Recalled · Class II · 2015NS3000 Holmium Laser System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
New Star Lasers, Inc. · GEX Powered Laser Surgical Instrument
Cleared on Oct 31, 2014 as substantially equivalent to K132607 (New Star Lasers, Inc., 2014) and 1 other predicate after a 74-day review. Its family tree reaches back 2 generations to 2004. It has been recalled once.
- Decided
- Oct 31, 2014
- Review
- 74 days
- Class
- Class II
- Product code
- GEX
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 2 generations to K033437 (2004). No later device cites it as a predicate yet.
View FIG. 1 as a table
Line of descent and top descendants
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in GEX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new GEX device
- #1 K222064 The Alma Soprano Titaniumscore 38.1 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K132607 | NS 2500 Holmium Laser SystemSummary page 4…Class II FDA Panel: General and Plastic Surgery Product Code: GEX III. Predicate NS2500™ Laser System (K132607) IV. Device Description The New Star Model NS3000 is a compact, portable, self-contained system that produces a… | 2014 |
| * | K033437 | Sphinx Laser SystemsSummary page 6…opsy. VII. Summary of Technological Characteristics Technical K132607 K142286 Characteristics K033437 NS2500 New Star Laser NS3000 New Star Laser (Predicate)… | 2004 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 13, 2015 | Class II | Z-1918-2015 | Software design (manufacturing process) | During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value displayed on the… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 10.2 years on median.
- None of the 2 ancestors has a recall.
How does its category compare?
All GEX devicesGEX Powered Laser Surgical Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · GEX
83in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · GEX
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · GEX
3in 2024
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · GEX
2,720in 2025
MAUDE reports naming product code GEX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 73 | 79 | 9 | 342 |
| 2008 | 107 | 86 | 2 | 232 |
| 2009 | 69 | 100 | 0 | 209 |
| 2010 | 50 | 96 | 4 | 1,711 |
| 2011 | 41 | 83 | 18 | 3,024 |
| 2012 | 45 | 118 | 5 | 2,000 |
| 2013 | 46 | 144 | 3 | 2,236 |
| 2014 | 61 | 147 | 6 | 2,036 |
| 2015 | 47 | 112 | 0 | 2,099 |
| 2016 | 51 | 114 | 1 | 2,585 |
| 2017 | 67 | 90 | 5 (1) | 2,977 |
| 2018 | 51 | 78 | 5 | 3,525 |
| 2019 | 41 | 88 | 6 | 2,709 |
| 2020 | 52 | 90 | 1 | 2,183 |
| 2021 | 53 | 123 | 0 | 3,465 |
| 2022 | 60 | 152 | 0 | 2,288 |
| 2023 | 52 | 87 | 6 | 2,337 |
| 2024 | 82 | 90 | 3 | 2,187 |
| 2025 | 83 | 90 | 0 | 2,720 |
| 2026 | 57 | 90 | 2 | — |
What is it indicated for?
The NS3000 laser system is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery. Urology: Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: urethral strictures, bladder…
Read the full statement (145 words)
The NS3000 laser system is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.
Urology: Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: urethral strictures, bladder neck incisions (BNI), ablation and resection of bladder tumors, uretheral tumors and ureteral tumors, condylomas, and lesions of external genitalia.
Lithotripsy and Percutaneous Urinary Lithotripsy: Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi, endoscopic fragmentation of kidney calculi, and treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
General Surgery: Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: skin incision, excision of external and internal lesions, complete or partial resection of internal organs, tumors and lesions, and biopsy.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K142286 · printed Nov 5, 2026
Cite this page
510(k) Report. “K142286: NS3000 Holmium Laser System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K142286
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026