510(k) Report

510(k) · Cleared

Recalled · Class I · 2017

Skintact, various other tradenames

K142803510(k) TraditionalCleared Dec 2, 201464-day reviewClass III21 CFR 870.5310Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Leonhard Lang GmbH · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 2, 2014 as substantially equivalent to K072233 (Leonhard Lang GmbH, 2007) and 1 other predicate after a 64-day review. Its family tree reaches back 3 generations to 1996. 1 of the 7 devices in its family, counting itself, has a Class I recall. It has been recalled 2 times, including 2 Class I recalls.

Decided
Dec 2, 2014
Review
64 days
Class
Class III
Product code
MKJ
Recalls
2 (2 Class I)
Predicates
2

Its longest line of descent runs back 3 generations to K955628 (1996). 6 ancestors in all, the oldest K914955 (1992). No later device cites it as a predicate yet. 1 of the 7 devices in this family, counting this one, has a Class I recall.

  1. Root · 3 generations backK955628 Fore Runner External Defibrillator/Defibrillator PadsHeartstream, Inc. · 1996
  2. 2 generations backK041883 Skintact Multifunction Electrodes, Model DF 21Leonhard Lang GmbH · 2004
  3. 1 generation backK072233 Skintact Multifunctional Electrodes with DH02 Gel with Different…Leonhard Lang GmbH · 2007 · Recalled
  4. This deviceK142803 Skintact, various other tradenamesLeonhard Lang GmbH · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K955628Fore Runner External Defibrillator/Defibrillator PadsHeartstream, Inc.1996—
−2K041883Skintact Multifunction Electrodes, Model DF 21Leonhard Lang GmbH2004—
−1K072233Skintact Multifunctional Electrodes with DH02 Gel with Different Connectors…Leonhard Lang GmbH20072
0 (this device)K142803Skintact, various other tradenamesLeonhard Lang GmbH2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199710
200810
FIG. 1 — Lineage of K142803, 1996–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072233Skintact Multifunctional Electrodes with DH02 Gel with Different Connectors Compatible…
Summary page 4
…rgy, (including paddles) PRO CODE: MKJ Regulation No.: 21CFR 870.5310 Classification: Class III SUBSTANTIALLY EQUIVALENT TO: K072233 Skintact® Multifunction Electrodes with DH02 Gel available with different connectors compatible…
Leonhard Lang GmbH20072
*K0820909131 Defibrillation Electrodes
Summary page 4
…available with different connectors compatible with different devices, Leonhard Lang GmbH, decision date: 10/05/2007 K082090 9131 Defibrillation Electrodes, Cardiac Science Corporation, decision date: 12/12/2008 INDICATI…
Cardiac Science Corporation2008—

Also cleared against K072233: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2017Class IZ-3109-2017Component design/selectionCertain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening…
Sep 1, 2016Class IZ-0004-2017Device DesignThere is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 2 companies; a new generation was cleared every 7.2 years on median.
  • 2 of the 6 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K072233.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

Skintact® Multifunction Electrodes are for use on adults and children weighing more than 25 kg for external defibrillation, pacing, monitoring and cardioversion. The device is non-sterile and single use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142803: Skintact, various other tradenames.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142803

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026