510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Saber Percutaneous Transluminal Angioplasty Dilatation Catheter

K133843510(k) TraditionalCleared Jun 27, 2014191-day reviewClass II21 CFR 870.1250Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cordis Corporation · LIT Catheter, Angioplasty, Peripheral, Transluminal

Cleared on Jun 27, 2014 as substantially equivalent to K971010 (Cordis Corp., 1997) and 2 other predicates after a 191-day review. Its family tree reaches back 8 generations to 2000. 6 later devices descend from it. It has been recalled once.

Decided
Jun 27, 2014
Review
191 days
Class
Class II
Product code
LIT
Recalls
1
Predicates
3

Its longest line of descent runs back 8 generations to K001134 (2000). 27 ancestors in all, the oldest K971010 (1997). 6 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2010: 2 other ancestors1997: 2 other ancestors2009: 1 other ancestor2004: 2 other ancestors2008: 1 other ancestor2006: 1 other ancestor2005: 4 other ancestors2003: 1 other ancestor2001: 3 other ancestors2000: 1 other ancestor1998: 1 other ancestor2015: 1 descendant cleared2022: 1 descendant cleared2017: 1 descendant cleared2016: 2 descendants cleared2019: 1 descendant clearedK001134 · Gazelle Balloon Dilatation Catheter · Boston Scientific Corp · 2000K021735 · Boston Scientific Ultra-Soft SV Balloon Dilatation Catheter · Boston Scientific Scimed, Inc. · 2002 · recalledK050389 · Ultra-Soft SV Balloon Dilatation Catheter · Boston Scientific Corp · 2005K053118 · Sterling Monorail Pta Balloon Dilatation Catheter, Model H74939031-Xxxxxx · Boston Scientific Corp · 2005×2: 2 generations cleared close together, K063563 (2007), K071189 (2007) ×2: 2 generations cleared close together, K112797 (2012), K121442 (2012) 1995200020052010201520202025
  1. Root · 8 generations backK001134 Gazelle Balloon Dilatation CatheterBoston Scientific Corp · 2000
  2. 7 generations backK021735 Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc. · 2002 · Recalled
  3. 6 generations backK050389 Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp · 2005
  4. 5 generations backK053118 Sterling Monorail Pta Balloon Dilatation Catheter, Model…Boston Scientific Corp · 2005
  5. 4 generations backK063563 Cordis Amila Percutaneous Transluminal Angioplasty (Pta) Balloon…Cordis Europa, N.V. · 2007
  6. 3 generations backK071189 Cordis Aviator Plus Pta Balloon Dilatation CatheterCordis Europa, N.V. · 2007
  7. 2 generations backK112797 Powerflex Pro Percutaneous Transluminal Angioplasty CatheterCordis Corp. · 2012
  8. 1 generation backK121442 Powerflex Pro Percutaneous Transluminal Angioplasty CatheterCordis Corporation · 2012 · Recalled
  9. This deviceK133843 Saber Percutaneous Transluminal Angioplasty Dilatation CatheterCordis Corporation · 2014
  10. Descendant · cites it directly · 4 of its ownK151153 Pirouette 018Arravasc Limited · 2015
  11. Descendant · 2 generations on · 2 of its ownK162316 Pirouette 014Arravasc, Ltd. · 2016
  12. Descendant · 2 generations on · 2 of its ownK161427 Pirouette 035Arravasc Limited · 2016
  13. Descendant · 2 generations on · 1 of its ownK173621 Advocate PTA CatheterMerit Medical Systems, Inc. · 2017
  14. Descendant · cites it directly · 0 of its ownK221245 PTA Balloon Dilatation CatheterKossel Medtech (Suzhou) Co., Ltd. · 2022

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2015: 12016: 22017: 12019: 12022: 1
2015high 2 (2016) · low 0 (2018)2026

Most influential descendants

Showing 5 of 6 · See all
  • K151153Pirouette 018Arravasc Limited · 2015 · cites it directly4descendants
  • K162316Pirouette 014Arravasc, Ltd. · 2016 · 2 generations on2descendants
  • K161427Pirouette 035Arravasc Limited · 2016 · 2 generations on2descendants
  • K173621Advocate PTA CatheterMerit Medical Systems, Inc. · 2017 · 2 generations on1descendant
  • K221245PTA Balloon Dilatation CatheterKossel Medtech (Suzhou) Co., Ltd. · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K001134Gazelle Balloon Dilatation CatheterBoston Scientific Corp2000—
−7K021735Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc.200223
−6K050389Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp2005—
−5K053118Sterling Monorail Pta Balloon Dilatation Catheter, Model H74939031-XxxxxxBoston Scientific Corp2005—
−4K063563Cordis Amila Percutaneous Transluminal Angioplasty (Pta) Balloon Dilatation…Cordis Europa, N.V.2007—
−3K071189Cordis Aviator Plus Pta Balloon Dilatation CatheterCordis Europa, N.V.2007—
−2K112797Powerflex Pro Percutaneous Transluminal Angioplasty CatheterCordis Corp.2012—
−1K121442Powerflex Pro Percutaneous Transluminal Angioplasty CatheterCordis Corporation20122
0 (this device)K133843Saber Percutaneous Transluminal Angioplasty Dilatation CatheterCordis Corporation2014—
+1K151153Pirouette 018Arravasc Limited2015—
+2K162316Pirouette 014Arravasc, Ltd.2016—
+2K161427Pirouette 035Arravasc Limited2016—
+2K173621Advocate PTA CatheterMerit Medical Systems, Inc.2017—
+1K221245PTA Balloon Dilatation CatheterKossel Medtech (Suzhou) Co., Ltd.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199720
199810
200010
200130
200310
200420
200540
200610
200810
200910
201020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201620
201710
201910
202210
FIG. 1 — Lineage of K133843, 2000–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LIT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LIT device

