510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Protura Couch Software

K132084510(k) TraditionalCleared Sep 26, 201383-day reviewClass II21 CFR 892.5770Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Med-Tec, Inc. D/B/A Civco Medical Solutions · JAI Couch, Radiation Therapy, Powered

Cleared on Sep 26, 2013 after an 83-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). It has been recalled 2 times.

Decided
Sep 26, 2013
Review
83 days
Class
Class II
Product code
JAI
Recalls
2
Predicates
1

The summary cites one number, which could not be confirmed as its predicate (see the table below).

FIG. 1 — Lineage of K132084Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAI device

  1. #1 K231612 Eve Patient Positioner Systemscore 7.9 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK1221Protura Couch Software
Number could not be read
Summary page 1
…09.70/A K1221SOFTWARE 8270AI MEDICA…

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 8, 2017Class IIZ-0671-2018OtherWhen an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the…
Mar 31, 2016Class IIZ-1992-2016Software designWhen using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.

No ancestor or descendant has a Class I recall.

How does its category compare?

All JAI devices

JAI Couch, Radiation Therapy, Powered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · JAI

1in 2024

2007: 22012: 42013: 12016: 12017: 12019: 12024: 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAI

61in 2012

2012: 61 days median
2007high 61 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAI

2in 2023

2007: 12008: 22009: 12010: 22013: 22017: 12023: 2
2007high 2 (2008) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAI

1in 2024

2007: 42008: 22009: 32012: 22013: 52014: 22015: 12016: 22017: 32019: 12024: 1
2007high 5 (2013) · low 0 (2010)2025

MAUDE reports naming product code JAI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—14
20080—22
20090—13
20100—20
20110—00
201246102
20131—25
20140—02
20150—01
20161—02
20171—13
20180—00
20191—01
20200—00
20210—00
20220—00
20230—20
20241—01
20250—00
20260—0—

What is it indicated for?

The Protura Couch Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K132084: Protura Couch Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K132084

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)