# K133898: Niox Vero Airway Inflammation Monitor

- Pathway: 510(k) (Traditional)
- Applicant: Aerocrine AB
- Decided: Nov 6, 2014 after 321 days
- Product code: MXA (System, Test, Breath Nitric Oxide)
- Device class: 2
- Regulation: 21 CFR 862.3080
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 3 generations back to DEN030001 (2003). 3 ancestors in all.
3 later devices descend from it; 1 cites it directly.
0 of 7 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K101034: Niox Mino Model 09-1000, Aerocrine AB (2010) Summary page 4.

## Most influential descendants
- K150233: NIOX VERO Airway Inflammation Monitor (2015), 2 descendants of its own
- K170983: Niox Vero (2017), 1 descendant of its own
- K203695: NObreath® (2021), 0 descendants of its own

## Indications for use
NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society. Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, approximately 7 - 17 years, and adults 18 years and older. FeNO measurements provide the physician with means of evaluating an asthma patient’s response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX VERO should not be used in critical care, emergency care or in anesthesiology.