510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Discovery IQ

K141477510(k) TraditionalCleared Sep 18, 2014106-day reviewClass II21 CFR 892.1200Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · KPS System, Tomography, Computed, Emission

Cleared on Sep 18, 2014 as substantially equivalent to K133657 (Ge Medical Systems, LLC, 2014) after a 106-day review. Its family tree reaches back 7 generations to 2001. It has been recalled 3 times.

Decided
Sep 18, 2014
Review
106 days
Class
Class II
Product code
KPS
Recalls
3
Predicates
1

Its longest line of descent runs back 7 generations to K010641 (2001). 9 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 7 generations backK010641 GE CT-Pet SystemGe Medical Systems, Inc. · 2001
  2. 6 generations backK023988 GE Discovery LS SystemGE Medical Systems · 2003 · Recalled
  3. 5 generations backK040172 Modification to GE Discovery LS SystemGE Medical Systems · 2004 · Recalled
  4. 4 generations backK041543 GE Discovery STGE Healthcare · 2004 · Recalled
  5. 3 generations backK050559 GE Discovery VCT SystemGE Medical Systems · 2005 · Recalled
  6. 2 generations backK081496 Optima Pet/CT 560Ge Medical Systems, LLC · 2008 · Recalled
  7. 1 generation backK133657 Discovery Pet/CT 710Ge Medical Systems, LLC · 2014 · Recalled
  8. This deviceK141477 Discovery IQGe Medical Systems, LLC · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K010641GE CT-Pet SystemGe Medical Systems, Inc.2001—
−6K023988GE Discovery LS SystemGE Medical Systems20031
−5K040172Modification to GE Discovery LS SystemGE Medical Systems20045
−4K041543GE Discovery STGE Healthcare20041
−3K050559GE Discovery VCT SystemGE Medical Systems20051
−2K081496Optima Pet/CT 560Ge Medical Systems, LLC20087
−1K133657Discovery Pet/CT 710Ge Medical Systems, LLC20144
0 (this device)K141477Discovery IQGe Medical Systems, LLC2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210
200410
FIG. 1 — Lineage of K141477, 2001–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133657Discovery Pet/CT 710
Summary page 4
…ers on systems with the 2m scan range option. Identification of Legally Marketed Equivalent Devices Discovery PET/CT 710 GE Medical Systems LLC K133657 5-2…
Ge Medical Systems, LLC20144

Also cleared against K133657: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 12, 2019Class IIZ-1748-2019Radiation Control for Health and Safety ActGE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
May 24, 2018Class IIZ-2578-2018Component change controlSome CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
May 19, 2017Class IIZ-2192-2017Software designIt was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 1.2 years on median.
  • 7 of the 9 ancestors have a recall of any class.
  • 1 other device also cites its predicate K133657.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

GE Discovery IQ PET/CT system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The system is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions in patients of all ages. This…

Read the full statement (127 words)

GE Discovery IQ PET/CT system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.

The system is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions in patients of all ages. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning.

The system can also be used as a stand-alone head and whole body multi-slice computed tomography (CT) diagnostic imaging systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141477: Discovery IQ.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141477

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026