510(k) Report

510(k) · Cleared

Cleared

smARTrack Feeding Tube System

K142327510(k) TraditionalCleared Jun 8, 2015292-day reviewClass II21 CFR 876.5980

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Art Healthcare, Ltd. · KNT Tubes, Gastrointestinal (And Accessories)

Cleared on Jun 8, 2015 as substantially equivalent to K113351 (Corpak Medsystems, 2012) and 2 other predicates after a 292-day review. Its family tree reaches back 3 generations to 1983. 1 of the 9 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it.

Decided
Jun 8, 2015
Review
292 days
Class
Class II
Product code
KNT
Recalls
None
Predicates
3

Its longest line of descent runs back 3 generations to K831868 (1983). 6 ancestors in all. 2 later devices descend from it, 1 citing it directly. 1 of the 9 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor2008: 1 other ancestor2023: 1 descendant cleared2024: 1 descendant clearedK831868 · Dobbhoff Feeding Tube W/Hydromer Lub · Biosearch Medical Products, Inc. · 1983K112511 · Dobbhoff Dual Port Feeding Tube · Covidien Lp, Formerly Registered AS Kendall · 2012K123555 · Kangaroo Feeding Tubes with Iris Technology · Covidien · 201419801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall13 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K123555 · Kangaroo Feeding Tubes with Iris Technology · Covidien · 2014K112511 · Dobbhoff Dual Port Feeding Tube · Covidien Lp, Formerly Registered AS Kendall · 2012K831868 · Dobbhoff Feeding Tube W/Hydromer Lub · Biosearch Medical Products, Inc. · 19830
  1. Root · 3 generations backK831868 Dobbhoff Feeding Tube W/Hydromer LubBiosearch Medical Products, Inc. · 1983
  2. 2 generations backK112511 Dobbhoff Dual Port Feeding TubeCovidien Lp, Formerly Registered AS Kendall · 2012
  3. 1 generation backK123555 Kangaroo Feeding Tubes with Iris TechnologyCovidien · 2014
  4. This deviceK142327 smARTrack Feeding Tube SystemArt Healthcare, Ltd. · 2015
  5. Descendant · cites it directly · 1 of its ownK230206 Entarik Feeding Tube System; Entarik Feeding TubeGravitas Medical, Inc. · 2023
  6. Descendant · 2 generations on · 0 of its ownK241169 Entarik NI Feeding Tube SystemGravitas Medical, Inc. · 2024

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K230206Entarik Feeding Tube System; Entarik Feeding TubeGravitas Medical, Inc. · 2023 · cites it directly1descendant
  • K241169Entarik NI Feeding Tube SystemGravitas Medical, Inc. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K831868Dobbhoff Feeding Tube W/Hydromer LubBiosearch Medical Products, Inc.1983—
−2K112511Dobbhoff Dual Port Feeding TubeCovidien Lp, Formerly Registered AS Kendall2012—
−1K123555Kangaroo Feeding Tubes with Iris TechnologyCovidien2014—
0 (this device)K142327smARTrack Feeding Tube SystemArt Healthcare, Ltd.2015—
+1K230206Entarik Feeding Tube System; Entarik Feeding TubeGravitas Medical, Inc.2023—
+2K241169Entarik NI Feeding Tube SystemGravitas Medical, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210
200810
201211

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
FIG. 1 — Lineage of K142327, 1983–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KNT device

  1. #1 K261890 52Fr ViSiGi LUX™; 52Fr ViSiGi 3D®score 12.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K113351Cortrak Enteral Access Device
Summary page 5
…2 Classification Panel: Gastroenterology/Urology 3. PREDICATE DEVICES Main predicate:  CORTRAK 2 Enteral Access Device by CORPAK MedSystems; K113351 Reference devices:  Kangaroo™ Feeding Tube with IRIS Technology by Covidien; K123555  ZepHr Impedance/…
Corpak Medsystems2012Class I
*K123555Kangaroo Feeding Tubes with Iris Technology
Summary page 5
…e:  CORTRAK 2 Enteral Access Device by CORPAK MedSystems; K113351 Reference devices:  Kangaroo™ Feeding Tube with IRIS Technology by Covidien; K123555  ZepHr Impedance/pH Reflux Monitoring System by Sandhill Scientific Inc.; K012232…
Covidien2014—
*K012232Insight, Model S980000
Summary page 5
… Kangaroo™ Feeding Tube with IRIS Technology by Covidien; K123555  ZepHr Impedance/pH Reflux Monitoring System by Sandhill Scientific Inc.; K012232 K142327…
Sandhill Scientific, Inc.20022

