# K140110: Flextip Plus Epidural Catheter;Flexblock Continous Peripheral Nerve Block Catheter Kit/Set

- Pathway: 510(k) (Traditional)
- Applicant: Teleflex, Inc.
- Decided: Mar 27, 2015 after 436 days
- Product code: BSO (Catheter, Conduction, Anesthetic)
- Device class: 2
- Regulation: 21 CFR 868.5120
- Recalls of this device: 9 (1 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K884552 (1989). 2 ancestors in all.
1 later device descends from it; 1 cites it directly.
3 of 4 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K103658: Flextip Plus Closed Tip Epidural Catherter, Teleflex, Inc. (2012) Summary page 5.

## Most influential descendants
- K181782: Medline Reinforced Epidural Catheter (2019), 0 descendants of its own

## Recalls of this device
- Apr 21, 2026 · Class I · Z-2371-2026: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffect
- Mar 11, 2026 · Class II · Z-1913-2026: Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured
- May 8, 2020 · Class II · Z-2288-2020: Nine complaints reported that the rotatable collar on the filter had detached, causing a leak.
- Jan 15, 2020 · Class II · Z-1174-2020: There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Jan 15, 2020 · Class II · Z-1177-2020: There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Jan 15, 2020 · Class II · Z-1172-2020: There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Dec 20, 2019 · Class II · Z-1237-2020: Product lidstock contains the incorrect expiration date for the product
- Dec 20, 2019 · Class II · Z-1238-2020: Product lidstock contains the incorrect expiration date for the product
- Dec 20, 2019 · Class II · Z-1236-2020: Product lidstock contains the incorrect expiration date for the product

## Indications for use
The Arrow Epidural Catheter permits access to the epidural space for the administration of epidural anesthetic. The epidural catheter is intended for use up to 72 hours.