510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Evis Exera II Duodenovideoscope Olympus TJF Type Q180V

K143153510(k) TraditionalCleared Jan 15, 2016438-day reviewClass II21 CFR 876.1500Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corp. · FDT Duodenoscope And Accessories, Flexible/Rigid

Cleared on Jan 15, 2016 as substantially equivalent to K080403 (Olympus Medical Systems Corporation, 2008) after a 438-day review. Its family tree reaches back 1 generation to 2008. 14 later devices descend from it. It has been recalled 3 times.

Decided
Jan 15, 2016
Review
438 days
Class
Class II
Product code
FDT
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K080403 (2008). 14 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2019: 1 descendant cleared2020: 3 descendants cleared2024: 1 descendant cleared2022: 2 descendants cleared2026: 3 descendants cleared2025: 4 descendants clearedK080403 · XTJF Type Q160VF1 · Olympus Medical Systems Corporation · 2008 · recalled20052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 3 descendants3 generations forward: 4 descendants4 generations forward: 3 descendants K080403 · XTJF Type Q160VF1 · Olympus Medical Systems Corporation · 2008 · recalled0
  1. Root · 1 generation backK080403 XTJF Type Q160VF1Olympus Medical Systems Corporation · 2008 · Recalled
  2. This deviceK143153 Evis Exera II Duodenovideoscope Olympus TJF Type Q180VOlympus Medical Systems Corp. · 2016
  3. Descendant · cites it directly · 5 of its ownK193182 Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp. · 2020 · Recalled
  4. Descendant · 2 generations on · 4 of its ownK202661 Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp. · 2020 · Recalled
  5. Descendant · cites it directly · 4 of its ownK193202 EXALT Model D Single-Use Duodenoscope, EXALT ControllerBoston Scientific Corporation · 2019 · Recalled
  6. Descendant · 2 generations on · 3 of its ownK213606 EvoEndo Single-Use Endoscopy SystemEvoEndo, Inc. · 2022
  7. Descendant · cites it directly · 2 of its ownK201098 Ambu Duodeno SystemAmbu Innovation GmbH · 2020

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2019: 12020: 32022: 22024: 12025: 42026: 3
2019high 4 (2025) · low 0 (2021)2026

Most influential descendants

Showing 5 of 14 · See all
  • K193182Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp. · 2020 · cites it directly5descendants
  • K202661Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp. · 2020 · 2 generations on4descendants
  • K193202EXALT Model D Single-Use Duodenoscope, EXALT ControllerBoston Scientific Corporation · 2019 · cites it directly4descendants
  • K213606EvoEndo Single-Use Endoscopy SystemEvoEndo, Inc. · 2022 · 2 generations on3descendants
  • K201098Ambu Duodeno SystemAmbu Innovation GmbH · 2020 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K080403XTJF Type Q160VF1Olympus Medical Systems Corporation20082
0 (this device)K143153Evis Exera II Duodenovideoscope Olympus TJF Type Q180VOlympus Medical Systems Corp.2016—
+1K193182Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp.20204
+2K202661Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp.20201
+1K193202EXALT Model D Single-Use Duodenoscope, EXALT ControllerBoston Scientific Corporation20196
+2K213606EvoEndo Single-Use Endoscopy SystemEvoEndo, Inc.2022—
+1K201098Ambu Duodeno SystemAmbu Innovation GmbH2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202030
202220
202410
202540
202630
FIG. 1 — Lineage of K143153, 2008–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FDT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FDT device

  1. #1 K251861 FUJIFILM Endoscope Model ED-S100TP; FUJIFILM…score 13.7 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080403XTJF Type Q160VF1
Summary page 5
…ope, accessories, narrow band spectrum III. Predicate Device Information  Device Trade Name: XTJF Q160VF1 DUODENOVIDEOSCOPE  510(k) Number K080403  Decision Date 05/20/2008  Manufacturer AIZU OLYMPUS CO., LTD IV. Device Description T…
Olympus Medical Systems Corporation20082

What descends from it? (14)

See all 14

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K193182Evis Exera III Duodenovideoscope Olympus TJF-Q190VOlympus Medical Systems Corp.20205
K193202EXALT Model D Single-Use Duodenoscope, EXALT ControllerBoston Scientific Corporation20194
K201098Ambu Duodeno SystemAmbu Innovation GmbH20202
K233886Ambu® aScope™ Duodeno 2, Ambu® aBox™ 2Ambu A/S20241

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 25, 2023Class IIZ-0346-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Nov 2, 2021Class IIZ-0379-2022Under Investigation by firmUse of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.
Dec 17, 2018Class IIZ-1010-2019Under Investigation by firmThe adhesive used in the repair of the endoscope was incorrectly prepared. It is unknown what affect the nonconforming adhesive mixture would have on the durability or effectiveness of the gluing operation for the endoscope repair.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K193182 (Olympus Medical Systems Corp., 2020) has gone furthest, with 5 descendants of its own.

How does its category compare?

All FDT devices

FDT Duodenoscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FDT

2in 2025

2008: 12009: 12011: 12016: 12017: 22018: 32019: 52020: 42021: 22022: 12024: 22025: 22026 (partial year): 2
2007high 5 (2019) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FDT

72in 2020

2018: 63 days median2019: 88 days median2020: 72 days median
2007high 88 (2019) · low 63 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FDT

7so far in 2026, the most since 2007

2015: 12016: 12018: 12021: 12023: 32025: 12026 (partial year): 7
2007high 7 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FDT

3,946in 2025

2007: 42008: 112009: 182010: 122011: 102012: 152013: 182014: 362015: 2462016: 2472017: 7912018: 4812019: 1,3472020: 8222021: 1,2702022: 2,6192023: 3,0502024: 3,0192025: 3,946
2007high 3,946 (2025) · low 4 (2007)2025

MAUDE reports naming product code FDT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—04
20081—011
20091—018
20100—012
20111—010
20120—015
20130—018
20140—036
20150—1246
20161—1247
20172—0791
20183631481
201958801,347
20204720822
20212—11,270
20221—02,619
20230—33,050
20242—03,019
20252—13,946
20262—7—

What is it indicated for?

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K143153: Evis Exera II Duodenovideoscope Olympus TJF Type Q180V.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K143153

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026