Endo Vive 3S Low Profile Balloon Kit, Endo Vive 3S Bolus Extension Sets, Endo Vive 3S Continuous Extension Sets, Endo Vive 3S Medication Extension Set

K142297510(k) TraditionalCleared Jan 8, 2015143-day reviewClass II21 CFR 876.5980Recalls 104

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Xeridiem (Formerly Mri) · PIF Gastrointestinal Tubes With Enteral Specific Connectors

Cleared on Jan 8, 2015 as substantially equivalent to K043114 (Kimberly-Clark Corp., 2004) and 1 other predicate after a 143-day review. Its family tree reaches back 1 generation to 1998. It has been recalled 104 times.

Decided
Jan 8, 2015
Review
143 days
Class
Class II
Product code
PIF
Recalls
104
Predicates
2

Its longest line of descent runs back 1 generation to K971434 (1998). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2004: 1 other ancestorK971434 · A.M.T. G-Tube Balloon Replacement Gastrostomy Feeding Device · Applied Medical Technologies · 199819952000200520102015
Other ancestors, by generation 1 generation back: 1 other ancestor K971434 · A.M.T. G-Tube Balloon Replacement Gastrostomy Feeding Device · Applied Medical Technologies · 19980
  1. Root · 1 generation backK971434 A.M.T. G-Tube Balloon Replacement Gastrostomy Feeding DeviceApplied Medical Technologies · 1998
  2. This deviceK142297 Endo Vive 3S Low Profile Balloon Kit, Endo Vive 3S Bolus Extension…Xeridiem (Formerly Mri) · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K971434A.M.T. G-Tube Balloon Replacement Gastrostomy Feeding DeviceApplied Medical Technologies1998—
0 (this device)K142297Endo Vive 3S Low Profile Balloon Kit, Endo Vive 3S Bolus Extension Sets, Endo…Xeridiem (Formerly Mri)2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200410
FIG. 1 — Lineage of K142297, 1998–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PIF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PIF device

  1. #1 K253715 Edi Catheter ENFit 6Fr/48 cm (6893166); Edi…score 15.5 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043114Kimberly-Clark Mic-Key Low Profile Gastrostomy Tube
Summary page 4
…FR 876.5980, Product Code KNT] Classification Panel: Gastroenterology/Urology Class: Class II 510(k) Number: K043114 Secondary Predicate Device Information Trade Name: AMT Mini ONE® Balloon Button Low Profile Gastrostomy D…
Kimberly-Clark Corp.2004—
*K971434A.M.T. G-Tube Balloon Replacement Gastrostomy Feeding Device
Summary page 4
…FR 876.5980, Product Code KNT] Classification Panel: Gastroenterology/Urology Class: Class II 510(k) Number: K971434 Xeridiem 01/06/15 Page 5-2 === Page 5…
Applied Medical Technologies1998—

Also cleared against K043114: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 23, 2015Class IIZ-0924-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0947-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0933-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0981-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0950-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0917-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0938-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0963-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0954-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0974-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0979-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0955-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0912-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0916-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0958-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0930-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0929-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0945-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0940-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0952-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0942-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0949-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0941-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0951-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0975-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0968-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0962-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0978-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0980-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0956-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0915-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0914-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0922-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0935-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0926-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0928-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0932-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0961-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0944-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0931-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0946-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0920-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0948-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0939-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0973-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0919-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0927-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0943-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0977-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…
Dec 23, 2015Class IIZ-0964-2016Device DesignA feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • 1 other device also cites its predicate K043114.

How does its category compare?

All PIF devices

PIF Gastrointestinal Tubes With Enteral Specific Connectors: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · PIF

2in 2025

2014: 32015: 92016: 32017: 52018: 22019: 72020: 42021: 42022: 42023: 42024: 22025: 22026 (partial year): 3
2007high 9 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PIF

242in 2023

2014: 119 days median2015: 143 days median2016: 267 days median2017: 174 days median2019: 135 days median2020: 165 days median2021: 242 days median2022: 167 days median2023: 242 days median2026 (partial year): 239 days median
2007high 267 (2016) · low 119 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PIF

1in 2024

2015: 1062024: 1
2007high 106 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PIF

83in 2025

2015: 122016: 452017: 172018: 172019: 282020: 702021: 892022: 772023: 962024: 1102025: 83
2007high 110 (2024) · low 0 (2007)2025

MAUDE reports naming product code PIF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
2014311900
2015914310612
20163267045
20175174017
20182—017
20197135028
20204165070
20214242089
20224167077
20234242096
20242—1110
20252—083
202632390—

What is it indicated for?

The EndoVive™ 3S Low Profile Balloon is indicated for use in adult and pediatric patients who require long term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.

The EndoVive™ 3S Extension Set is indicated for the delivery of nutrition, hydration and/or medication into the stomach through a low profile balloon and also provides a mechanism for gastric decompression.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142297: Endo Vive 3S Low Profile Balloon Kit, Endo Vive 3S Bolus Extension Sets, Endo Vive 3S Continuous Extension Sets, Endo Vive 3S Medication Extension Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K142297

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.