510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

LIFEPAK 15 monitor/defibrillator

K142430510(k) TraditionalCleared Dec 19, 2014112-day reviewClass III21 CFR 870.5310Recalls 7 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Physio-Control, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 19, 2014 as substantially equivalent to K082937 (Physio-Control, Inc., 2009) and 1 other predicate after a 112-day review. Its family tree reaches back 11 generations to 1993. 1 of the 35 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 7 times, including 3 Class I recalls.

Decided
Dec 19, 2014
Review
112 days
Class
Class III
Product code
MKJ
Recalls
7 (3 Class I)
Predicates
2

Its longest line of descent runs back 11 generations to K921669 (1993). 32 ancestors in all, the oldest K802288 (1980). 2 later devices descend from it, 1 citing it directly. 1 of the 35 devices in this family, counting this one, has a Class I recall.

  1. Root · 11 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 7 generations between
  3. 3 generations backK063510 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2007 · Recalled
  4. 2 generations backK082937 Lifepak 15 Monitor/DefibrillatorPhysio-Control, Inc. · 2009 · Recalled
  5. 1 generation backK103567 Lifepak 15 Monitor/DefibrillatorPhysio-Control, Inc. · 2011
  6. This deviceK142430 LIFEPAK 15 monitor/defibrillatorPhysio-Control, Inc. · 2014
  7. Descendant · cites it directly · 1 of its ownK240162 LIFEPAK® 35 AC Power Adapter (41335-000001)Stryker Physio-Control · 2024
  8. Descendant · 2 generations on · 0 of its ownK253099 LIFEPAK 15 AC Power Adapter (41577-000333)Physio-Control, Inc. · 2025
Show all 11 generations
  1. Root · 11 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 10 generations backK964750 Eagle 4000 Patient MonitorMarquette Electronics, Inc. · 1997
  3. 9 generations backK973486 Life Pak 12Physio-Control Corp. · 1998 · Recalled
  4. 8 generations backK991910 Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · Recalled
  5. 7 generations backK012274 Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · Recalled
  6. 6 generations backK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled
  7. 5 generations backK040775 Lifepak 12Medtronic Physio-Control Corp. · 2004 · Recalled
  8. 4 generations backK041459 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2004
  9. 3 generations backK063510 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2007 · Recalled
  10. 2 generations backK082937 Lifepak 15 Monitor/DefibrillatorPhysio-Control, Inc. · 2009 · Recalled
  11. 1 generation backK103567 Lifepak 15 Monitor/DefibrillatorPhysio-Control, Inc. · 2011
  12. This deviceK142430 LIFEPAK 15 monitor/defibrillatorPhysio-Control, Inc. · 2014

Most influential descendants

Showing 2 of 2 · See all
  • K240162LIFEPAK® 35 AC Power Adapter (41335-000001)Stryker Physio-Control · 2024 · cites it directly1descendant
  • K253099LIFEPAK 15 AC Power Adapter (41577-000333)Physio-Control, Inc. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K921669Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc.1993—
−10K964750Eagle 4000 Patient MonitorMarquette Electronics, Inc.1997—
−9K973486Life Pak 12Physio-Control Corp.19983
−8K991910Lifepak 12 BiphasicPhysio-Control Corp.19993
−7K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp.200217
−6K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031
−5K040775Lifepak 12Medtronic Physio-Control Corp.20041
−4K041459Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc.2004—
−3K063510Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc.20072
−2K082937Lifepak 15 Monitor/DefibrillatorPhysio-Control, Inc.20093
−1K103567Lifepak 15 Monitor/DefibrillatorPhysio-Control, Inc.2011—
0 (this device)K142430LIFEPAK 15 monitor/defibrillatorPhysio-Control, Inc.2014—
+1K240162LIFEPAK® 35 AC Power Adapter (41335-000001)Stryker Physio-Control2024—
+2K253099LIFEPAK 15 AC Power Adapter (41577-000333)Physio-Control, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
199010
199110
199220
199410
199540
199620
199820
199920
200140

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202510
FIG. 1 — Lineage of K142430, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082937Lifepak 15 Monitor/Defibrillator
Summary page 8
…are below: Predicate Device 510(k) Number Clearance Date ® LIFEPAK 15 monitor/defibrillator K082937 March 11, 2009 ® LIFEPAK 15 monitor/defibrillator K103567 March 22,…
Physio-Control, Inc.20093
*K103567Lifepak 15 Monitor/Defibrillator
Summary page 8
…® LIFEPAK 15 monitor/defibrillator K082937 March 11, 2009 ® LIFEPAK 15 monitor/defibrillator K103567 March 22, 2011 Description: The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response…
Physio-Control, Inc.2011—

Also cleared against K082937: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K240162LIFEPAK® 35 AC Power Adapter (41335-000001)Stryker Physio-Control20241

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 20, 2019Class IIZ-0816-2020Device DesignMonitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button.
Dec 20, 2019Class IIZ-0817-2020Device DesignMonitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button.
Feb 1, 2019Class IZ-0863-2019Software designCertain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no…
Feb 1, 2019Class IZ-0862-2019Software designCertain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no…
Jan 16, 2017Class IIZ-1144-2017Process controlThe End-Tidal CO2 (EtCO2) reading can intermittently show a value of XXX after start-up or during device operation.
Jan 12, 2017Class IZ-1123-2017OtherThe firm has become aware of a potential device issue where the LIFEPAK 15 device may no longer be able to deliver defibrillation energy due to a potential failure of an internal component (Relay) installed on the Therapy Printed Circuit…
Jan 27, 2016Class IIZ-1993-2016Software designThe firm became aware that when using EtC02 in the kPa or % setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially obscure a portion of the leading digit of the EtC02…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 15 of the 32 ancestors have a recall of any class.
  • Of its descendants, K240162 (Stryker Physio-Control, 2024) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K082937.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

Manual Defibrillation: Indications Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia. Contraindications Defibrillation is…

Read the full statement (404 words)

Manual Defibrillation:

Indications

Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.

Contraindications

Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as idioventricular or ventricular escape rhythms, and in the treatment of asystole.

Automated External Defibrillation:

Indications

AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient’s ECG rhythm. In AED mode, the LIFEPAK 15 monitor/defibrillator is not intended for use on pediatric patients less than 8 years old.

Noninvasive Pacing

Indications

Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.

Contraindications

Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.

12-lead Electrocardiography:

Indications

The 12-lead electrocardiogram is used to identify, diagnose and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute ST-elevation myocardial infarction (STEMI).

Pulse Oximetry

Indications

Pulse Oximetry is indicated for use in any patient who is at risk of developing hypoxemia, carboxyhemoglobinemia, or methemoglobinemia. SpO, monitoring may be used during no motion and motion conditions, and in patients who are well or poorly perfused. SpCO and SpMet accuracies have not been validated under motion or low perfusion conditions.

Noninvasive Blood Pressure Monitoring:

Indications

Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension and monitoring BP trends in patient conditions such as, but not limited to, shock, acute dysrhythmia, or major fluid imbalance.

End-Tidal CO2 monitoring:

Indications

EtCO, monitoring is used to detect trends in the level of expired CO,. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.

Invasive Pressure Monitoring:

Indications

Invasive pressure monitoring is indicated for use in patients who require continuous monitoring of physiological pressures in order to rapidly assess changes in the patient’s condition or response to therapy. It may also be used to aid in medical diagnosis.

Temperature Monitoring:

Indications

Temperature monitoring is indicated for use in patients who require continuous monitoring of body temperature.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K142430: LIFEPAK 15 monitor/defibrillator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142430

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026