Cardiovascular Pressure Measurement (CPM) System, Navvus Catheter

K132474510(k) TraditionalCleared Jan 30, 2014176-day reviewClass II21 CFR 870.2870

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Acist Medical Systems · DXO Transducer, Pressure, Catheter Tip

Cleared on Jan 30, 2014 as substantially equivalent to K062769 (Radi Medical Systems AB, 2006) after a 176-day review. Its family tree reaches back 6 generations to 2001. 10 later devices descend from it.

Decided
Jan 30, 2014
Review
176 days
Class
Class II
Product code
DXO
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K003122 (2001). 13 ancestors in all, the oldest K002067 (2001). 10 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2003: 2 other ancestors2002: 1 other ancestor2001: 3 other ancestors2019: 2 descendants cleared2015: 2 descendants cleared2016: 2 descendants cleared2024: 1 descendant cleared2017: 1 descendant cleared2018: 1 descendant cleared2023: 1 descendant clearedK003122 · Smartflow, Model Sfip 2000 · Florence Medical, Ltd. · 2001K012947 · Smartflow · Florence Medical, Ltd. · 2001K020127 · Smartflow Pulse Transmission Coefficient · Florence Medical, Ltd. · 2002K022188 · Radianalyzer, Model 12710 · Radi Medical Systems AB · 2002K031662 · Pressure Wire Sensor, Models 12004/12304 · Radi Medical Systems AB · 2003K062769 · Pressurewire Certus, Models 12006 and 12306 · Radi Medical Systems AB · 2006200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor4 generations back: 4 other ancestors5 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 3 descendants3 generations forward: 3 descendants4 generations forward: 1 descendant K062769 · Pressurewire Certus, Models 12006 and 12306 · Radi Medical Systems AB · 2006K031662 · Pressure Wire Sensor, Models 12004/12304 · Radi Medical Systems AB · 2003K022188 · Radianalyzer, Model 12710 · Radi Medical Systems AB · 2002K020127 · Smartflow Pulse Transmission Coefficient · Florence Medical, Ltd. · 2002K012947 · Smartflow · Florence Medical, Ltd. · 2001K003122 · Smartflow, Model Sfip 2000 · Florence Medical, Ltd. · 2001−50
  1. Root · 6 generations backK003122 Smartflow, Model Sfip 2000Florence Medical, Ltd. · 2001
  2. 5 generations backK012947 SmartflowFlorence Medical, Ltd. · 2001
  3. 4 generations backK020127 Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd. · 2002
  4. 3 generations backK022188 Radianalyzer, Model 12710Radi Medical Systems AB · 2002
  5. 2 generations backK031662 Pressure Wire Sensor, Models 12004/12304Radi Medical Systems AB · 2003
  6. 1 generation backK062769 Pressurewire Certus, Models 12006 and 12306Radi Medical Systems AB · 2006
  7. This deviceK132474 Cardiovascular Pressure Measurement (CPM) System, Navvus CatheterAcist Medical Systems · 2014
  8. Descendant · cites it directly · 4 of its ownK141275 Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2015
  9. Descendant · 2 generations on · 3 of its ownK160272 Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2016
  10. Descendant · cites it directly · 2 of its ownK141615 Endophys Blood Pressure MonitorEndophys, Inc. · 2015
  11. Descendant · cites it directly · 1 of its ownK190473 ACIST RXi System and Navvus II CatheterAcist Medical Systems, Inc. · 2019
  12. Descendant · 3 generations on · 1 of its ownK173399 Endophys Pressure Sensing Sheath Kit- 8FEndophys Holdings, LLC · 2017

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2015: 22016: 22017: 12018: 12019: 22023: 12024: 1
2015high 2 (2015) · low 0 (2020)2026

Most influential descendants

Showing 5 of 10 · See all
  • K141275Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2015 · cites it directly4descendants
  • K160272Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2016 · 2 generations on3descendants
  • K141615Endophys Blood Pressure MonitorEndophys, Inc. · 2015 · cites it directly2descendants
  • K190473ACIST RXi System and Navvus II CatheterAcist Medical Systems, Inc. · 2019 · cites it directly1descendant
  • K173399Endophys Pressure Sensing Sheath Kit- 8FEndophys Holdings, LLC · 2017 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K003122Smartflow, Model Sfip 2000Florence Medical, Ltd.2001—
−5K012947SmartflowFlorence Medical, Ltd.2001—
−4K020127Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd.2002—
−3K022188Radianalyzer, Model 12710Radi Medical Systems AB2002—
−2K031662Pressure Wire Sensor, Models 12004/12304Radi Medical Systems AB2003—
−1K062769Pressurewire Certus, Models 12006 and 12306Radi Medical Systems AB2006—
0 (this device)K132474Cardiovascular Pressure Measurement (CPM) System, Navvus CatheterAcist Medical Systems2014—
+1K141275Endophys Pressure Sensing Sheath KitEndophys, Inc.2015—
+2K160272Endophys Pressure Sensing Sheath KitEndophys, Inc.2016—
+1K141615Endophys Blood Pressure MonitorEndophys, Inc.2015—
+1K190473ACIST RXi System and Navvus II CatheterAcist Medical Systems, Inc.2019—
+3K173399Endophys Pressure Sensing Sheath Kit- 8FEndophys Holdings, LLC2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200130
200210
200320
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201520
201620
201710
201810
201920
202310
202410
FIG. 1 — Lineage of K132474, 2001–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXO device

  1. #1 K241418 OptoMonitor 3score 14.8 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062769Pressurewire Certus, Models 12006 and 12306
Summary page 5
…ire® ACIST Rapid Exchange (RXi) System and Characteristic (RADI Medical Systems, St. Jude avsCter Medical, Minnesota, USA) (K062769)NavsCter PressureWire is designed to fit inside a percutaneous catheter for the purpose of…
Radi Medical Systems AB2006—

Also cleared against K062769: 1 other device →

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141275Endophys Pressure Sensing Sheath KitEndophys, Inc.20154
K141615Endophys Blood Pressure MonitorEndophys, Inc.20152
K190473ACIST RXi System and Navvus II CatheterAcist Medical Systems, Inc.20191

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 8 months on median.
  • None of the 13 ancestors has a recall.
  • Of its descendants, K141275 (Endophys, Inc., 2015) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K062769.

How does its category compare?

All DXO devices

DXO Transducer, Pressure, Catheter Tip: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · DXO

1in 2025

2007: 12008: 12010: 12011: 12012: 12013: 12014: 32015: 22016: 22017: 22018: 12019: 42020: 22022: 12024: 12025: 12026 (partial year): 1
2007high 4 (2019) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXO

120in 2019

2014: 176 days median2019: 120 days median
2007high 176 (2014) · low 120 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXO

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXO

183in 2025

2007: 32008: 22009: 82010: 742011: 1112012: 822013: 932014: 852015: 1142016: 1522017: 1862018: 1792019: 1532020: 1582021: 1162022: 1442023: 1942024: 1972025: 183
2007high 197 (2024) · low 2 (2008)2025

MAUDE reports naming product code DXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20081—02
20090—08
20101—074
20111—0111
20121—082
20131—093
20143176085
20152—1114
20162—0152
20172—0186
20181—0179
201941200153
20202—0158
20210—0116
20221—0144
20230—0194
20241—0197
20251—0183
20261—0—

What is it indicated for?

The ACIST Rapid Exchange (RXI) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K132474: Cardiovascular Pressure Measurement (CPM) System, Navvus Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K132474

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.