EZ-IO Intraosseous Infusion System

K141117510(k) TraditionalCleared Jul 8, 201469-day reviewClass II21 CFR 880.5570Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vidacare, LLC · FMI Needle, Hypodermic, Single Lumen

Cleared on Jul 8, 2014 after a 69-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). 2 of the 10 devices in its family, counting itself, have a Class I recall. 9 later devices descend from it. It has been recalled 3 times.

Decided
Jul 8, 2014
Review
69 days
Class
Class II
Product code
FMI
Recalls
3
Predicates
1

9 later devices descend from it, 3 citing it directly. 2 of the 10 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2019: 2 descendants cleared, 1 with a Class I recall12018: 1 descendant cleared2022: 2 descendants cleared2021: 2 descendants cleared, 1 with a Class I recall12024: 1 descendant cleared2023: 1 descendant cleared2010201520202025
  1. This deviceK141117 EZ-IO Intraosseous Infusion SystemVidacare, LLC · 2014
  2. Descendant · cites it directly · 3 of its ownK180395 EZ-IO Intraosseous Vascular Access SystemArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2018
  3. Descendant · cites it directly · 2 of its ownK191976 Piper GO-IO Intraosseous Infusion SystemPiper Access, LLC · 2019 · Recalled (Class I)
  4. Descendant · 2 generations on · 1 of its ownK202492 EZ-IO Intraosseous Vascular Access SystemTeleflex Medical · 2021
  5. Descendant · 2 generations on · 1 of its ownK203193 BD Intraosseous Infusion SystemBard Access Systems, Inc (Bas)[Wholly-Owned Subsidiary of Bd] · 2021 · Recalled (Class I)
  6. Descendant · cites it directly · 1 of its ownK191488 SAM IO Intraosseous Access SystemSam® Medical Products, Inc. · 2019

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2018: 12019: 22021: 22022: 22023: 12024: 1
2018high 2 (2019) · low 0 (2020)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 9 · See all
  • K180395EZ-IO Intraosseous Vascular Access SystemArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2018 · cites it directly3descendants
  • K191976Piper GO-IO Intraosseous Infusion SystemPiper Access, LLC · 2019 · cites it directly2descendants
  • K202492EZ-IO Intraosseous Vascular Access SystemTeleflex Medical · 2021 · 2 generations on1descendant
  • K203193BD Intraosseous Infusion SystemBard Access Systems, Inc (Bas)[Wholly-Owned Subsidiary of Bd] · 2021 · 2 generations on1descendant
  • K191488SAM IO Intraosseous Access SystemSam® Medical Products, Inc. · 2019 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K141117EZ-IO Intraosseous Infusion SystemVidacare, LLC2014—
+1K180395EZ-IO Intraosseous Vascular Access SystemArrow International, Inc. (Subsidiary of Teleflex, Inc.)2018—
+1K191976Piper GO-IO Intraosseous Infusion SystemPiper Access, LLC2019Class I
+2K202492EZ-IO Intraosseous Vascular Access SystemTeleflex Medical2021—
+2K203193BD Intraosseous Infusion SystemBard Access Systems, Inc (Bas)[Wholly-Owned Subsidiary of Bd]2021Class I
+1K191488SAM IO Intraosseous Access SystemSam® Medical Products, Inc.2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
201921
202121
202220
202310
202410
FIG. 1 — Lineage of K141117, 2014–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMI device

  1. #1 K252886 Pen Needlescore 14.5 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK132638FDA’s K132638 is Dimension Creatinine (CRE2) Flex Reagent Cartridge
Our reading of the summary gave K132638, FDA’s number for Dimension Creatinine (CRE2) Flex Reagent Cartridge (Siemens Healthcare Diagnostics, Inc.). Its FDA name has none of the distinctive words the summary uses for the device it names, “EZ-IO Intraosseous Infusion System”. Kept off the family tree.
Summary page 2
…s the addition of the distal fe'mur as an insertion site for the pediatric population utilizing the same insertion techniqu~e and devices previously cleared via 510(k) K132638, which includes the insertion sites of the proximal tibia, distal tibia and proximal humerus. Proposed Expanded Indication Dis…
Siemens Healthcare Diagnostics, Inc.20141

What descends from it? (9)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180395EZ-IO Intraosseous Vascular Access SystemArrow International, Inc. (Subsidiary of Teleflex, Inc.)20183
K191976Piper GO-IO Intraosseous Infusion System Class IPiper Access, LLC20192
K191488SAM IO Intraosseous Access SystemSam® Medical Products, Inc.20191

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 8, 2019Class IIZ-0688-2020Package design/selectionThe safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the…
Oct 8, 2019Class IIZ-0689-2020Package design/selectionThe safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the…
Oct 8, 2019Class IIZ-0690-2020Package design/selectionThe safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K191976Piper GO-IO Intraosseous Infusion SystemPiper Access, LLC · 2019 · Cites it directly · ◆ in the 2019 descendants bar of FIG. 12 Class I recalls
  • K203193BD Intraosseous Infusion SystemBard Access Systems, Inc (Bas)[Wholly-Owned Subsidiary of Bd] · 2021 · 2 generations on · ◆ in the 2021 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K180395 (Arrow International, Inc. (Subsidiary of Teleflex, Inc.), 2018) has gone furthest, with 3 descendants of its own.

How does its category compare?

All FMI devices

FMI Needle, Hypodermic, Single Lumen: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FMI

11in 2025

2007: 152008: 102009: 72010: 152011: 122012: 172013: 112014: 142015: 102016: 202017: 92018: 202019: 122020: 112021: 182022: 232023: 182024: 32025: 112026 (partial year): 12
2007high 23 (2022) · low 3 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMI

141in 2025

2007: 71 days median2008: 85 days median2009: 177 days median2010: 119 days median2011: 113 days median2012: 107 days median2013: 87 days median2014: 155 days median2015: 89 days median2016: 217 days median2017: 253 days median2018: 176 days median2019: 186 days median2020: 263 days median2021: 143 days median2022: 189 days median2023: 112 days median2024: 90 days median2025: 141 days median2026 (partial year): 119 days median
2007high 263 (2020) · low 71 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMI

1in 2024

2007: 22008: 42009: 32010: 792011: 52012: 12013: 12014: 32015: 12017: 162019: 52020: 32021: 72023: 22024: 12026 (partial year): 1, 1 Class I
2007high 79 (2010) · low 0 (2016)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · FMI

2,424in 2025

2007: 2082008: 1772009: 2632010: 2492011: 2402012: 2152013: 2552014: 2872015: 5672016: 9412017: 1,5472018: 2,3982019: 4,8962020: 4,2452021: 4,5322022: 4,8852023: 4,9862024: 4,0532025: 2,424
2007high 4,986 (2023) · low 177 (2008)2025

MAUDE reports naming product code FMI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715712208
200810854177
200971773263
20101511979249
2011121135240
2012171071215
201311871255
2014141553287
201510891567
2016202170941
20179253161,547
20182017602,398
20191218654,896
20201126334,245
20211814374,532
20222318904,885
20231811224,986
202439014,053
20251114102,424
2026121191 (1)—

What is it indicated for?

The EZ-IO Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adults and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141117: EZ-IO Intraosseous Infusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141117

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026