510(k) Report

510(k) · Cleared

Cleared

FaStep Marijuana Tests (Strip, Panel Dip, Quick Cup, Turn-Key Split Cup), FaStep Methamphetamine Tests (Strip, Panel Dip, Quick Cup, Turn-Key Split Cup)

K142408510(k) TraditionalCleared Sep 25, 201428-day reviewClass II21 CFR 862.3870

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Polymed Therapeutics, Inc. · LDJ Enzyme Immunoassay, Cannabinoids

Cleared on Sep 25, 2014 as substantially equivalent to K052115 (First Check Diagnostics, LLC, 2005) after a 28-day review. Its family tree reaches back 2 generations to 2005.

Decided
Sep 25, 2014
Review
28 days
Class
Class II
Product code
LDJ
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K050186 (2005). No later device cites it as a predicate yet.

  1. Root · 2 generations backK050186 Ameditech Immutest Multi-Drug Screen Panel IiiAmeditech, Inc. · 2005 · Recalled
  2. 1 generation backK052115 First Check Multi Drug Cup 12 for Marijuana (THC), Cocaine (Coc)…First Check Diagnostics, LLC · 2005 · Recalled
  3. This deviceK142408 FaStep Marijuana Tests (Strip, Panel Dip, Quick Cup, Turn-Key Split…Polymed Therapeutics, Inc. · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K050186Ameditech Immutest Multi-Drug Screen Panel IiiAmeditech, Inc.20053
−1K052115First Check Multi Drug Cup 12 for Marijuana (THC), Cocaine (Coc), Amphetamine…First Check Diagnostics, LLC200586
0 (this device)K142408FaStep Marijuana Tests (Strip, Panel Dip, Quick Cup, Turn-Key Split Cup)…Polymed Therapeutics, Inc.2014—
FIG. 1 — Lineage of K142408, 2005–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LDJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDJ device

  1. #1 K232732 Rapid Marijuana (THC) Test Strip 20, Rapid…score 9.4 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052115First Check Multi Drug Cup 12 for Marijuana (THC), Cocaine (Coc), Amphetamine (Amp)…
Summary page 4
…R, 862.3870 Cannabinoid Test System Toxicology LAF 21 CFR, 862.3610 Methamphetamine Test System Toxicology 5. Predicate Devices: k052115 FIRST CHECK DIAGNOSTICS LLC 6. Intended Use: FaStep Marijuana Tests are immunochromatographic a…
First Check Diagnostics, LLC200586

Also cleared against K052115: 12 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 8.7 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • 12 other devices also cite its predicate K052115.

How does its category compare?

All LDJ devices

LDJ Enzyme Immunoassay, Cannabinoids: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LDJ

2in 2024

2008: 22009: 12011: 42012: 42013: 32014: 22015: 32016: 12019: 12023: 12024: 2
2007high 4 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDJ

48in 2015

2011: 219 days median2012: 193 days median2013: 162 days median2015: 48 days median
2007high 219 (2011) · low 48 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDJ

1in 2023

2008: 42023: 1
2007high 4 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LDJ

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2025

MAUDE reports naming product code LDJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—40
20091—00
20100—00
2011421900
2012419300
2013316200
20142—00
201534801
20161—00
20170—00
20180—00
20191—00
20200—00
20210—00
20220—00
20231—10
20242—00
20250—00
20260—0—

What is it indicated for?

FaStep Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-nor-A9-THC-9-COOH in human urine at a cut-off concentration of 50ng/mL. The test is available in a Strip format, a Panel Dip format, a Quick Cup format and a Turn-Key Split Cup format. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the…

Read the full statement (221 words)

FaStep Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-nor-A9-THC-9-COOH in human urine at a cut-off concentration of 50ng/mL. The test is available in a Strip format, a Panel Dip format, a Quick Cup format and a Turn-Key Split Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

FaStep Methamphetamine Tests are immunochromatographic assays for the qualitative determination of methamphetamine in human urine at a cut-off concentration of 1000 ng/mL. The test is available in a Strip format, a Panel Dip format, a Quick Cup format and a Turn-Key Split Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K142408: FaStep Marijuana Tests (Strip, Panel Dip, Quick Cup, Turn-Key Split Cup), FaStep Methamphetamine Tests (Strip, Panel Dip, Quick Cup, Turn-Key Split Cup).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142408

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026