510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Modified HD Guide Catheter

K133177510(k) TraditionalCleared Feb 25, 2014131-day reviewClass II21 CFR 870.1250Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Concentric Medical, Inc. · DQY Catheter, Percutaneous

Cleared on Feb 25, 2014 as substantially equivalent to K090335 (Concentric Medical, Inc., 2009) and 2 other predicates after a 131-day review. Its family tree reaches back 4 generations to 2001. 28 later devices descend from it. It has been recalled 3 times.

Decided
Feb 25, 2014
Review
131 days
Class
Class II
Product code
DQY
Recalls
3
Predicates
3

Its longest line of descent runs back 4 generations to K003880 (2001). 8 ancestors in all. 28 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2010: 1 other ancestor2002: 1 other ancestor2001: 1 other ancestor2015: 1 descendant cleared2024: 7 descendants cleared2018: 2 descendants cleared2019: 3 descendants cleared2023: 3 descendants cleared2022: 1 descendant cleared2026: 4 descendants cleared2025: 4 descendants cleared2021: 1 descendant cleared2020: 2 descendants clearedK003880 · Concentroc HD Guide Catheter, Models 90017 (4F), 90019 (5F), 90021 (6F), 90022 (7F), 90023 (8F), 90024 (9F) · Concentric Medical, Inc. · 2001K080583 · Modified HD Guide Catheter · Concentric Medical, Inc. · 2008K090335 · HD Guide Catheter · Concentric Medical, Inc. · 2009K110483 · Modified HD Guide Catheter · Concentric Medical, Inc. · 2011 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 9 descendants4 generations forward: 5 descendants5 generations forward: 6 descendants6 generations forward: 1 descendant K110483 · Modified HD Guide Catheter · Concentric Medical, Inc. · 2011 · recalledK090335 · HD Guide Catheter · Concentric Medical, Inc. · 2009K080583 · Modified HD Guide Catheter · Concentric Medical, Inc. · 2008K003880 · Concentroc HD Guide Catheter, Models 90017 (4F), 90019 (5F), 90021 (6F), 90022 (7F), 90023 (8F), 90024 (9F) · Concentric Medical, Inc. · 20010+5
  1. Root · 4 generations backK003880 Concentroc HD Guide Catheter, Models 90017 (4F), 90019 (5F), 90021…Concentric Medical, Inc. · 2001
  2. 3 generations backK080583 Modified HD Guide CatheterConcentric Medical, Inc. · 2008
  3. 2 generations backK090335 HD Guide CatheterConcentric Medical, Inc. · 2009
  4. 1 generation backK110483 Modified HD Guide CatheterConcentric Medical, Inc. · 2011 · Recalled
  5. This deviceK133177 Modified HD Guide CatheterConcentric Medical, Inc. · 2014
  6. Descendant · cites it directly · 26 of its ownK151667 AXS Catalyst Distal Access Catheter 058x115cm, AXS Catalyst Distal…Stryker · 2015 · Recalled
  7. Descendant · 2 generations on · 19 of its ownK180715 React 68 CatheterMicro Therapeutics, Inc. d/b/a ev3 Neurovascular · 2018
  8. Descendant · 2 generations on · 14 of its ownK183463 AXS Catalyst Distal Access Catheter 068 x 115cm, AXS Catalyst Distal…Stryker · 2019
  9. Descendant · 3 generations on · 6 of its ownK190833 AXS Vecta Aspiration CatheterStryker Neurovascular · 2019
  10. Descendant · 3 generations on · 5 of its ownK191237 CERENOVUS Large Bore CatheterCodman & Shurtleff, Inc. · 2019

Showing 5 of 28 descendants. See all descendants

Descendants cleared per year

2015: 12018: 22019: 32020: 22021: 12022: 12023: 32024: 72025: 42026: 4
2015high 7 (2024) · low 0 (2016)2026

