# K141213: Ameriwater Ozone Disinfection System

- Pathway: 510(k) (Traditional)
- Applicant: Ameriwater
- Decided: Dec 10, 2014 after 212 days
- Product code: FIN (Tank, Holding, Dialysis)
- Device class: 2
- Regulation: 21 CFR 876.5820
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 3 generations back to K882564 (1989). 8 ancestors in all.
No later device cites it as a predicate yet.
0 of 9 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K093641: TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC (2010) Summary page 4.
- K043207: Ozone Generator, Ge Infrastructure Water & Process Technologies (2005) Summary page 6.

## Indications for use
The AmeriWater Ozone Disinfection System is an optional accessory for the AmeriWater Bicarb Solution Mix and Distribution System (SDS) cleared for market under K051031 and the AmeriWater Water Purification System (WPS) for Hemodialysis cleared for market under K991519. It is intended for use in hospitals and dialysis clinics for the disinfection of the AmeriWater SDS and WPS system. The disinfection process is completed during off-hours when the SDS or WPS are not being used for patient treatment. The AmeriWater Ozone Disinfection System provides dissolved aqueous ozone concentrations of at least 0.5 ppm with a contact time of at least 30 minutes for disinfection of the WPS and SDS systems (the entire disinfection process takes approximately 2 hours to complete and longer contact times may be necessary based on microbial loads). The Ozone Disinfection System is designed to produce water with a microbial count of 50 CFU/ml or less.

Source page: https://510kreport.com/device/K141213