510(k) · Cleared
Recalled · Class I · 2026TRUE METRIX GO Self Monitoring Blood Glucose System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Trividia Health, Inc. · NBW System, Test, Blood Glucose, Over The Counter
Cleared on Apr 1, 2016 after a 473-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has a Class I recall. It has been recalled once, including a Class I recall.
- Decided
- Apr 1, 2016
- Review
- 473 days
- Class
- Class II
- Product code
- NBW
- Recalls
- 1 (1 Class I)
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NBW device
- #1 K260545 CONTOUR® PLUS BLUE Blood Glucose Monitoring…score 14.6 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 6, 2026 | Class I | Z-1548-2026 | Software design | The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a)… |
No ancestor or descendant has a Class I recall.
How does its category compare?
All NBW devicesNBW System, Test, Blood Glucose, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · NBW
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NBW
243in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NBW
1in 2025
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · NBW
4,745in 2025
MAUDE reports naming product code NBW, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 29 | 85 | 9 | 7,991 |
| 2008 | 37 | 128 | 2 | 7,704 |
| 2009 | 38 | 140 | 5 | 10,963 |
| 2010 | 31 | 132 | 11 | 11,631 |
| 2011 | 30 | 328 | 1 | 22,909 |
| 2012 | 30 | 230 | 1 | 23,449 |
| 2013 | 47 | 167 | 5 (2) | 35,408 |
| 2014 | 24 | 202 | 1 | 38,524 |
| 2015 | 32 | 237 | 0 | 62,111 |
| 2016 | 20 | 233 | 1 | 76,621 |
| 2017 | 25 | 244 | 0 | 37,538 |
| 2018 | 19 | 168 | 0 | 16,911 |
| 2019 | 10 | 127 | 14 | 14,472 |
| 2020 | 7 | 121 | 0 | 10,703 |
| 2021 | 7 | 465 | 0 | 9,198 |
| 2022 | 12 | 410 | 0 | 8,103 |
| 2023 | 5 | 193 | 0 | 7,256 |
| 2024 | 7 | 243 | 0 | 6,220 |
| 2025 | 2 | — | 1 | 4,745 |
| 2026 | 5 | 190 | 0 | — |
What is it indicated for?
The TRUE METRIX GO Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The TRUE METRIX GO Self Monitoring Blood Glucose System is intended to be used by a single person and not shared. The TRUE METRIX GO Self Monitoring Blood Glucose System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home…
Read the full statement (150 words)
The TRUE METRIX GO Self Monitoring Blood Glucose System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The TRUE METRIX GO Self Monitoring Blood Glucose System is intended to be used by a single person and not shared.
The TRUE METRIX GO Self Monitoring Blood Glucose System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The TRUE METRIX GO Self Monitoring Blood Glucose System should not be used for the diagnosis or screening of diabetes, for neonatal use or for alternative site testing.
The TRUE METRIX Self Monitoring Blood Glucose Test Strips are for use with the TRUE METRIX GO Self Monitoring Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.
Indications for use as stated in the FDA 510(k) statement.
https://510kreport.com/device/K143548 · printed Nov 5, 2026
Cite this page
510(k) Report. “K143548: TRUE METRIX GO Self Monitoring Blood Glucose System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K143548
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026