Most influential descendants

Showing 5 of 28 · See all
  • K151667AXS Catalyst Distal Access Catheter 058x115cm, AXS Catalyst Distal Access Catheter…Stryker · 2015 · cites it directly26descendants
  • K180715React 68 CatheterMicro Therapeutics, Inc. d/b/a ev3 Neurovascular · 2018 · 2 generations on19descendants
  • K183463AXS Catalyst Distal Access Catheter 068 x 115cm, AXS Catalyst Distal Access Catheter 068…Stryker · 2019 · 2 generations on14descendants
  • K190833AXS Vecta Aspiration CatheterStryker Neurovascular · 2019 · 3 generations on6descendants
  • K191237CERENOVUS Large Bore CatheterCodman & Shurtleff, Inc. · 2019 · 3 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K003880Concentroc HD Guide Catheter, Models 90017 (4F), 90019 (5F), 90021 (6F), 90022…Concentric Medical, Inc.2001—
−3K080583Modified HD Guide CatheterConcentric Medical, Inc.2008—
−2K090335HD Guide CatheterConcentric Medical, Inc.2009—
−1K110483Modified HD Guide CatheterConcentric Medical, Inc.20111
0 (this device)K133177Modified HD Guide CatheterConcentric Medical, Inc.2014—
+1K151667AXS Catalyst Distal Access Catheter 058x115cm, AXS Catalyst Distal Access…Stryker20153
+2K180715React 68 CatheterMicro Therapeutics, Inc. d/b/a ev3 Neurovascular2018—
+2K183463AXS Catalyst Distal Access Catheter 068 x 115cm, AXS Catalyst Distal Access…Stryker2019—
+3K190833AXS Vecta Aspiration CatheterStryker Neurovascular2019—
+3K191237CERENOVUS Large Bore CatheterCodman & Shurtleff, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200210
201010
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201820
201930
202020
202110
202210
202330
202470
202540
202640
FIG. 1 — Lineage of K133177, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090335HD Guide Catheter
Summary page 1
…roduct Code: DQY and OQO Legally Marketed Reference Device Predicate Devices: (Clearance Date) K090335 (May 6, 2009) Concentric HD Guide Catheter KI 10483 (April 4, 2011) Modified H…
Concentric Medical, Inc.2009—
*K110483Modified HD Guide Catheter
Summary page 2
…hnological Characteristics and Product Feature Comparison: The subject device has the same technological characteristics as the predicate devices (K090335 and K1 10483). The device design, materials, fundamental scientific technology, materials and processes for packaging and sterilization h…
Concentric Medical, Inc.20111
*K112404Concentric Balloon Guide Catheter
Summary page 2
…aging and sterilization have not been changed from the previous predicate devices (K090335 and KI 10483). The subject device differs from the predicate device, K1 12404 primarily in that it does not include a balloon at the distal tip or a second lumen. A tabular comparison of the speci…
Concentric Medical, Inc.20121

Also cleared against K090335: 3 other devices →

What descends from it? (28)

See all 28

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151667AXS Catalyst Distal Access Catheter 058x115cm, AXS Catalyst Distal Access Catheter…Stryker201526
K242301Socrates 38 CatheterScientia Vascular, Inc.20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 23, 2024Class IIZ-1908-2024Process designNeurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin…
Apr 3, 2024Class IIZ-1740-2024Release of Material/Component prior to receiving test resultsStryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.
Aug 19, 2015Class IIZ-0116-2016Employee errorOne Distal Access Catheter intended for distribution in Korea, was found at a US location.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 1 company; a new generation was cleared every 2.9 years on median.
  • 4 of the 8 ancestors have a recall of any class.
  • Of its descendants, K151667 (Stryker, 2015) has gone furthest, with 26 descendants of its own.
  • 3 other devices also cite its predicate K090335.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The Concentric HID Guide Catheter is indicated for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K133177: Modified HD Guide Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133177

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026