# K140603: Onepass Endoscopic Vessel Harvesting Sysyetm

- Pathway: 510(k) (Traditional)
- Applicant: Saphena Medical, Inc.
- Decided: Jul 31, 2014 after 143 days
- Product code: GEI (Electrosurgical, Cutting & Coagulation & Accessories)
- Device class: 2
- Regulation: 21 CFR 878.4400
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 5 generations back to K945975 (1994). 16 ancestors in all.
No later device cites it as a predicate yet.
0 of 17 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K091733: Vasoview 6 Pro Endoscopic Vessel Harvesting System, Model VH-2500, Maquet Cardiovascular, LLC (2009) Summary page 1.

## Recalls of this device
- Sep 17, 2024 · Class II · Z-0285-2025: A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into 

## Indications for use
The Saphena Medical OnePass Endoscopic Vessel Harvesting System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extra peritoneall or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thorascopic procedures include exposure and dissection of structures external tothe parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.

Source page: https://510kreport.com/device/K140603
Methodology: https://510kreport.com/methodology