# K142955: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open, SOMATON Emotion 6/16, SOMATOM Sensation 64/Sensation Cardiac , SOMATOM Perspective SOMATOM Scope/ Scope Power

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Medical Solutions USA, Inc.
- Decided: Nov 24, 2015 after 406 days
- Product code: JAK (System, X-Ray, Tomography, Computed)
- Device class: 2
- Regulation: 21 CFR 892.1750
- Recalls of this device: 16 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 8 generations back to K013561 (2001). 29 ancestors in all.
27 later devices descend from it; 5 cite it directly.
0 of 57 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K133589: Somatom Force, Siemens Medical Solutions USA, Inc. (2014) Summary page 18.
- K121072: Somatom Definition Flash, Siemens Medical Systems, Inc. (2012) Summary page 19.
- K120579: Somatom Definition Edge, Siemens Medical Systems, Inc. (2012) Summary page 19.
- K081022: Somatom Definition, Model as/as+, Siemens Medical Solutions USA, Inc. (2008) Summary page 18.
- K122471: Somatom Definition, Siemens Medical Systems, Inc. (2012) Summary page 19.
- K130901: Somatom Definition as Open, Siemens Medical Systems, Inc. (2014) Summary page 19.
- K133424: Somatom Emotion 6 CT Systems, Somatom Emotion 16, Siemens Medical Solutions USA, Inc. (2014) Summary page 19.
- K133590: Somatom Perspective, Siemens Medical Solutions USA, Inc. (2014) Summary page 20.
- K113287: Somatom Perspective, Siemens Medical Solutions USA, Inc. (2012) Summary page 19.
- K140912: Somatom Scope/Somatom Scope Power, Siemens Medical Solutions USA, Inc. (2014) Summary page 20.
- K982349: Somatom Plus 4WITH Volume Zoom CT Scanners, Siemens Medical Solutions USA, Inc. (1998) Summary page 20.

## Most influential descendants
- K161196: SOMATOM Drive (2016), 24 descendants of its own
- K162302: Somatom Confidence ® RT Pro (2016), 21 descendants of its own
- K173630: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Deinition AS/AS+, SOMATOM Confidence (2018), 20 descendants of its own
- K190578: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Edge Plus, SOMATOM Definition AS/AS+, SOMATOM Confidence (2019), 19 descendants of its own
- K163297: Scan&GO (2017), 19 descendants of its own

## Recalls of this device
- Nov 24, 2020 · Class II · Z-0771-2021: System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
- Jun 22, 2020 · Class II · Z-3025-2020: Sporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan
- Oct 7, 2019 · Class II · Z-0406-2020: Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes
- Feb 15, 2018 · Class II · Z-1446-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1449-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1455-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1456-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1450-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1445-2018: To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1457-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1458-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Feb 15, 2018 · Class II · Z-1447-2018: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on 
- Nov 8, 2017 · Class II · Z-1745-2018: There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, S
- Nov 8, 2017 · Class II · Z-1747-2018: There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, S
- Nov 8, 2017 · Class II · Z-1746-2018: There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, S
- Nov 8, 2017 · Class II · Z-1748-2018: There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, S

## Indications for use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Siemens SOMATOM Definition AS/ AS+ (Project P46) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Siemens SOMATOM Definition AS Open systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Siemens SOMATOM Definition Flash system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. In addition the SOMATOM Definition Flash is able to produce additional image planes and analysis results by executing optional post processing features, which operate on DICOM images. The images and results delivered by the system can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Siemens SOMATOM Definition Edge (Project P46F) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The SOMATOM P45 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The SOMATOM Emotion 16 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Siemens SOMATOM Perspective 64/128 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. In addition the SOMATOM Perspective is able to produce additional image planes and analysis results by executing optional post-processing features, which operate on DICOM images. For cardiac imaging, which is an option on the system, a new reconstruction algorithm (iTRIM - Iterative Temporal Resolution Improvement Method) is used. iTRIM improves the temporal resolution of cardiac CT images compared to conventional cardiac CT image reconstruction. Actual results obtained with iTRIM can vary depending on the particular clinical situation. The images and results delivered by the SOMATOM Perspective can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Siemens SOMATOM Perspective 16/32 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. In addition the SOMATOM Perspective is able to produce additional image planes and analysis results by executing optional post-processing features, which operate on DICOM images. For cardiac imaging, which is an option on the system, a new reconstruction algorithm (iTRIM - Iterative Temporal Resolution Improvement Method) is used. iTRIM improves the temporal resolution of cardiac CT images compared to conventional cardiac CT image reconstruction. Actual results obtained with iTRIM can vary depending on the particular clinical situation. The images and results delivered by the SOMATOM Perspective can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The SOMATOM Scope and the SOMATOM Scope Power are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles. The images and results delivered by the SOMATOM Scope and the SOMATOM Scope Power can be used by a trained physician as an aid in diagnosis. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. The Somatom Plus 4 with Volume Zoom package is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.