510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Advia Centaur TSH

K141999510(k) SpecialCleared Apr 22, 2015273-day reviewClass II21 CFR 862.1690Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics, Inc. · JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Cleared on Apr 22, 2015 as substantially equivalent to K910981 (Ciba Corning Diagnostics Corp., 1991) after a 273-day review. Its family tree reaches back 1 generation to 1991. It has been recalled 3 times.

Decided
Apr 22, 2015
Review
273 days
Class
Class II
Product code
JLW
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K910981 (1991). No later device cites it as a predicate yet.

K910981 · Ciba Corning Acs TSH Immunoassay · Ciba Corning Diagnostics Corp. · 1991199019952000200520102015
  1. Root · 1 generation backK910981 Ciba Corning Acs TSH ImmunoassayCiba Corning Diagnostics Corp. · 1991
  2. This deviceK141999 Advia Centaur TSHSiemens Healthcare Diagnostics, Inc. · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K910981Ciba Corning Acs TSH ImmunoassayCiba Corning Diagnostics Corp.1991—
0 (this device)K141999Advia Centaur TSHSiemens Healthcare Diagnostics, Inc.2015—
FIG. 1 — Lineage of K141999, 1991–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JLW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JLW device

  1. #1 K251543 Atellica® IM TSH3-Ultra II (TSH3ULII)score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K910981Ciba Corning Acs TSH Immunoassay
Summary page 4
…862.1690 4. Predicate Device Information The ADVIA Centaur TSH assay was originally cleared by the FDA on 04/30/1991 (k910981) as the ACS TSH Immunoassay. 5. Description of Device Modifications No changes were made to the ADVIA Centaur TSH…
Ciba Corning Diagnostics Corp.1991—
*—ADVIA Centaur XP system
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 2, 2017Class IIZ-1537-2017Under Investigation by firmMultiple software issues, which may affect the operation and workflow of the system. Potential for an apparent delay to testing when these issues occur. The software issues affecting the analyzer may impact all analytes available on the…
May 26, 2016Class IIZ-2256-2016Under Investigation by firmEight (8) issues were identified which may affect the results generated by the system software version.
Apr 28, 2016Class IIZ-0072-2017Under Investigation by firmThe ADVIA Centaur XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.

How does its category compare?

All JLW devices

JLW Radioimmunoassay, Thyroid-Stimulating Hormone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · JLW

1in 2025

2008: 52009: 22010: 12012: 12013: 12014: 12015: 32016: 32017: 42019: 12022: 12023: 12024: 22025: 12026 (partial year): 1
2007high 5 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JLW

194in 2017

2008: 143 days median2015: 142 days median2016: 238 days median2017: 194 days median
2007high 238 (2016) · low 142 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JLW

2in 2025

2007: 12010: 12011: 12012: 32016: 12018: 42019: 12020: 22024: 22025: 2
2007high 4 (2018) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JLW

103in 2025

2007: 192008: 102009: 412010: 342011: 1672012: 1442013: 1682014: 1552015: 2032016: 1832017: 2352018: 2492019: 2162020: 1452021: 1462022: 1372023: 1622024: 1692025: 103
2007high 249 (2018) · low 10 (2008)2025

MAUDE reports naming product code JLW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—119
20085143010
20092—041
20101—134
20110—1167
20121—3144
20131—0168
20141—0155
201531420203
201632381183
201741940235
20180—4249
20191—1216
20200—2145
20210—0146
20221—0137
20231—0162
20242—2169
20251—2103
20261—0—

What is it indicated for?

For in vitro diagnostic use in the quantitative determination of thyroid-stimulating hormone (TSH, thyrotropin) in serum using the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141999: Advia Centaur TSH.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K141999

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026