510(k) Report

510(k) · Cleared

Recalled · Class I · 2016

NeuroBlate System

K143457510(k) SpecialCleared Feb 19, 201578-day reviewClass II21 CFR 878.4810Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Monteris Medical, Inc. · GEX Powered Laser Surgical Instrument

Cleared on Feb 19, 2015 as substantially equivalent to K141983 (Monteris Medical Corp., 2014) after a 78-day review. Its family tree reaches back 8 generations to 1999. 2 of the 44 devices in its family, counting itself, have a Class I recall. 17 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Feb 19, 2015
Review
78 days
Class
Class II
Product code
GEX
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 8 generations to K983831 (1999). 26 ancestors in all, the oldest K913475 (1991). 17 later devices descend from it, 4 citing it directly. 2 of the 44 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2007: 1 other ancestor2006: 4 other ancestors2003: 1 other ancestor1999: 1 other ancestor2001: 3 other ancestors1998: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor2000: 1 other ancestor1991: 2 other ancestors2018: 1 descendant cleared2016: 1 descendant cleared2017: 4 descendants cleared, 1 with a Class I recall12020: 2 descendants cleared2023: 2 descendants cleared2022: 1 descendant cleared2024: 1 descendant cleared2026: 3 descendants cleared2025: 2 descendants clearedK983831 · VECTORVISION2 · Brainlab AG · 1999K014256 · Kolibri Igs Systems · Brainlab AG · 2002K042721 · Kolibri Spine · Brainlab AG · 2005K062086 · Navivision · Brainlab AG · 2006K081509 · Monteris Medical Autolitt Laser Probe · Monteris Medical, · 2009×2: 2 generations cleared close together, K120561 (2013), K131278 (2013) K141983 · Neuroblate System · Monteris Medical Corp. · 2014 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor4 generations back: 2 other ancestors5 generations back: 5 other ancestors6 generations back: 6 other ancestors7 generations back: 4 other ancestors 1 generation forward: 4 descendants2 generations forward: 5 descendants, 1 with a Class I recall13 generations forward: 3 descendants4 generations forward: 5 descendants K141983 · Neuroblate System · Monteris Medical Corp. · 2014 · recalledK131278 · Monteris Medical Neuroblate System · Monteris Medical, Inc. · 2013 · recalledK120561 · Neuroblate System · Monteris Medical, Inc. · 2013 · recalledK081509 · Monteris Medical Autolitt Laser Probe · Monteris Medical, · 2009K062086 · Navivision · Brainlab AG · 2006K042721 · Kolibri Spine · Brainlab AG · 2005K014256 · Kolibri Igs Systems · Brainlab AG · 2002K983831 · VECTORVISION2 · Brainlab AG · 1999−50
  1. Root · 8 generations backK983831 VECTORVISION2Brainlab AG · 1999
  2. 7 generations backK014256 Kolibri Igs SystemsBrainlab AG · 2002
  3. 6 generations backK042721 Kolibri SpineBrainlab AG · 2005
  4. 5 generations backK062086 NavivisionBrainlab AG · 2006
  5. 4 generations backK081509 Monteris Medical Autolitt Laser ProbeMonteris Medical, · 2009
  6. 3 generations backK120561 Neuroblate SystemMonteris Medical, Inc. · 2013 · Recalled
  7. 2 generations backK131278 Monteris Medical Neuroblate SystemMonteris Medical, Inc. · 2013 · Recalled
  8. 1 generation backK141983 Neuroblate SystemMonteris Medical Corp. · 2014 · Recalled
  9. This deviceK143457 NeuroBlate SystemMonteris Medical, Inc. · 2015
  10. Descendant · cites it directly · 16 of its ownK162762 NeuroBlate SystemMonteris Medical, Inc. · 2016
  11. Descendant · cites it directly · 15 of its ownK170724 NeuroBlate SystemMonteris Medical, · 2017
  12. Descendant · 2 generations on · 14 of its ownK171255 Monteris Medical NeuroBlate SystemMonteris Medical, Inc. · 2017 · Recalled (Class I)
  13. Descendant · cites it directly · 13 of its ownK172881 NeuroBlate SystemMonteris Medical, · 2017
  14. Descendant · 2 generations on · 12 of its ownK173305 NeuroBlate SystemMonteris Medical, · 2017

