# K142129: Rapidfret Oral Fluid Assay for Cocaine, Rapidfret Oral Fluid Cocaine Calibrator Set, Rapidfret Oral Fluid Cocaine Control Set

- Pathway: 510(k) (Traditional)
- Applicant: Biophor Diagnostics, Inc.
- Decided: Jul 24, 2015 after 354 days
- Product code: DIO (Enzyme Immunoassay, Cocaine And Cocaine Metabolites)
- Device class: 2
- Regulation: 21 CFR 862.3250
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K001197 (2000). 2 ancestors in all.
No later device cites it as a predicate yet.
0 of 3 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K101742: Thermo Scientific Cedia Cocaine Oft Assay, Microgenics Corp. (2011) Summary page 4.

## Indications for use
The RapidFRET Oral Fluid Assay for Cocaine is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for cocaine at 20 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay is calibrated with Cocaine and provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only. The RapidFRET Oral Fluid Cocaine Calibrator Set and RapidFRET Oral Fluid Cocaine Control Set are intended for use with the RapidFRET Oral Fluid Assay for Cocaine and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.