# K132320: Intellivue CL Respiration Pod

- Pathway: 510(k) (Traditional)
- Applicant: Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
- Decided: Mar 26, 2014 after 244 days
- Product code: BZQ (Monitor, Breathing Frequency)
- Device class: 2
- Regulation: 21 CFR 868.2375
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 33 generations back to K941811 (1994). 265 ancestors in all.
20 later devices descend from it; 1 cites it directly.
2 of 286 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K122223: New: Intellivue CL Respiration Pod New: Specialized Assessory: Mobile CL Resp Attachment Modified: Intellivue Patient MO, Philips Medizin Systeme Boeblingen GmbH (2013) Summary page 1.

## Most influential descendants
- K182543: Wearable Device ; Cradle ; Small Strap, Medium Strap, Large Strap ; Multiple Device Charger ; Hospital Starter Package (2018), 19 descendants of its own
- K190073: Wearable Device, Home Package in Carry Case with Homehub, Home Package in Carry Case without Homehub (2019), 18 descendants of its own
- K191272: Wearable Device, Wearable Package (2019), 16 descendants of its own
- K210133: Current Health Monitoring System Gen 2 (G2) (2021), 8 descendants of its own
- K221282: Empatica Health Monitoring Platform (2022), 7 descendants of its own

## Indications for use
The IntelliVue CL Respiration Pod is indicated for use by health care professionals whenever there is a need for intermittent or spot-check acquisition and monitoring of physiological patient parameters respiration rate and pulse rate wirelessly in specific hospital areas. The IntelliVue CL Respiration Pod is mainly indicated for use in general medical and surgery wards and in waiting areas of emergency rooms. It is not indicated for use in hospital areas in which continuous patient monitoring is needed, such as intensive care units or operating rooms. The intended use of the IntelliVue CL Respiration Pod when used together with a patient monitor is for intermittent or spot-check monitoring and recording of, and to generate alarms for, respiration rate and pulse rate of adult patients. The IntelliVue CL Respiration Pod is also intended for acquisition of respiration rate and pulse rate data of adult patients for a clinical information management system. The IntelliVue CL Respiration Pod is intended for use by health care professionals. It is not intended for home use. It is not a therapeutic device. The IntelliVue CL Respiration Pod is not intended for use on patients with extremely high values for respiration rate (above 60 rpm). The IntelliVue CL Respiration Pod is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias, e.g. very fast atrial fibrillation or ventricular tachycardia (rapid irregular pulse rate). For monitoring of these patients, a device for continuous ECG monitoring is necessary. The IntelliVue CL Respiration Pod is not a substitute for an ECG monitor. Warning: Do not use the CL Respiration Pod on patients with rapid, irregular heart rates greater than 110 bpm. Use under these conditions has not been clinically validated. Rx only: US Federal Law restricts these devices to sale by or on the order of a physician.

Source page: https://510kreport.com/device/K132320