510(k) · Cleared
Recalled · Class II · 2020Biomet Cannulated Screw System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Biomet, Inc. · HWC Screw, Fixation, Bone
Cleared on Jun 3, 2014 as substantially equivalent to K082874 (Biomet Trauma, 2009) and 7 other predicates after a 56-day review. Its family tree reaches back 4 generations to 1992. 16 later devices descend from it. It has been recalled 5 times.
- Decided
- Jun 3, 2014
- Review
- 56 days
- Class
- Class II
- Product code
- HWC
- Recalls
- 5
- Predicates
- 8
Its longest line of descent runs back 4 generations to K922322 (1992). 20 ancestors in all, the oldest K872859 (1987). 16 later devices descend from it, 6 citing it directly.
- Root · 4 generations backK922322 Alphatec 3.5MM Znnulated Cortical Lag ScrewAlphatec Mfg., Inc. · 1992
- 3 generations backK013534 Self-Countersinking Bone ScrewBiomet Orthopedics, Inc. · 2001
- 2 generations backK024086 Propeller Head Small Cannulated Screw SystemBiomet Orthopedics, Inc. · 2003
- 1 generation backK082874 Biodrive Cannulated Screw SystemBiomet Trauma · 2009
- This deviceK140891 Biomet Cannulated Screw SystemBiomet, Inc. · 2014
- Descendant · cites it directly · 8 of its ownK143188 3.Omm Biomet Cannulated Screw SystemBiomet Manufacturing Corp · 2014 · Recalled
- Descendant · cites it directly · 3 of its ownK193214 WasherLoc and No-Profile Screw and Washer Systems, Biomet Cannulated…Biomet, Inc. · 2020
- Descendant · 2 generations on · 3 of its ownK180808 Omni Foot Plating SystemExtremity Medical, LLC · 2018
- Descendant · cites it directly · 2 of its ownK213199 Asnis® III Cannulated Screw System and Asnis® PRO Cannulated Screw…Stryker GmbH · 2022
- Descendant · 2 generations on · 2 of its ownK211290 Cannulated Fasteners and NutsOsteoCentric Technologies · 2021
Showing 5 of 16 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 16 · See all- K1431883.Omm Biomet Cannulated Screw SystemBiomet Manufacturing Corp · 2014 · cites it directly8descendants
- K193214WasherLoc and No-Profile Screw and Washer Systems, Biomet Cannulated Screw System…Biomet, Inc. · 2020 · cites it directly3descendants
- K180808Omni Foot Plating SystemExtremity Medical, LLC · 2018 · 2 generations on3descendants
- K213199Asnis® III Cannulated Screw System and Asnis® PRO Cannulated Screw SystemStryker GmbH · 2022 · cites it directly2descendants
- K211290Cannulated Fasteners and NutsOsteoCentric Technologies · 2021 · 2 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K922322 | Alphatec 3.5MM Znnulated Cortical Lag Screw | Alphatec Mfg., Inc. | 1992 | — |
| −3 | K013534 | Self-Countersinking Bone Screw | Biomet Orthopedics, Inc. | 2001 | — |
| −2 | K024086 | Propeller Head Small Cannulated Screw System | Biomet Orthopedics, Inc. | 2003 | — |
| −1 | K082874 | Biodrive Cannulated Screw System | Biomet Trauma | 2009 | — |
| 0 (this device) | K140891 | Biomet Cannulated Screw System | Biomet, Inc. | 2014 | — |
| +1 | K143188 | 3.Omm Biomet Cannulated Screw System | Biomet Manufacturing Corp | 2014 | 1 |
| +1 | K193214 | WasherLoc and No-Profile Screw and Washer Systems, Biomet Cannulated Screw… | Biomet, Inc. | 2020 | — |
| +2 | K180808 | Omni Foot Plating System | Extremity Medical, LLC | 2018 | — |
| +1 | K213199 | Asnis® III Cannulated Screw System and Asnis® PRO Cannulated Screw System | Stryker GmbH | 2022 | — |
| +2 | K211290 | Cannulated Fasteners and Nuts | OsteoCentric Technologies | 2021 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1987 | 1 | 0 |
| 1990 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1996 | 4 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 2 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2007 | 1 | 0 |
| 2010 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 3 | 0 |
| 2022 | 5 | 0 |
| 2023 | 2 | 0 |
| 2024 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HWC device
