel Electronics, Ltd.1996—
−4K983112Respiratory Humidifier, Model # MR850Fisher & Paykel Healthcare, Ltd.19985
−3K020332Respiratory Humidifier, Model MR850Fisher & Paykel Healthcare, Ltd.20031
−2K033710MR850 Respiratory Humidifier, Model MR850JHUFisher & Paykel Healthcare, Ltd.2004—
−1K061736Comfort Flo Humidification System and Nasal CannulaTeleflex Medical2006—
0 (this device)K131912Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification System…Teleflex, Inc.2014—
+1K162242Comfort Flo Humidification SystemTeleflexmedical, Inc.2017Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199110
199210
199420
200020
200110
200310
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201711
FIG. 1 — Lineage of K131912, 1992–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTT device

  1. #1 K222197 F&P Optiflow Junior 2/2+ Nasal Cannula…score 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063758Choncha Therm Neptune Heated Humifier, Choncha Therm Neptune Heated Humifier…
Summary page 2
…E. Predicate Device This submission demonstrates substantial equivalence to the following predicate devices: * ConchaTherm Neptune Heated Humidifier - K063758 * Minimal Compliance Column - K993355 * Comfort Flo Humidification System - K061 736 F. Device Description Th…
Teleflex Medical20072
*K993355Hudson Rci Concha Column, Cat. NO. 385-40 and Hudson Rci Circuit, Cat. NO. 780-40…
Summary page 2
…onstrates substantial equivalence to the following predicate devices: * ConchaTherm Neptune Heated Humidifier - K063758 * Minimal Compliance Column - K993355 * Comfort Flo Humidification System - K061 736 F. Device Description The ConchaTherm Neptune Heated Humidifier is an…
Hudson Respiratory Care, Inc.20001
*K061736Comfort Flo Humidification System and Nasal Cannula
Summary page 2
…dicate devices: * ConchaTherm Neptune Heated Humidifier - K063758 * Minimal Compliance Column - K993355 * Comfort Flo Humidification System - K061 736 F. Device Description The ConchaTherm Neptune Heated Humidifier is an active heated humidifier designed to provide heat an…
Teleflex Medical2006—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162242Comfort Flo Humidification System Class ITeleflexmedical, Inc.20170

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 13, 2019Class IZ-1115-2020Device DesignThere is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
Dec 13, 2019Class IZ-1114-2020Device DesignThere is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
Apr 1, 2015Class IIZ-1558-2015Nonconforming Material/ComponentCustomer complaints were received of power to the unit ceasing and the display going blank.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162242Comfort Flo Humidification SystemTeleflexmedical, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 4.2 years on median.
  • 5 of the 16 ancestors have a recall of any class.
  • Of its descendants, K162242 (Teleflexmedical, Inc., 2017) has gone furthest, with 0 descendants of its own.

How does its category compare?

All BTT devices

BTT Humidifier, Respiratory Gas, (Direct Patient Interface): FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · BTT

4in 2025

2007: 62008: 42009: 12010: 42011: 32012: 22013: 62014: 72015: 12016: 32017: 32018: 22019: 12020: 22021: 22022: 32023: 72024: 72025: 42026 (partial year): 4
2007high 7 (2014) · low 1 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTT

104in 2025

2007: 73 days median2008: 199 days median2010: 113 days median2011: 116 days median2013: 171 days median2014: 231 days median2016: 157 days median2017: 303 days median2022: 290 days median2023: 258 days median2024: 219 days median2025: 104 days median2026 (partial year): 184 days median
2007high 303 (2017) · low 73 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTT

1in 2025

2007: 12008: 12010: 22011: 12014: 5, 4 Class I2025: 12026 (partial year): 1
2007high 5 (2014) · low 0 (2009)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · BTT

1,743in 2025

2007: 3652008: 4562009: 5582010: 5622011: 6652012: 5832013: 7182014: 7182015: 6702016: 1,0942017: 1,0072018: 9182019: 8942020: 1,1882021: 1,5102022: 7492023: 7162024: 5012025: 1,743
2007high 1,743 (2025) · low 365 (2007)2025

MAUDE reports naming product code BTT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076731365
200841991456
20091—0558
201041132562
201131161665
20122—0583
201361710718
201472315 (4)718
20151—0670
2016315701,094
2017330301,007
20182—0918
20191—0894
20202—01,188
20212—01,510
202232900749
202372580716
202472190501
2025410411,743
202641841—

What is it indicated for?

To provide a continuous flow of heated and humidified gas to spontaneously breathing patients. When used with the Hudson RCI ConchaTherm Neptune Heated Humidifier and Hudson RCI ventilator circuits, the ConchaSmart Column provides heated humidification for patients with and without an artificial airway in place. The ConchaTherm® Neptune® is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal…

Read the full statement (128 words)

To provide a continuous flow of heated and humidified gas to spontaneously breathing patients.

When used with the Hudson RCI ConchaTherm Neptune Heated Humidifier and Hudson RCI ventilator circuits, the ConchaSmart Column provides heated humidification for patients with and without an artificial airway in place.

The ConchaTherm® Neptune® is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits.

The ConchaTherm® Neptune® can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or non-flammable anesthesia gases to help maintain patient body temperature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131912: Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification System, Conchasmart Column.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131912

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026