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2016: 12017: 42018: 12020: 22022: 12023: 22024: 12025: 22026: 3
2016high 4 (2017) · low 0 (2019)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 17 · See all
  • K162762NeuroBlate SystemMonteris Medical, Inc. · 2016 · cites it directly16descendants
  • K170724NeuroBlate SystemMonteris Medical, · 2017 · cites it directly15descendants
  • K171255Monteris Medical NeuroBlate SystemMonteris Medical, Inc. · 2017 · 2 generations on14descendants
  • K172881NeuroBlate SystemMonteris Medical, · 2017 · cites it directly13descendants
  • K173305NeuroBlate SystemMonteris Medical, · 2017 · 2 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K983831VECTORVISION2Brainlab AG1999—
−7K014256Kolibri Igs SystemsBrainlab AG2002—
−6K042721Kolibri SpineBrainlab AG2005—
−5K062086NavivisionBrainlab AG2006—
−4K081509Monteris Medical Autolitt Laser ProbeMonteris Medical,2009—
−3K120561Neuroblate SystemMonteris Medical, Inc.20131
−2K131278Monteris Medical Neuroblate SystemMonteris Medical, Inc.20131
−1K141983Neuroblate SystemMonteris Medical Corp.20142
0 (this device)K143457NeuroBlate SystemMonteris Medical, Inc.2015—
+1K162762NeuroBlate SystemMonteris Medical, Inc.2016—
+1K170724NeuroBlate SystemMonteris Medical,2017—
+2K171255Monteris Medical NeuroBlate SystemMonteris Medical, Inc.2017Class I
+1K172881NeuroBlate SystemMonteris Medical,2017—
+2K173305NeuroBlate SystemMonteris Medical,2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199120
199510
199610
199710
199810
199910
200010
200130
200310
200640
200710
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201741
201810
202020
202210
202320
202410
202520
202630
FIG. 1 — Lineage of K143457, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEX device

  1. #1 K222064 The Alma Soprano Titaniumscore 38.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141983Neuroblate System
Summary page 4
…ice Manufacturer: Monteris Medical Predicate Device Common Name: Monteris NeuroBlate™ System Predicate Device Premarket Notification # K141983 Predicate Device Regulation: 21 CFR 878.4810 Laser surgical instru…
Monteris Medical Corp.20142

Also cleared against K141983: 4 other devices →

What descends from it? (17)

See all 17

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162762NeuroBlate SystemMonteris Medical, Inc.201616
K170724NeuroBlate SystemMonteris Medical,201715
K172881NeuroBlate SystemMonteris Medical,201713
K182036NeuroBlate SystemMonteris Medical,201811

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 25, 2016Class IZ-1614-2016Device DesignMonteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K171255Monteris Medical NeuroBlate SystemMonteris Medical, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 9 of the 26 ancestors have a recall of any class.
  • Of its descendants, K162762 (Monteris Medical, Inc., 2016) has gone furthest, with 16 descendants of its own.
  • 4 other devices also cite its predicate K141983.

How does its category compare?

All GEX devices

GEX Powered Laser Surgical Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · GEX

83in 2025

2007: 732008: 1072009: 692010: 502011: 412012: 452013: 462014: 612015: 472016: 512017: 672018: 512019: 412020: 522021: 532022: 602023: 522024: 822025: 832026 (partial year): 57
2007high 107 (2008) · low 41 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEX

90in 2025

2007: 79 days median2008: 86 days median2009: 100 days median2010: 96 days median2011: 83 days median2012: 118 days median2013: 144 days median2014: 147 days median2015: 112 days median2016: 114 days median2017: 90 days median2018: 78 days median2019: 88 days median2020: 90 days median2021: 123 days median2022: 152 days median2023: 87 days median2024: 90 days median2025: 90 days median2026 (partial year): 90 days median
2007high 152 (2022) · low 78 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEX

3in 2024

2007: 92008: 22010: 42011: 182012: 52013: 32014: 62016: 12017: 5, 1 Class I2018: 52019: 62020: 12023: 62024: 32026 (partial year): 2
2007high 18 (2011) · low 0 (2009)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · GEX

2,720in 2025

2007: 3422008: 2322009: 2092010: 1,7112011: 3,0242012: 2,0002013: 2,2362014: 2,0362015: 2,0992016: 2,5852017: 2,9772018: 3,5252019: 2,7092020: 2,1832021: 3,4652022: 2,2882023: 2,3372024: 2,1872025: 2,720
2007high 3,525 (2018) · low 209 (2009)2025

MAUDE reports naming product code GEX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200773799342
2008107862232
2009691000209
2010509641,711
20114183183,024
20124511852,000
20134614432,236
20146114762,036
20154711202,099
20165111412,585
201767905 (1)2,977
2018517853,525
2019418862,709
2020529012,183
20215312303,465
20226015202,288
2023528762,337
2024829032,187
2025839002,720
202657902—

What is it indicated for?

The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI…

Read the full statement (119 words)

The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through

interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI

visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible

(conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI

images.

When interpreted by a trained physician, this System provides information that may be useful in the determination or

assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™

System analysis.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K143457: NeuroBlate System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K143457

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026