510(k) Report

510(k) · Cleared

Recalled · Class I · 2014

Diamondback 360 Peripheral Orbital Atherectomy System

K133399510(k) SpecialCleared Nov 22, 201316-day reviewClass II21 CFR 870.4875Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardiovascular Systems Incorporated · MCW Catheter, Peripheral, Atherectomy

Cleared on Nov 22, 2013 as substantially equivalent to K110389 (Cardiovascular Systems, Inc., 2011) and 2 other predicates after a 16-day review. Its family tree reaches back 10 generations to 1987. 1 of the 23 devices in its family, counting itself, has a Class I recall. 4 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Nov 22, 2013
Review
16 days
Class
Class II
Product code
MCW
Recalls
1 (1 Class I)
Predicates
3

Its longest line of descent runs back 10 generations to K871011 (1987). 18 ancestors in all. 4 later devices descend from it, 3 citing it directly. 1 of the 23 devices in this family, counting this one, has a Class I recall.

  1. Root · 10 generations backK871011 Simpson Peripheral Atherocath (TM)Devices For Vascular Intervention, Inc. · 1987
  2. 6 generations between
  3. 3 generations backK110389 Stealth 360 Orbital Pad SystemCardiovascular Systems, Inc. · 2011 · Recalled
  4. 2 generations backK122987 Stealth 360 Orbital Pad SystemCardiovascular Systems, Inc. · 2012
  5. 1 generation backK131092 Stealth 360 Orbital Pad SystemCardio Systems, Inc. · 2013
  6. This deviceK133399 Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems Incorporated · 2013
  7. Descendant · cites it directly · 2 of its ownK150732 Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc. · 2015
  8. Descendant · cites it directly · 1 of its ownK151260 ViperWire Advance Guide Wire, Diamondback 360 Peripheral Orbital…Cardiovascular System, Inc. · 2015 · Recalled
  9. Descendant · 2 generations on · 0 of its ownK180416 ViperWire Advance Guide Wire, Stealth 360 Orbital Peripheral…Cardiovascular Systems, Inc. · 2018
  10. Descendant · cites it directly · 0 of its ownK133999 Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc. · 2014
Show all 10 generations
  1. Root · 10 generations backK871011 Simpson Peripheral Atherocath (TM)Devices For Vascular Intervention, Inc. · 1987
  2. 9 generations backK024243 Reform Peripheral Catheter System, Model 02200; Reform Peripheral…Fox Hollow Technologies · 2003 · Recalled
  3. 8 generations backK043553 Silverhawk Peripheral Plaque Excision System, Models 02550,04800…Fox Hollow Technologies · 2005
  4. 7 generations backK071350 Diamondback 360 Degree Orbital Atherectomy SystemCardiovascular Systems, Inc. · 2007
  5. 6 generations backK072748 Modification to: Diamondback 360 Orbital Atherectomy SystemCardiovascular Systems, Inc. · 2007
  6. 5 generations backK081944 Modification to Diamondback 360 Degree Orbital Atherectomy SystemCardio Systems, Inc. · 2008
  7. 4 generations backK090521 Diamondback 360 DBXCardiovascular Systems, Inc. · 2009 · Recalled
  8. 3 generations backK110389 Stealth 360 Orbital Pad SystemCardiovascular Systems, Inc. · 2011 · Recalled
  9. 2 generations backK122987 Stealth 360 Orbital Pad SystemCardiovascular Systems, Inc. · 2012
  10. 1 generation backK131092 Stealth 360 Orbital Pad SystemCardio Systems, Inc. · 2013
  11. This deviceK133399 Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems Incorporated · 2013

Most influential descendants

Showing 4 of 4 · See all
  • K150732Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc. · 2015 · cites it directly2descendants
  • K151260ViperWire Advance Guide Wire, Diamondback 360 Peripheral Orbital Atherectomy System…Cardiovascular System, Inc. · 2015 · cites it directly1descendant
  • K180416ViperWire Advance Guide Wire, Stealth 360 Orbital Peripheral Atherectomy Device (PAD)…Cardiovascular Systems, Inc. · 2018 · 2 generations on0descendants
  • K133999Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K871011Simpson Peripheral Atherocath (TM)Devices For Vascular Intervention, Inc.1987—
−9K024243Reform Peripheral Catheter System, Model 02200; Reform Peripheral…Fox Hollow Technologies20031
−8K043553Silverhawk Peripheral Plaque Excision System, Models 02550,04800, 05200…Fox Hollow Technologies2005—
−7K071350Diamondback 360 Degree Orbital Atherectomy SystemCardiovascular Systems, Inc.2007—
−6K072748Modification to: Diamondback 360 Orbital Atherectomy SystemCardiovascular Systems, Inc.2007—
−5K081944Modification to Diamondback 360 Degree Orbital Atherectomy SystemCardio Systems, Inc.2008—
−4K090521Diamondback 360 DBXCardiovascular Systems, Inc.20092
−3K110389Stealth 360 Orbital Pad SystemCardiovascular Systems, Inc.20111
−2K122987Stealth 360 Orbital Pad SystemCardiovascular Systems, Inc.2012—
−1K131092Stealth 360 Orbital Pad SystemCardio Systems, Inc.2013—
0 (this device)K133399Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems Incorporated2013—
+1K150732Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc.2015—
+1K151260ViperWire Advance Guide Wire, Diamondback 360 Peripheral Orbital Atherectomy…Cardiovascular System, Inc.20152
+2K180416ViperWire Advance Guide Wire, Stealth 360 Orbital Peripheral Atherectomy…Cardiovascular Systems, Inc.2018—
+1K133999Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198830
199010
199720
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201520
201810
FIG. 1 — Lineage of K133399, 1987–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MCW device

