510(k) Report

510(k) · Cleared

Recalled · Class III · 2023

Randox Immunoassay Premium Plus Control Levels 1,2 and Level 3 and Immunoassay Premium Plus Tri-Level Control

K140522510(k) TraditionalCleared May 22, 201479-day reviewClass I21 CFR 862.1660Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Randox Laboratories Limited · JJY Multi-Analyte Controls, All Kinds (Assayed)

Cleared on May 22, 2014 as substantially equivalent to K040379 (Randox Laboratories, Ltd., 2004) after a 79-day review. Its family tree reaches back 1 generation to 2004. It has been recalled 5 times.

Decided
May 22, 2014
Review
79 days
Class
Class I
Product code
JJY
Recalls
5
Predicates
1

Its longest line of descent runs back 1 generation to K040379 (2004). No later device cites it as a predicate yet.

K040379 · Randox Immunoassay Control · Randox Laboratories, Ltd. · 2004 · recalled2000200520102015
  1. Root · 1 generation backK040379 Randox Immunoassay ControlRandox Laboratories, Ltd. · 2004 · Recalled
  2. This deviceK140522 Randox Immunoassay Premium Plus Control Levels 1,2 and Level 3 and…Randox Laboratories Limited · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K040379Randox Immunoassay ControlRandox Laboratories, Ltd.20041
0 (this device)K140522Randox Immunoassay Premium Plus Control Levels 1,2 and Level 3 and Immunoassay…Randox Laboratories Limited2014—
FIG. 1 — Lineage of K140522, 2004–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040379Randox Immunoassay Control
Summary page 2
…4. PREDICATE DEVICE PROPRIETARY NAMES AND 510 (k) NUMBERS Predicate Device Proprietary Name: Randox Immunoassay Controls Levels 1, 2 and 3 510 (k)Numbers: K040379 5. INTENDED USE This product is intended for in vitro diagnostic use as assayed quality control material to monit…
Randox Laboratories, Ltd.20041

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 4, 2023Class IIIZ-2022-2023Nonconforming Material/ComponentDelay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has…
May 4, 2023Class IIIZ-2021-2023Nonconforming Material/ComponentDelay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has…
Feb 7, 2019Class IIIZ-1106-2019Employee errorThe target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
Aug 11, 2017Class IIIZ-0970-2018Device DesignThe ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.
Aug 11, 2017Class IIIZ-0969-2018Device DesignThe ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.

How does its category compare?

All JJY devices

JJY Multi-Analyte Controls, All Kinds (Assayed): FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · JJY

3in 2017

2007: 112008: 72009: 82010: 102011: 142012: 132013: 162014: 102015: 42016: 12017: 3
2007high 16 (2013) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JJY

78in 2017

2007: 49 days median2008: 54 days median2009: 80 days median2010: 109 days median2011: 120 days median2012: 40 days median2013: 82 days median2014: 49 days median2015: 153 days median2017: 78 days median
2007high 153 (2015) · low 40 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JJY

2in 2025

2007: 52009: 22010: 62011: 12012: 32013: 22014: 22015: 12016: 32017: 42018: 32019: 42020: 62022: 32023: 32025: 22026 (partial year): 1
2007high 6 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJY

13in 2025

2007: 22008: 12009: 32010: 32011: 62012: 202013: 52014: 42015: 42016: 132017: 122018: 72019: 12020: 72021: 12023: 42024: 12025: 13
2007high 20 (2012) · low 0 (2022)2025

MAUDE reports naming product code JJY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007114952
200875401
200988023
20101010963
20111412016
20121340320
2013168225
2014104924
2015415314
20161—313
2017378412
20180—37
20190—41
20200—67
20210—01
20220—30
20230—34
20240—01
20250—213
20260—1—

What is it indicated for?

This product is intended for in vitro diagnostic use as assayed quality control materials to monitor the accuracy and reproducibility of the analytes listed in the package insert.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140522: Randox Immunoassay Premium Plus Control Levels 1,2 and Level 3 and Immunoassay Premium Plus Tri-Level Control.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K140522

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026