- #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (8)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K082874 | Biodrive Cannulated Screw SystemSummary page 2…V alloy is a commonly used material in orthopedic implants, and is a material that was used in predicate devices cleared in K082874, K903810, K(926047, and K(872859. Summary of The Biomet Cannulated Screw System is substantially equivalent to Subst… | 2009 |
| * | K926047 | Ace 8.0MM Cannulated Cancellous Bone ScrewSummary page 1…to currently marketed BioDrive Cannulated Screw System (1(082874), DePuy/Ace 8.0mm Cannulated Cancellous Bone Screw (K(926047), DePuy/Ace Cannulated Self Tapping Cancellous Bone Screw (K(903810), Ace/DePuy… | 1993 |
| * | K903810 | Ace Cannulated Self Tapping Cancellous Bone ScrewSummary page 2…s a commonly used material in orthopedic implants, and is a material that was used in predicate devices cleared in K082874, K903810, K(926047, and K(872859. Summary of The Biomet Cannulated Screw System is substantially equivalent to Substantial… | 1990 |
| * | K872859 | Ace Cancellous Bone ScrewSummary page 1…ew (K(926047), DePuy/Ace Cannulated Self Tapping Cancellous Bone Screw (K(903810), Ace/DePuy Ace Cancellous Bone Screw (K872859), Synthes Sterile 4.5mm Cannulated Screws (K(963172), Synthes Sterile 3.5mm and… | 1987 |
| * | K963172 | Synthes Sterile 4.5 MM Cannulated ScrewsSummary page 1…ellous Bone Screw (K(903810), Ace/DePuy Ace Cancellous Bone Screw (K872859), Synthes Sterile 4.5mm Cannulated Screws (K(963172), Synthes Sterile 3.5mm and 4.0mm Cannulated Screws (K(963192), Synthes 6.5mm C… | 1996 |
| * | K963192 | Synthes Sterile 3.5 MM and 4.0 MM Cannulated ScrewsSummary page 1…rew (K872859), Synthes Sterile 4.5mm Cannulated Screws (K(963172), Synthes Sterile 3.5mm and 4.0mm Cannulated Screws (K(963192), Synthes 6.5mm Cannulated Screw (K(021932), Synthes 7.3mm Cannulated Slipped C… | 1996 |
| * | K021932 | Synthes 6.5 MM Cannulated ScrewSummary page 1…ated Screws (K(963172), Synthes Sterile 3.5mm and 4.0mm Cannulated Screws (K(963192), Synthes 6.5mm Cannulated Screw (K(021932), Synthes 7.3mm Cannulated Slipped Capital Femoral Epiphysis Screws (K(092909). Device Description:… | 2002 |
| * | K092909 | Synthes 7.3 MM Cannulated Slipped Capital Femoral Epiphysis Screws (Scfe)Summary page 1…63192), Synthes 6.5mm Cannulated Screw (K(021932), Synthes 7.3mm Cannulated Slipped Capital Femoral Epiphysis Screws (K(092909). Device Description: The Biomet Cannulated Screw System consists of bone screws, associated washe… | 2010 |
What descends from it? (16)
See all 166 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K143188 | 3.Omm Biomet Cannulated Screw System | 2014 | 8 |
| K193214 | WasherLoc and No-Profile Screw and Washer Systems, Biomet Cannulated Screw System… | 2020 | 3 |
| K213199 | Asnis® III Cannulated Screw System and Asnis® PRO Cannulated Screw System | 2022 | 2 |
| K222409 | CAPTIVATE® Compression Screws | 2022 | 1 |
| K212979 | CalcShift™ Displacement Calcaneal Osteotomy System | 2022 | 0 |
| K213301 | InCore Subtalar System | 2021 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 25, 2020 | Class II | Z-0617-2021 | Process control | Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device. |
| Nov 25, 2020 | Class II | Z-0618-2021 | Process control | Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device. |
| May 14, 2020 | Class II | Z-2304-2020 | Device Design | Distributed to the field without having completed design history files or design transfer activities |
| May 14, 2020 | Class II | Z-2303-2020 | Device Design | Distributed to the field without having completed design history files or design transfer activities |
| Dec 7, 2017 | Class II | Z-1064-2018 | Error in labeling | One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 6.1 years on median.
