510(k) Report

510(k) · Cleared

Cleared

PulmoVia Working Channel

K150280510(k) TraditionalCleared Jun 11, 2015126-day reviewClass II21 CFR 874.4680

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sanovas, Inc. · EOQ Bronchoscope (Flexible Or Rigid)

Cleared on Jun 11, 2015 as substantially equivalent to K131234 (Broncus Medical, Inc., 2013) and 1 other predicate after a 126-day review. Its family tree reaches back 3 generations to 1993. 1 of the 7 devices in its family, counting itself, has a Class I recall.

Decided
Jun 11, 2015
Review
126 days
Class
Class II
Product code
EOQ
Recalls
None
Predicates
2

Its longest line of descent runs back 3 generations to K931154 (1993). 6 ancestors in all, the oldest K923487 (1993). No later device cites it as a predicate yet. 1 of the 7 devices in this family, counting this one, has a Class I recall.

  1. Root · 3 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 2 generations backK060243 Olympus Guide Sheath, Model XBO1-836-13Olympus Medical Systems Corporation · 2006 · Recalled (Class I)
  3. 1 generation backK131234 Lungpoint Tools (Lungpoint Sheath, Lungpoint Dilation Balloon)Broncus Medical, Inc. · 2013
  4. This deviceK150280 PulmoVia Working ChannelSanovas, Inc. · 2015

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K931154Evis 200 SystemOlympus Corp.1993—
−2K060243Olympus Guide Sheath, Model XBO1-836-13Olympus Medical Systems Corporation2006Class I
−1K131234Lungpoint Tools (Lungpoint Sheath, Lungpoint Dilation Balloon)Broncus Medical, Inc.2013—
0 (this device)K150280PulmoVia Working ChannelSanovas, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
200210
201110
FIG. 1 — Lineage of K150280, 1993–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EOQ device

  1. #1 K223299 Ambu® aScope™ 5 Broncho HD 5.0/2.2, Ambu®…score 18.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131234Lungpoint Tools (Lungpoint Sheath, Lungpoint Dilation Balloon)
Summary page 4
…tion: Class II Product Code: EOQ 3. PREDICATE DEVICES • LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon), K131234 • KimVent BAL Cath Bronchial Aspirate Sampling Catheter, K112562 • Olympus Guide Sheath, K060243 4. DEVICE D…
Broncus Medical, Inc.2013—
*K112562Bal Cath Bronchial Aspirate Sampling Catheter
Summary page 4
…REDICATE DEVICES • LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon), K131234 • KimVent BAL Cath Bronchial Aspirate Sampling Catheter, K112562 • Olympus Guide Sheath, K060243 4. DEVICE DESCRIPTION The PulmoVia Working Channel is a single-use disposable, t…
Kimberly-Clark2011—
†K060243Olympus Guide Sheath, Model XBO1-836-13Olympus Medical Systems Corporation2006Class I

Also cleared against K131234: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K060243Olympus Guide Sheath, Model XBO1-836-13Olympus Medical Systems Corporation · 2006 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 7.3 years on median.
  • 1 of the 6 ancestors has a recall of any class; 1 has a Class I recall.
  • 3 other devices also cite its predicate K131234.

How does its category compare?

All EOQ devices

EOQ Bronchoscope (Flexible Or Rigid): FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · EOQ

20in 2025

2007: 22008: 22009: 32012: 22013: 42014: 32015: 62016: 32017: 42018: 72019: 112020: 102021: 152022: 112023: 122024: 132025: 202026 (partial year): 9
2007high 20 (2025) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EOQ

147in 2025

2009: 24 days median2013: 178 days median2014: 302 days median2015: 206 days median2016: 262 days median2017: 136 days median2018: 212 days median2019: 147 days median2020: 103 days median2021: 169 days median2022: 134 days median2023: 160 days median2024: 130 days median2025: 147 days median2026 (partial year): 206 days median
2007high 302 (2014) · low 24 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EOQ

25in 2025

2011: 12012: 42016: 12019: 12020: 12021: 92022: 52023: 18, 4 Class I2024: 42025: 25, 24 Class I
2007high 25 (2025) · low 0 (2007)2026

Class I (28 since 2007) Class II/III not classified

Adverse event reports per year · EOQ

6,546in 2025

2007: 392008: 872009: 1142010: 1002011: 962012: 932013: 932014: 982015: 1322016: 2972017: 3372018: 2792019: 6542020: 5152021: 1,3582022: 2,5412023: 5,0642024: 6,0512025: 6,546
2007high 6,546 (2025) · low 39 (2007)2025

MAUDE reports naming product code EOQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—039
20082—087
20093240114
20100—0100
20110—196
20122—493
20134178093
20143302098
201562060132
201632621297
201741360337
201872120279
2019111471654
2020101031515
20211516991,358
20221113452,541
20231216018 (4)5,064
20241313046,051
20252014725 (24)6,546
202692060—

What is it indicated for?

The PulmoVia Working Channel is intended to be used as an access channel through which a bronchoscope and other endoscopic tools may be introduced to treat targeted tissue of the airways. The PulmoVia Working Channel can also provide means for bronchoalveolar lavage. The PulmoVia Working Channel is indicated for adult patients.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K150280: PulmoVia Working Channel.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K150280

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026