# K143400: SOMATOM Definition AS/AS+

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Medical Solutions USA, Inc.
- Decided: Apr 8, 2015 after 131 days
- Product code: JAK (System, X-Ray, Tomography, Computed)
- Device class: 2
- Regulation: 21 CFR 892.1750
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 6 generations back to K982349 (1998). 8 ancestors in all.
32 later devices descend from it; 4 cite it directly.
0 of 41 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K120579: Somatom Definition Edge, Siemens Medical Systems, Inc. (2012) Summary page 5.
- K130901: Somatom Definition as Open, Siemens Medical Systems, Inc. (2014) Summary page 5.

## Most influential descendants
- K152036: SOMATOM Definition Edge, SOMATOM Definition AS/AS+ (2015), 28 descendants of its own
- K173630: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Deinition AS/AS+, SOMATOM Confidence (2018), 20 descendants of its own
- K173607: SOMATOM Edge Plus (2018), 20 descendants of its own
- K190578: SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Edge Plus, SOMATOM Definition AS/AS+, SOMATOM Confidence (2019), 19 descendants of its own
- K192061: SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open Pro, Scan & GO (2019), 16 descendants of its own

## Recalls of this device
- Jul 25, 2019 · Class II · Z-2440-2019: SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifi
- Oct 19, 2016 · Class II · Z-0604-2017: A problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becom
- Mar 22, 2016 · Class II · Z-1350-2016: Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the c
- Mar 2, 2016 · Class II · Z-1519-2016: Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of addit
- Aug 14, 2015 · Class II · Z-0021-2016: software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve a

## Indications for use
The Siemens SOMATOM Definition AS/ AS+ (Project P46) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes™ taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultancous translation of the patient.)