510(k) · Cleared
ClearedX-Repairsl
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Synthasome, Inc. · FTL Mesh, Surgical, Polymeric
Cleared on Jul 2, 2014 as substantially equivalent to K083307 (Synthasome, Inc., 2009) after a 26-day review. Its family tree reaches back 10 generations to 1992.
- Decided
- Jul 2, 2014
- Review
- 26 days
- Class
- Class II
- Product code
- FTL
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 10 generations to K921895 (1992). 37 ancestors in all, the oldest K810428 (1981). No later device cites it as a predicate yet.
- Root · 10 generations backK921895 Tissue Patch, Tissue GraftBio-Vascular, Inc. · 1992
- 6 generations between
- 3 generations backK052830 SportmeshArtimplant AB · 2006
- 2 generations backK071887 Sportmesh or Artelon Tissue ReinforcementArtimplant AB · 2007
- 1 generation backK083307 Synthasome X-RepairSynthasome, Inc. · 2009
- This deviceK141499 X-RepairslSynthasome, Inc. · 2014
Show all 10 generations
- Root · 10 generations backK921895 Tissue Patch, Tissue GraftBio-Vascular, Inc. · 1992
- 9 generations backK961810 Supple Peri-Guard PericardiumBio-Vascular, Inc. · 1996
- 8 generations backK980431 SurgisisCook Biotech, Inc. · 1998
- 7 generations backK992159 Surgisis SlingCook Biotech, Inc. · 1999
- 6 generations backK001738 Depuy Restore Orthobiologic Soft Tissue ImplantDePuy Orthopaedics, Inc. · 2000
- 5 generations backK020049 Fortaflex Surgical MeshOrganogenesis, Inc. · 2002
- 4 generations backK042809 CuffpatchOrganogenesis, Inc. · 2004
- 3 generations backK052830 SportmeshArtimplant AB · 2006
- 2 generations backK071887 Sportmesh or Artelon Tissue ReinforcementArtimplant AB · 2007
- 1 generation backK083307 Synthasome X-RepairSynthasome, Inc. · 2009
- This deviceK141499 X-RepairslSynthasome, Inc. · 2014
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −10 | K921895 | Tissue Patch, Tissue Graft | Bio-Vascular, Inc. | 1992 | — |
| −9 | K961810 | Supple Peri-Guard Pericardium | Bio-Vascular, Inc. | 1996 | — |
| −8 | K980431 | Surgisis | Cook Biotech, Inc. | 1998 | — |
| −7 | K992159 | Surgisis Sling | Cook Biotech, Inc. | 1999 | — |
| −6 | K001738 | Depuy Restore Orthobiologic Soft Tissue Implant | DePuy Orthopaedics, Inc. | 2000 | — |
| −5 | K020049 | Fortaflex Surgical Mesh | Organogenesis, Inc. | 2002 | — |
| −4 | K042809 | Cuffpatch | Organogenesis, Inc. | 2004 | — |
| −3 | K052830 | Sportmesh | Artimplant AB | 2006 | — |
| −2 | K071887 | Sportmesh or Artelon Tissue Reinforcement | Artimplant AB | 2007 | — |
| −1 | K083307 | Synthasome X-Repair | Synthasome, Inc. | 2009 | — |
| 0 (this device) | K141499 | X-Repairsl | Synthasome, Inc. | 2014 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1983 | 1 | 0 |
| 1985 | 1 | 0 |
| 1988 | 1 | 0 |
| 1991 | 3 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1996 | 3 | 0 |
| 1997 | 1 | 0 |
| 1998 | 6 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
| 2007 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FTL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FTL device
- #1 K220091 Poly-Tape/Infinity-Lock Soft Tissue…score 16.8 · 14 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K083307 | Synthasome X-RepairSummary page 1…21 CFR 878.3300. class 11,FTL OWW Mesh, Surgical, Polymeric Predicate Device Identification: X-Repair (K083307) Device Description: The X-RepairSL is a bioabsorbable, rectangular, single-layered, flexible, woven surgical mesh manufactured f… | 2009 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 22 years and passes through 6 companies; a new generation was cleared every 1.7 years on median.
- 1 of the 37 ancestors has a recall of any class.
- 3 other devices also cite its predicate K083307.
How does its category compare?
All FTL devicesFTL Mesh, Surgical, Polymeric: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · FTL
6in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FTL
30in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FTL
1in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FTL
3,487in 2025
MAUDE reports naming product code FTL, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 20 | 92 | 2 | 1,737 |
| 2008 | 23 | 114 | 1 | 1,569 |
| 2009 | 15 | 97 | 1 | 1,906 |
| 2010 | 18 | 132 | 12 | 2,175 |
| 2011 | 15 | 90 | 8 | 3,608 |
| 2012 | 12 | 213 | 2 | 14,700 |
| 2013 | 8 | 90 | 1 | 26,519 |
| 2014 | 11 | 70 | 3 | 2,683 |
| 2015 | 12 | 121 | 0 | 2,526 |
| 2016 | 3 | 186 | 0 | 2,417 |
| 2017 | 6 | 132 | 1 | 3,947 |
| 2018 | 12 | 105 | 1 | 11,893 |
| 2019 | 4 | 99 | 0 | 21,724 |
| 2020 | 5 | 138 | 1 | 18,819 |
| 2021 | 2 | — | 0 | 16,284 |
| 2022 | 6 | 195 | 0 | 11,439 |
| 2023 | 4 | 29 | 0 | 8,025 |
| 2024 | 2 | — | 0 | 8,565 |
| 2025 | 6 | 30 | 0 | 3,487 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
X-RepairSL is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists.
X-RepairSL is also intended for reinforcement of soft tissues that are repaired by suture or suture anchors, during tendon repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, or quadriceps tendons.
X-RepairSL is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K141499 · printed Nov 6, 2026
Cite this page
510(k) Report. “K141499: X-Repairsl.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141499
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026