# K141102: Hospira Mednet Medication Management Suite

- Pathway: 510(k) (Special)
- Applicant: Hospira, Inc.
- Decided: Jul 8, 2014 after 70 days
- Product code: FRN (Pump, Infusion)
- Device class: 2
- Regulation: 21 CFR 880.5725
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K904518 (1990). 11 ancestors in all.
3 later devices descend from it; 1 cites it directly.
4 of 15 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K042609: Hospira Mednet Medication Management Suite, Hospira, Inc. (2004) Summary page 1.

## Most influential descendants
- K223606: LifeShield™ Infusion Safety Software Suite (2023), 2 descendants of its own
- K242117: LifeShield Infusion Safety Software Suite (2025), 1 descendant of its own
- K252130: LifeShield Infusion Safety Software Suite (2026), 0 descendants of its own

## Recalls of this device
- Oct 30, 2017 · Class II · Z-0195-2018: Issue 1: The MedNet Meds 6.1 and 6.21 programs, under certain conditions, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a me
- Jul 28, 2016 · Class II · Z-2790-2016: Hospira MedNet 6.1 software, in combination with the SapphirePlus 13.1x infusion pump, may result in incorrect bolus amount calculations for drugs whose concentrations are listed in Million Units.

## Indications for use
The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira Infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physician's orders.

Source page: https://510kreport.com/device/K141102