- 3 of the 20 ancestors have a recall of any class.
- Of its descendants, K143188 (Biomet Manufacturing Corp, 2014) has gone furthest, with 8 descendants of its own.
- 5 other devices also cite its predicate K082874.
How does its category compare?
All HWC devicesHWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · HWC
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HWC
111in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HWC
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HWC
2,893in 2025
MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 77 | 25 | 669 |
| 2008 | 33 | 76 | 33 | 660 |
| 2009 | 27 | 113 | 29 | 651 |
| 2010 | 23 | 110 | 7 | 765 |
| 2011 | 35 | 88 | 12 | 1,104 |
| 2012 | 20 | 99 | 30 | 3,339 |
| 2013 | 25 | 90 | 14 | 3,850 |
| 2014 | 31 | 88 | 2 | 2,549 |
| 2015 | 41 | 85 | 5 | 2,711 |
| 2016 | 26 | 131 | 4 | 2,285 |
| 2017 | 40 | 127 | 4 | 2,561 |
| 2018 | 20 | 106 | 1 | 3,662 |
| 2019 | 30 | 108 | 5 | 5,463 |
| 2020 | 28 | 75 | 4 | 4,820 |
| 2021 | 26 | 84 | 2 | 4,446 |
| 2022 | 26 | 167 | 13 | 3,934 |
| 2023 | 24 | 104 | 0 | 3,718 |
| 2024 | 17 | 90 | 0 | 3,575 |
| 2025 | 15 | 111 | 1 | 2,893 |
| 2026 | 21 | 125 | 2 | — |
What is it indicated for?
Small Cannulated Screws (4.0mm and smaller diameter) are intended for use in: 1. Fixation of small bones, including those in the foot, patella, ankle, wrist and elbow. 2. Arthrodesis of the foot, wrist and elbow. 3. Small and long bone osteotomies. 4. Fracture fixation of small bones, small bone fragments and long bones. Large Cannulated Screws (5mm and larger in diameter) are intended for use in: 1. Fixation of fractures in long bones and long bone…
Read the full statement (152 words)
Small Cannulated Screws (4.0mm and smaller diameter) are intended for use in:
1. Fixation of small bones, including those in the foot, patella, ankle, wrist and elbow.
2. Arthrodesis of the foot, wrist and elbow.
3. Small and long bone osteotomies.
4. Fracture fixation of small bones, small bone fragments and long bones.
Large Cannulated Screws (5mm and larger in diameter) are intended for use in:
1. Fixation of fractures in long bones and long bone fragments.
2. Long bone osteotomies (femur, tibia, foot, ankle, olecranon).
3. Arthrodesis, and fracture fixation of the foot and ankle, such as Jones fractures of the fifth metatarsal, and Calcaneal fractures.
Large Cannulated Screws (6.5mm and larger in diameter) are intended for use in:
1. Slipped capital femoral epiphysis
2. Pediatric femoral neck fractures
3. Tibial plateau fractures
4. SI joint disruptions
5. Intercondylar femur fractures
6. Subtalar arthrodesis
7. Fixation of pelvis and iliosacral joint.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K140891 · printed Nov 4, 2026
Cite this page
510(k) Report. “K140891: Biomet Cannulated Screw System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K140891
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026