Also cleared against K113351: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K230206Entarik Feeding Tube System; Entarik Feeding TubeGravitas Medical, Inc.20231

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K113351Cortrak Enteral Access DeviceCorpak Medsystems · 2012 · Direct predicate · ◆ in the 2012 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 32 years and passes through 4 companies; a new generation was cleared every 1.9 years on median.
  • 3 of the 6 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K230206 (Gravitas Medical, Inc., 2023) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K113351.

How does its category compare?

All KNT devices

KNT Tubes, Gastrointestinal (And Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · KNT

4in 2025

2007: 102008: 102009: 112010: 122011: 72012: 122013: 72014: 162015: 102016: 62017: 92018: 72019: 102020: 52021: 62022: 72023: 52024: 92025: 42026 (partial year): 3
2007high 16 (2014) · low 4 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KNT

136in 2025

2007: 118 days median2008: 115 days median2009: 102 days median2010: 104 days median2011: 211 days median2012: 107 days median2013: 128 days median2014: 259 days median2015: 140 days median2016: 99 days median2017: 247 days median2018: 149 days median2019: 126 days median2020: 102 days median2021: 121 days median2022: 110 days median2023: 162 days median2024: 103 days median2025: 136 days median2026 (partial year): 105 days median
2007high 259 (2014) · low 99 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KNT

13in 2025

2007: 12008: 822009: 92010: 32011: 22012: 12015: 12016: 32017: 42018: 22019: 22020: 12021: 22022: 1, 1 Class I2023: 22025: 132026 (partial year): 20
2007high 82 (2008) · low 0 (2013)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · KNT

1,451in 2025

2007: 3002008: 8672009: 8182010: 5172011: 5312012: 5282013: 7472014: 1,0302015: 1,0482016: 1,2512017: 1,6722018: 1,9012019: 1,8382020: 1,6412021: 1,6552022: 1,6502023: 1,8632024: 1,3442025: 1,451
2007high 1,901 (2018) · low 300 (2007)2025

MAUDE reports naming product code KNT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007101181300
20081011582867
2009111029818
2010121043517
201172112531
2012121071528
201371280747
20141625901,030
20151014011,048
201669931,251
2017924741,672
2018714921,901
20191012621,838
2020510211,641
2021612121,655
202271101 (1)1,650
2023516221,863
2024910301,344
20254136131,451
2026310520—

What is it indicated for?

The smARTrack™ Feeding Tube System is designed to aid qualified operators in the placement of the smARTrack Feeding Tube into the stomach of patients requiring enteral feeding. The smARTrack™ tube is equipped with sensors designed to provide information about the location of the tube tip relative to the lower esophageal sphincter (LES) thus assisting in reducing the incidence of misplacement during first positioning. The smARTrack™ tube also monitors the…

Read the full statement (171 words)

The smARTrack™ Feeding Tube System is designed to aid qualified operators in the placement of the smARTrack Feeding Tube into the stomach of patients requiring enteral feeding. The smARTrack™ tube is equipped with sensors designed to provide information about the location of the tube tip relative to the lower esophageal sphincter (LES) thus assisting in reducing the incidence of misplacement during first positioning. The smARTrack™ tube also monitors the position continuously during the course of the feeding and automatically and in real time alerts of tube migration. The smARTrack console is equipped with a motorized mechanism which automatically and in real-time stops feeding if the feeding tube moves out of position during ongoing use. Furthermore, smARTrack™ Feeding Tube System can guide operator to correctly re-position the tube.

The smARTrack™ Feeding Tube is intended for gastric decompression, gastric lavage, and the administration of nutrition, fluids and medications by the naso-enteric route for patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142327: smARTrack Feeding Tube System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142327

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026