510(k) Report

510(k) · Cleared

Recalled · Class I · 2022

Revolution Full Face Mask

K142554510(k) AbbreviatedCleared Jan 9, 2015120-day reviewClass II21 CFR 868.5905Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · BZD Ventilator, Non-Continuous (Respirator)

Cleared on Jan 9, 2015 as substantially equivalent to K082866 (Respironics Inc., Sleep & Home Respiratory Group, 2009) after a 120-day review. Its family tree reaches back 3 generations to 1999. 2 of the 7 devices in its family, counting itself, have a Class I recall. It has been recalled once, including a Class I recall.

Decided
Jan 9, 2015
Review
120 days
Class
Class II
Product code
BZD
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 3 generations to K984428 (1999). 6 ancestors in all, the oldest K954207 (1996). No later device cites it as a predicate yet. 2 of the 7 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2008: 1 other ancestor2000: 1 other ancestor1996: 1 other ancestorK984428 · Sullivan Autoset Nasal Cpap System · Resmed, Ltd. · 1999K023068 · Papillon Mask Set · Map Medizintechnik Fur Arzt Und Patient GmbH · 2003K082866 · Revolutionfull Face Mask · Respironics Inc., Sleep & Home Respiratory Group · 2009 · Class I recall19952000200520102015
  1. Root · 3 generations backK984428 Sullivan Autoset Nasal Cpap SystemResmed, Ltd. · 1999
  2. 2 generations backK023068 Papillon Mask SetMap Medizintechnik Fur Arzt Und Patient GmbH · 2003
  3. 1 generation backK082866 Revolutionfull Face MaskRespironics Inc., Sleep & Home Respiratory Group · 2009 · Recalled (Class I)
  4. This deviceK142554 Revolution Full Face MaskRespironics, Inc. · 2015

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K984428Sullivan Autoset Nasal Cpap SystemResmed, Ltd.1999—
−2K023068Papillon Mask SetMap Medizintechnik Fur Arzt Und Patient GmbH2003—
−1K082866Revolutionfull Face MaskRespironics Inc., Sleep & Home Respiratory Group2009Class I
0 (this device)K142554Revolution Full Face MaskRespironics, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200010
200810
FIG. 1 — Lineage of K142554, 1999–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZD device

  1. #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082866Revolutionfull Face Mask
Summary page 4
…ification Name / BZD – Ventilator, non-continuous (respirator) Product Code Predicate Device(s) Respironics Revolution Full Face Mask (K082866) Indication for Use This mask is intended to provide an interface for application of CPAP or bi-level therapy t…
Respironics Inc., Sleep & Home Respiratory Group2009Class I

Also cleared against K082866: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 6, 2022Class IZ-1781-2022Error in labelingNo contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K082866Revolutionfull Face MaskRespironics Inc., Sleep & Home Respiratory Group · 2009 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 5.7 years on median.
  • 2 of the 6 ancestors have a recall of any class; 1 has a Class I recall.
  • 1 other device also cites its predicate K082866.

How does its category compare?

All BZD devices

BZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · BZD

7in 2025

2007: 212008: 172009: 232010: 122011: 232012: 212013: 232014: 142015: 112016: 122017: 52018: 52019: 62020: 62021: 62022: 42023: 52024: 102025: 72026 (partial year): 10
2007high 23 (2009) · low 4 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZD

189in 2025

2007: 83 days median2008: 108 days median2009: 90 days median2010: 122 days median2011: 125 days median2012: 133 days median2013: 127 days median2014: 172 days median2015: 224 days median2016: 217 days median2017: 221 days median2018: 280 days median2019: 231 days median2020: 137 days median2021: 192 days median2022: 253 days median2023: 204 days median2024: 132 days median2025: 189 days median2026 (partial year): 214 days median
2007high 280 (2018) · low 83 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZD

8in 2023

2007: 292008: 22009: 12011: 12019: 4, 4 Class I2023: 8, 7 Class I2026 (partial year): 2, 2 Class I
2007high 29 (2007) · low 0 (2010)2026

Class I (13 since 2007) Class II/III not classified

Adverse event reports per year · BZD

60,279in 2025

2007: 652008: 672009: 1472010: 1432011: 1892012: 1392013: 822014: 862015: 1142016: 1182017: 1482018: 1012019: 1882020: 1262021: 6,0662022: 90,7682023: 27,9872024: 68,2622025: 60,279
2007high 90,768 (2022) · low 65 (2007)2025

MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721832965
200817108267
200923901147
2010121220143
2011231251189
2012211330139
201323127082
201414172086
2015112240114
2016122170118
201752210148
201852800101
201962314 (4)188
202061370126
2021619206,066
20224253090,768
202352048 (7)27,987
202410132068,262
20257189060,279
2026102142 (2)—

What is it indicated for?

This mask is intended to provide an interface for application of CPAP or bi-level therapy to patients. The mask is for single patient use in the home or multi-patient use in the hospital/institutional environment. The mask is to be used on patients (>66lbs/30kg) for whom CPAP and bi-level therapy has been prescribed.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142554: Revolution Full Face Mask.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142554

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026