  1. #1 K241040 Amethyst HP PTA OTW 0.035 Catheterscore 13.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K971010Peripheral Transluminal Angioplasty Catheters
Summary page 1
…-Predicate Devices 510(k) Date Device Name Manufacturer Number Cleared . K971010 6/18/1997 SAVVY 4 PTA Balloon Catheter Cordis Corp.…
Cordis Corp.1997—
*K121442Powerflex Pro Percutaneous Transluminal Angioplasty Catheter
Summary page 2
…97 SAVVY 4 PTA Balloon Catheter Cordis Corp. K133843 Page 2 of 5 K121442 6/14/2012 Powerfiex® Pro Percutaneous Cordis Corp. Transluninal Angioplasty…
Cordis Corporation20122
*K103464Pacific Xtreme Ptca Balloon Dilatation Catheter
Summary page 2
…iex® Pro Percutaneous Cordis Corp. Transluninal Angioplasty Balloon Catheter TM K 103464 12/22/2010 Invatec PacifiC Xtremne PTA Medtronic Inc Balloon Catheter D. Device Description: The…
Medtronic, Inc.20101

Also cleared against K971010: 8 other devices →

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151153Pirouette 018Arravasc Limited20154
K221245PTA Balloon Dilatation CatheterKossel Medtech (Suzhou) Co., Ltd.20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 19, 2021Class IIZ-2235-2021Under Investigation by firmThe balloon protective sheath may potentially be contaminated with a foreign material left over from the manufacturing process. The contaminant may transfer over from the sheath to the balloon.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 5 companies; a new generation was cleared every 1.9 years on median.
  • 8 of the 27 ancestors have a recall of any class.
  • Of its descendants, K151153 (Arravasc Limited, 2015) has gone furthest, with 4 descendants of its own.
  • 8 other devices also cite its predicate K971010.

How does its category compare?

All LIT devices

LIT Catheter, Angioplasty, Peripheral, Transluminal: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LIT

3in 2025

2007: 62008: 82009: 72010: 132011: 102012: 192013: 142014: 122015: 192016: 112017: 92018: 62019: 42020: 92021: 42022: 82023: 62024: 62025: 32026 (partial year): 4
2007high 19 (2012) · low 3 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LIT

90in 2025

2007: 64 days median2008: 53 days median2009: 29 days median2010: 78 days median2011: 82 days median2012: 29 days median2013: 79 days median2014: 90 days median2015: 112 days median2016: 116 days median2017: 57 days median2018: 118 days median2019: 40 days median2020: 125 days median2021: 61 days median2022: 122 days median2023: 167 days median2024: 43 days median2025: 90 days median2026 (partial year): 205 days median
2007high 205 (2026) · low 29 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LIT

1in 2025

2009: 22010: 12013: 2, 2 Class I2014: 32016: 12018: 22019: 32021: 12024: 22025: 1
2007high 3 (2014) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LIT

2,073in 2025

2007: 5772008: 5132009: 6232010: 8092011: 8572012: 1,2772013: 1,2422014: 1,1782015: 1,2792016: 1,5292017: 2,0812018: 2,1882019: 2,4052020: 2,4432021: 2,4922022: 2,1552023: 1,8072024: 1,9132025: 2,073
2007high 2,492 (2021) · low 513 (2008)2025

MAUDE reports naming product code LIT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076640577
20088530513
20097292623
201013781809
201110820857
2012192901,277
201314792 (2)1,242
2014129031,178
20151911201,279
20161111611,529
201795702,081
2018611822,188
201944032,405
2020912502,443
202146112,492
2022812202,155
2023616701,807
202464321,913
202539012,073
202642050—

What is it indicated for?

The SABER PTA Dilatation catheter is intended to dilate stenoses in iliac, femnoral, ii-femoral, popliteal, infra papliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device is also indicated for post-dilation of balloon-expandable and self-expanding stents in the peripheral vasculature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133843: Saber Percutaneous Transluminal Angioplasty Dilatation Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133843

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026