# K133807: Mars Monitor Itc US, Mars Treatment Kit Type 1116/1 - X-Mars US, Prismaflex Control Unit

- Pathway: 510(k) (Traditional)
- Applicant: Gambro Renal Products, Inc.
- Decided: Mar 26, 2014 after 100 days
- Product code: FLD (Apparatus, Hemoperfusion, Sorbent)
- Device class: 2
- Regulation: 21 CFR 876.5870
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA’s file for it is a 510(k) statement, which names no predicate.

## Recalls of this device
- Dec 21, 2017 · Class II · Z-0411-2018: Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set

## Indications for use
The MARS is indicated for the treatment of drug overdose and poisonings. The only requirement is that the drug or chemical be dialyzable (in unbound form) and bound by charcoal and/or ion exchange resins. The MARS is indicated in the treatment of Hepatic Encephalopathy (HE) due to a de-compensation of a chronic liver disease. Clinical trials conducted with MARS treatments in HE patients having a decompensation of chronic liver disease demonstrated a transient effect from MARS treatments to significantly decrease their hepatic encephalopathy scores by at least 2 grades compared to standard medical therapy (SMT). Contraindication The MARS' is not indicated as a bridge to liver transplant. Safety and efficacy hats not been demonstrated for this indication in controlled, randomized clinical trials. The effectiveness of the MARS device in patients that are sedated could not be established in clinical studies and therefore cannot be predicted in sedated patients. The Prismaflex control unit is intended for: * Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. * Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. * Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.