510(k) Report

510(k) · Cleared

Cleared

E-CUBE 5

K142733510(k) TraditionalCleared Oct 28, 201435-day reviewClass II21 CFR 892.1550

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Alpinion Medical Systems Co., Ltd. · IYN System, Imaging, Pulsed Doppler, Ultrasonic

Cleared on Oct 28, 2014 as substantially equivalent to K132687 (Alpinion Medical Systems Co., Ltd., 2013) after a 35-day review. Its family tree reaches back 6 generations to 2001. 15 later devices descend from it.

Decided
Oct 28, 2014
Review
35 days
Class
Class II
Product code
IYN
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K010329 (2001). 15 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2016: 2 descendants cleared2019: 1 descendant cleared2017: 1 descendant cleared2018: 3 descendants cleared2022: 3 descendants cleared2020: 2 descendants cleared2021: 2 descendants cleared2023: 1 descendant clearedK010329 · LOGIQ 500 · General Electric Co. · 2001K014097 · GE Logiq 5 · GE Medical Systems · 2002 · recalledK032974 · GE Logiq 5 Pro or Expert · General Electric Co. · 2003K060993 · GE Logiq P5 and A5 · General Electric Co. · 2006K120060 · E-CUBE 9 · Alpinion Medical Systems Co., Ltd. · 2012K132687 · E-CUBE 7 · Alpinion Medical Systems Co., Ltd. · 2013200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 5 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 2 descendants7 generations forward: 1 descendant K132687 · E-CUBE 7 · Alpinion Medical Systems Co., Ltd. · 2013K120060 · E-CUBE 9 · Alpinion Medical Systems Co., Ltd. · 2012K060993 · GE Logiq P5 and A5 · General Electric Co. · 2006K032974 · GE Logiq 5 Pro or Expert · General Electric Co. · 2003K014097 · GE Logiq 5 · GE Medical Systems · 2002 · recalledK010329 · LOGIQ 500 · General Electric Co. · 2001−50+5
  1. Root · 6 generations backK010329 LOGIQ 500General Electric Co. · 2001
  2. 5 generations backK014097 GE Logiq 5GE Medical Systems · 2002 · Recalled
  3. 4 generations backK032974 GE Logiq 5 Pro or ExpertGeneral Electric Co. · 2003
  4. 3 generations backK060993 GE Logiq P5 and A5General Electric Co. · 2006
  5. 2 generations backK120060 E-CUBE 9Alpinion Medical Systems Co., Ltd. · 2012
  6. 1 generation backK132687 E-CUBE 7Alpinion Medical Systems Co., Ltd. · 2013
  7. This deviceK142733 E-CUBE 5Alpinion Medical Systems Co., Ltd. · 2014
  8. Descendant · cites it directly · 14 of its ownK153424 E-CUBE i7Alpinion Medical Systems Co., Ltd. · 2016
  9. Descendant · 2 generations on · 12 of its ownK161439 E-CUBE 11Alpinion Medical Systems Co., Ltd. · 2016
  10. Descendant · 2 generations on · 11 of its ownK172732 E-CUBE 8Alpinion Medical Systems Co., Ltd. · 2017
  11. Descendant · 3 generations on · 7 of its ownK181277 E-CUBE 12Alpinion Medical Systems Co., Ltd. · 2018
  12. Descendant · 4 generations on · 4 of its ownK200449 X-Cube 90Alpinion Medical Systems Co., Ltd. · 2020

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2016: 22017: 12018: 32019: 12020: 22021: 22022: 32023: 1
2016high 3 (2018) · low 0 (2024)2026