  1. #1 K233668 Auryon Atherectomy Catheter 1.5 mm XL…score 17.6 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K110389Stealth 360 Orbital Pad System
Summary page 1
…ember 05, 2013 Diamondback 360 Peripheral Orbital Atherectomy System MCW—Catheter, Peripheral, Atherectomy 21 CFR 870.4875—Intraluminal Artery Stripper 510(k) Number: K110389, K122987, & K131092 Manufacturer: Cardiovascular Systems, Inc. Trade Name: Stealth 360°® Orbital PAD System 6.1 Description of D…
Cardiovascular Systems, Inc.20111
*K122987Stealth 360 Orbital Pad System
Summary page 1
…2013 Diamondback 360 Peripheral Orbital Atherectomy System MCW—Catheter, Peripheral, Atherectomy 21 CFR 870.4875—Intraluminal Artery Stripper 510(k) Number: K110389, K122987, & K131092 Manufacturer: Cardiovascular Systems, Inc. Trade Name: Stealth 360°® Orbital PAD System 6.1 Description of Device Th…
Cardiovascular Systems, Inc.2012—
*K131092Stealth 360 Orbital Pad System
Summary page 1
…ondback 360 Peripheral Orbital Atherectomy System MCW—Catheter, Peripheral, Atherectomy 21 CFR 870.4875—Intraluminal Artery Stripper 510(k) Number: K110389, K122987, & K131092 Manufacturer: Cardiovascular Systems, Inc. Trade Name: Stealth 360°® Orbital PAD System 6.1 Description of Device The Diamondba…
Cardio Systems, Inc.2013—
†K090521Diamondback 360 DBXCardiovascular Systems, Inc.20092

Also cleared against K110389: 8 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150732Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc.20152
K151260ViperWire Advance Guide Wire, Diamondback 360 Peripheral Orbital Atherectomy System…Cardiovascular System, Inc.20151
K133999Diamondback 360 Peripheral Orbital Atherectomy SystemCardiovascular Systems, Inc.20140

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 23, 2014Class IZ-2155-2014Process controlCSI has initiated a recall on the Diamondback 360 Peripheral Orbital Atherectomy Device because it may contain a saline sheath that may experience cracking, fracture, and release particulate during use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 5 companies; a new generation was cleared every 1.6 years on median.
  • 4 of the 18 ancestors have a recall of any class.
  • Of its descendants, K150732 (Cardiovascular Systems, Inc., 2015) has gone furthest, with 2 descendants of its own.
  • 8 other devices also cite its predicate K110389.

How does its category compare?

All MCW devices

MCW Catheter, Peripheral, Atherectomy: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · MCW

3in 2025

2007: 52008: 62009: 82010: 62011: 62012: 52013: 52014: 72015: 62016: 52017: 72018: 82019: 32020: 22021: 62022: 62023: 42024: 32025: 32026 (partial year): 1
2007high 8 (2009) · low 2 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MCW

46in 2025

2007: 43 days median2008: 33 days median2009: 27 days median2010: 30 days median2011: 39 days median2012: 29 days median2013: 30 days median2014: 117 days median2015: 30 days median2016: 60 days median2017: 91 days median2018: 103 days median2019: 112 days median2021: 119 days median2022: 69 days median2023: 104 days median2024: 41 days median2025: 46 days median
2007high 119 (2021) · low 27 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MCW

2in 2024

2007: 22008: 12009: 32010: 12016: 22017: 22021: 2, 2 Class I2022: 2, 2 Class I2023: 12024: 2
2007high 3 (2009) · low 0 (2011)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MCW

677in 2025

2007: 502008: 1152009: 1052010: 1672011: 2042012: 1942013: 1342014: 2102015: 3772016: 4092017: 4152018: 5372019: 6182020: 6312021: 7342022: 5662023: 5742024: 6772025: 677
2007high 734 (2021) · low 50 (2007)2025

MAUDE reports naming product code MCW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007543250
20086331115
20098273105
20106301167
20116390204
20125290194
20135300134
201471170210
20156300377
20165602409
20177912415
201881030537
201931120618
20202—0631
202161192 (2)734
20226692 (2)566
202341041574
20243412677
20253460677
20261—0—

What is it indicated for?

The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K133399: Diamondback 360 Peripheral Orbital Atherectomy System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K133399

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026