510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Parietex Monofilament Polyester Mesh (new name: Parietex Lightweight Mesh), Parietex Composite Mono PM Mesh (new name: Parietex Composite Parastomal Mesh), Parietex Composite Ventral Patch, Symbotex Composite Mesh

K142908510(k) SpecialCleared Nov 14, 201439-day reviewClass II21 CFR 878.3300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sofradim Production · FTL Mesh, Surgical, Polymeric

Cleared on Nov 14, 2014 as substantially equivalent to K090858 (Sofradim Production, 2009) and 3 other predicates after a 39-day review. Its family tree reaches back 6 generations to 1999. 5 later devices descend from it. It has been recalled once.

Decided
Nov 14, 2014
Review
39 days
Class
Class II
Product code
FTL
Recalls
1
Predicates
4

Its longest line of descent runs back 6 generations to K991400 (1999). 18 ancestors in all, the oldest K930586 (1993). 5 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2011: 2 other ancestors2003: 1 other ancestor2002: 2 other ancestors2008: 1 other ancestor2005: 1 other ancestor2001: 2 other ancestors1998: 1 other ancestor1993: 1 other ancestor2022: 1 descendant cleared2015: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedK991400 · Parietene Polypropylene Mesh · Sofradim Production · 1999×2: 2 generations cleared close together, K033376 (2004), K040998 (2004) K081126 · Parietex Composite Mono PM Mesh · Sofradim Production · 2008 · recalledK090858 · Parietex Monofilament Polyester Mesh · Sofradim Production · 2009K131969 · Symbotex (TM) Composite Mesh · Sofradim Production · 20131990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 4 other ancestors3 generations back: 5 other ancestors4 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant K131969 · Symbotex (TM) Composite Mesh · Sofradim Production · 2013K090858 · Parietex Monofilament Polyester Mesh · Sofradim Production · 2009K081126 · Parietex Composite Mono PM Mesh · Sofradim Production · 2008 · recalledK040998 · Modification to Parietex Composite (Pco) Mesh · Sofradim Production · 2004K033376 · Ugytex Mesh · Sofradim Production · 2004K991400 · Parietene Polypropylene Mesh · Sofradim Production · 1999−50
  1. Root · 6 generations backK991400 Parietene Polypropylene MeshSofradim Production · 1999
  2. 5 generations backK033376 Ugytex MeshSofradim Production · 2004
  3. 4 generations backK040998 Modification to Parietex Composite (Pco) MeshSofradim Production · 2004
  4. 3 generations backK081126 Parietex Composite Mono PM MeshSofradim Production · 2008 · Recalled
  5. 2 generations backK090858 Parietex Monofilament Polyester MeshSofradim Production · 2009
  6. 1 generation backK131969 Symbotex (TM) Composite MeshSofradim Production · 2013
  7. This deviceK142908 Parietex Monofilament Polyester Mesh (new name: Parietex Lightweight…Sofradim Production · 2014
  8. Descendant · cites it directly · 4 of its ownK150091 Versatex Monofilament MeshSofradim Production · 2015 · Recalled
  9. Descendant · cites it directly · 3 of its ownK220586 Symbotex Composite Mesh, ProGrip Self-Gripping Polyester Mesh…Sofradim Production · 2022
  10. Descendant · 2 generations on · 2 of its ownK232373 Progrip™ Self-Gripping Polypropylene MeshSofradim Production · 2024
  11. Descendant · 3 generations on · 1 of its ownK243315 ProGrip™ Self-Gripping Polypropylene MeshSofradim Production · 2025
  12. Descendant · 4 generations on · 0 of its ownK253956 ProGrip™ advanced self-gripping polypropylene mesh (ADG1510…Sofradim Production · 2026

Most influential descendants

Showing 5 of 5 · See all
  • K150091Versatex Monofilament MeshSofradim Production · 2015 · cites it directly4descendants
  • K220586Symbotex Composite Mesh, ProGrip Self-Gripping Polyester Mesh, Parietex Hydrophilic 2D…Sofradim Production · 2022 · cites it directly3descendants
  • K232373Progrip™ Self-Gripping Polypropylene MeshSofradim Production · 2024 · 2 generations on2descendants
  • K243315ProGrip™ Self-Gripping Polypropylene MeshSofradim Production · 2025 · 3 generations on1descendant
  • K253956ProGrip™ advanced self-gripping polypropylene mesh (ADG1510, ADG2015, ADG3020, ADG4030)Sofradim Production · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K991400Parietene Polypropylene MeshSofradim Production1999—
−5K033376Ugytex MeshSofradim Production2004—
−4K040998Modification to Parietex Composite (Pco) MeshSofradim Production2004—
−3K081126Parietex Composite Mono PM MeshSofradim Production20081
−2K090858Parietex Monofilament Polyester MeshSofradim Production2009—
−1K131969Symbotex (TM) Composite MeshSofradim Production2013—
0 (this device)K142908Parietex Monofilament Polyester Mesh (new name: Parietex Lightweight Mesh)…Sofradim Production2014—
+1K150091Versatex Monofilament MeshSofradim Production20152
+1K220586Symbotex Composite Mesh, ProGrip Self-Gripping Polyester Mesh, Parietex…Sofradim Production2022—
+2K232373Progrip™ Self-Gripping Polypropylene MeshSofradim Production2024—
+3K243315ProGrip™ Self-Gripping Polypropylene MeshSofradim Production2025—
+4K253956ProGrip™ advanced self-gripping polypropylene mesh (ADG1510, ADG2015, ADG3020…Sofradim Production2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199810
200120
200220
200310
200510
200810
201120
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
202210
202410
202510
202610
FIG. 1 — Lineage of K142908, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FTL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FTL device