Most influential descendants

Showing 5 of 15 · See all
  • K153424E-CUBE i7Alpinion Medical Systems Co., Ltd. · 2016 · cites it directly14descendants
  • K161439E-CUBE 11Alpinion Medical Systems Co., Ltd. · 2016 · 2 generations on12descendants
  • K172732E-CUBE 8Alpinion Medical Systems Co., Ltd. · 2017 · 2 generations on11descendants
  • K181277E-CUBE 12Alpinion Medical Systems Co., Ltd. · 2018 · 3 generations on7descendants
  • K200449X-Cube 90Alpinion Medical Systems Co., Ltd. · 2020 · 4 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K010329LOGIQ 500General Electric Co.2001—
−5K014097GE Logiq 5GE Medical Systems20021
−4K032974GE Logiq 5 Pro or ExpertGeneral Electric Co.2003—
−3K060993GE Logiq P5 and A5General Electric Co.2006—
−2K120060E-CUBE 9Alpinion Medical Systems Co., Ltd.2012—
−1K132687E-CUBE 7Alpinion Medical Systems Co., Ltd.2013—
0 (this device)K142733E-CUBE 5Alpinion Medical Systems Co., Ltd.2014—
+1K153424E-CUBE i7Alpinion Medical Systems Co., Ltd.2016—
+2K161439E-CUBE 11Alpinion Medical Systems Co., Ltd.2016—
+2K172732E-CUBE 8Alpinion Medical Systems Co., Ltd.2017—
+3K181277E-CUBE 12Alpinion Medical Systems Co., Ltd.2018—
+4K200449X-Cube 90Alpinion Medical Systems Co., Ltd.2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201620
201710
201830
201910
202020
202120
202230
202310
FIG. 1 — Lineage of K142733, 2001–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYN device

  1. #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132687E-CUBE 7
Summary page 9
…Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX Predicate Device(s) K132687 E-CUBE 7 Diagnostic Ultrasound System Device Description: E-CUBE 5 product is an ultrasound imaging system for medical…
Alpinion Medical Systems Co., Ltd.2013—

Also cleared against K132687: 4 other devices →

What descends from it? (15)

See all 15

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153424E-CUBE i7Alpinion Medical Systems Co., Ltd.201614

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K153424 (Alpinion Medical Systems Co., Ltd., 2016) has gone furthest, with 14 descendants of its own.
  • 4 other devices also cite its predicate K132687.

How does its category compare?

All IYN devices

IYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · IYN

57in 2025

2007: 132008: 252009: 182010: 292011: 392012: 432013: 512014: 562015: 542016: 602017: 642018: 592019: 472020: 572021: 462022: 532023: 442024: 502025: 572026 (partial year): 44
2007high 64 (2017) · low 13 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYN

124in 2025

2007: 26 days median2008: 18 days median2009: 29 days median2010: 67 days median2011: 46 days median2012: 37 days median2013: 60 days median2014: 39 days median2015: 55 days median2016: 56 days median2017: 83 days median2018: 85 days median2019: 79 days median2020: 72 days median2021: 105 days median2022: 105 days median2023: 115 days median2024: 119 days median2025: 124 days median2026 (partial year): 107 days median
2007high 124 (2025) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYN

10in 2025

2007: 162008: 122009: 82010: 102011: 82012: 32013: 82014: 22015: 32016: 22018: 72020: 42021: 82022: 102023: 42024: 12025: 102026 (partial year): 2, 1 Class I
2007high 16 (2007) · low 0 (2017)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · IYN

610in 2025

2007: 32008: 22009: 92010: 172011: 182012: 202013: 182014: 482015: 1162016: 1452017: 1462018: 852019: 592020: 672021: 2132022: 1862023: 2292024: 2662025: 610
2007high 610 (2025) · low 2 (2008)2025

MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071326163
20082518122
2009182989
201029671017
20113946818
20124337320
20135160818
20145639248
201554553116
201660562145
201764830146
20185985785
20194779059
20205772467
2021461058213
20225310510186
2023441154229
2024501191266
20255712410610
2026441072 (1)—

What is it indicated for?

The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical

applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Trans-rectal(TR);

Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Peripheral Vascular (PV); and

Urology (including prostate).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142733: E-CUBE 5.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142733

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026