  1. #1 K220091 Poly-Tape/Infinity-Lock Soft Tissue…score 16.8 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090858Parietex Monofilament Polyester Mesh
Summary page 8
…Page 2 of 5 Legally marketed devices to which equivalence is claimed: Parietex™ Monofilament Polyester Mesh (K090858) Parietex™ Composite Mono PM Mesh (K081126) Parietex™ C…
Sofradim Production2009—
*K081126Parietex Composite Mono PM Mesh
Summary page 8
…ed devices to which equivalence is claimed: Parietex™ Monofilament Polyester Mesh (K090858) Parietex™ Composite Mono PM Mesh (K081126) Parietex™ Composite Ventral Patch (K120506) Symbotex™…
Sofradim Production20081
*K120506Parietex Composite Ventral Patch
Summary page 8
…er Mesh (K090858) Parietex™ Composite Mono PM Mesh (K081126) Parietex™ Composite Ventral Patch (K120506) Symbotex™ Composite Mesh (K131969) Reason for 510(k) Submission: The p…
Sofradim Production2012—
*K131969Symbotex (TM) Composite Mesh
Summary page 8
…te Mono PM Mesh (K081126) Parietex™ Composite Ventral Patch (K120506) Symbotex™ Composite Mesh (K131969) Reason for 510(k) Submission: The purpose of this 510(k) is to notify the Agency of the addition of another…
Sofradim Production2013—

Also cleared against K090858: 2 other devices →

What descends from it? (5)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150091Versatex Monofilament MeshSofradim Production20154
K220586Symbotex Composite Mesh, ProGrip Self-Gripping Polyester Mesh, Parietex Hydrophilic 2D…Sofradim Production20223

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 5, 2018Class IIZ-0461-2019Unknown/Undetermined by firmThe firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these reports, product…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 1 company; a new generation was cleared every 4.3 years on median.
  • 5 of the 18 ancestors have a recall of any class.
  • Of its descendants, K150091 (Sofradim Production, 2015) has gone furthest, with 4 descendants of its own.
  • 2 other devices also cite its predicate K090858.

How does its category compare?

All FTL devices

FTL Mesh, Surgical, Polymeric: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FTL

6in 2025

2007: 202008: 232009: 152010: 182011: 152012: 122013: 82014: 112015: 122016: 32017: 62018: 122019: 42020: 52021: 22022: 62023: 42024: 22025: 62026 (partial year): 2
2007high 23 (2008) · low 2 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FTL

30in 2025

2007: 92 days median2008: 114 days median2009: 97 days median2010: 132 days median2011: 90 days median2012: 213 days median2013: 90 days median2014: 70 days median2015: 121 days median2016: 186 days median2017: 132 days median2018: 105 days median2019: 99 days median2020: 138 days median2022: 195 days median2023: 29 days median2025: 30 days median
2007high 213 (2012) · low 29 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FTL

1in 2020

2007: 22008: 12009: 12010: 122011: 82012: 22013: 12014: 32017: 12018: 12020: 1
2007high 12 (2010) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FTL

3,487in 2025

2007: 1,7372008: 1,5692009: 1,9062010: 2,1752011: 3,6082012: 14,7002013: 26,5192014: 2,6832015: 2,5262016: 2,4172017: 3,9472018: 11,8932019: 21,7242020: 18,8192021: 16,2842022: 11,4392023: 8,0252024: 8,5652025: 3,487
2007high 26,519 (2013) · low 1,569 (2008)2025

MAUDE reports naming product code FTL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007209221,737
20082311411,569
2009159711,906
201018132122,175
2011159083,608
201212213214,700
2013890126,519
2014117032,683
20151212102,526
2016318602,417
2017613213,947
201812105111,893
2019499021,724
20205138118,819
20212—016,284
20226195011,439
202342908,025
20242—08,565
202563003,487
20262—0—

What is it indicated for?

Parietex™ lightweight mesh is indicated for inguinal and ventral hernia repair. The PARIETEX™ Parastomal Mesh is indicated for the reinforcement of soft tissues during surgical repair, and specifically for the repair of parastomal hernias. The non-absorbable polyester mesh provides long term reinforcement of soft tissues. The absorbable hydrophilic film minimizes tissue attachment to the mesh when in direct contact with the viscera. The Parietex™…

Read the full statement (183 words)

Parietex™ lightweight mesh is indicated for inguinal and ventral hernia repair.

The PARIETEX™ Parastomal Mesh is indicated for the reinforcement of soft tissues during surgical repair, and specifically for the repair of parastomal hernias.

The non-absorbable polyester mesh provides long term reinforcement of soft tissues.

The absorbable hydrophilic film minimizes tissue attachment to the mesh when in direct contact with the viscera.

The Parietex™ composite ventral patch is used for the reinforcement of soft tissues during surgical repair. It is indicated for the treatment of ventral defects (primary and incisional hernias).

The three-dimensional non-absorbable monofilament polyester mesh provides long term reinforcement of soft tissues. The absorbable hydrophilic film minimizes tissue attachment to the mesh in case of direct contact with the viscera.

Symbotex™ composite mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists in procedures involving primary abdominal wall and incisional hernia surgeries.

The non-absorbable three-dimensional polyester mesh provides long term reinforcement of soft tissues. On the opposite side, the absorbable hydrophilic film minimizes tissue attachment to the mesh in case of direct contact with the viscera.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K142908: Parietex Monofilament Polyester Mesh (new name: Parietex Lightweight Mesh), Parietex Composite Mono PM Mesh (new name: Parietex Composite Parastomal Mesh), Parietex Composite Ventral Patch, Symbotex Composite Mesh.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K142908

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026